Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius
BACKGROUND: The World Health Organization recommends reliable point-of-care (POC) diagnostic testing to eliminate schistosomiasis. Lateral flow immunoassay that detects schistosome circulating cathodic antigen (CCA) in urine to establish prevalence thresholds for intervention in endemic areas is rec...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Sociedade Brasileira de Medicina Tropical - SBMT
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8909435/ https://www.ncbi.nlm.nih.gov/pubmed/35239906 http://dx.doi.org/10.1590/0037-8682-0389-2021 |
_version_ | 1784666149458280448 |
---|---|
author | Favre, Tereza Cristina Beck, Lilian Christina Nóbrega Holsback Bezerra, Fernando Schemelzer Moraes Graeff-Teixeira, Carlos Coelho, Paulo Marcos Zech Enk, Martin Johannes Katz, Naftale Oliveira, Ricardo Riccio dos Reis, Mitermayer Galvão Pieri, Otávio Sarmento |
author_facet | Favre, Tereza Cristina Beck, Lilian Christina Nóbrega Holsback Bezerra, Fernando Schemelzer Moraes Graeff-Teixeira, Carlos Coelho, Paulo Marcos Zech Enk, Martin Johannes Katz, Naftale Oliveira, Ricardo Riccio dos Reis, Mitermayer Galvão Pieri, Otávio Sarmento |
author_sort | Favre, Tereza Cristina |
collection | PubMed |
description | BACKGROUND: The World Health Organization recommends reliable point-of-care (POC) diagnostic testing to eliminate schistosomiasis. Lateral flow immunoassay that detects schistosome circulating cathodic antigen (CCA) in urine to establish prevalence thresholds for intervention in endemic areas is recommended. Stored urine may be useful if surveying at-risk populations is delayed or interrupted by unforeseen circumstances, such as the current COVID-19 pandemic. This study evaluated the manufacturer’s claim that Schistosoma mansoni infection can be reliably diagnosed in urine samples stored at -20°C for one year. METHODS: Two-hundred-forty-two subjects from an endemic site in Brazil provided one urine sample each for testing with URINE CCA (SCHISTO) ECO TESTE® (POC-ECO) and one stool sample each for testing with Kato-Katz (KK) and Helmintex® (HTX) as a robust reference standard for infection status. At least 2 ml of urine from each participant was stored at -20°C; after one year, 76 samples were randomly selected for POC-ECO retesting. RESULTS: The POC-ECO agreement between freshly collected and stored urine was inadequate considering trace results as positive (Cohen’s kappa coefficient κ = 0.08) and negative (κ = 0.36). POC-ECO accuracy was not significantly greater than that of routine KK (54%; 95% confidence interval: 42.1%-65.5%). CONCLUSIONS: The precision and accuracy of POC-ECO have to be optimized in both freshly collected and stored urine before it can be recommended for use in control programs in Brazil. |
format | Online Article Text |
id | pubmed-8909435 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Sociedade Brasileira de Medicina Tropical - SBMT |
record_format | MEDLINE/PubMed |
spelling | pubmed-89094352022-03-21 Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius Favre, Tereza Cristina Beck, Lilian Christina Nóbrega Holsback Bezerra, Fernando Schemelzer Moraes Graeff-Teixeira, Carlos Coelho, Paulo Marcos Zech Enk, Martin Johannes Katz, Naftale Oliveira, Ricardo Riccio dos Reis, Mitermayer Galvão Pieri, Otávio Sarmento Rev Soc Bras Med Trop Major Article BACKGROUND: The World Health Organization recommends reliable point-of-care (POC) diagnostic testing to eliminate schistosomiasis. Lateral flow immunoassay that detects schistosome circulating cathodic antigen (CCA) in urine to establish prevalence thresholds for intervention in endemic areas is recommended. Stored urine may be useful if surveying at-risk populations is delayed or interrupted by unforeseen circumstances, such as the current COVID-19 pandemic. This study evaluated the manufacturer’s claim that Schistosoma mansoni infection can be reliably diagnosed in urine samples stored at -20°C for one year. METHODS: Two-hundred-forty-two subjects from an endemic site in Brazil provided one urine sample each for testing with URINE CCA (SCHISTO) ECO TESTE® (POC-ECO) and one stool sample each for testing with Kato-Katz (KK) and Helmintex® (HTX) as a robust reference standard for infection status. At least 2 ml of urine from each participant was stored at -20°C; after one year, 76 samples were randomly selected for POC-ECO retesting. RESULTS: The POC-ECO agreement between freshly collected and stored urine was inadequate considering trace results as positive (Cohen’s kappa coefficient κ = 0.08) and negative (κ = 0.36). POC-ECO accuracy was not significantly greater than that of routine KK (54%; 95% confidence interval: 42.1%-65.5%). CONCLUSIONS: The precision and accuracy of POC-ECO have to be optimized in both freshly collected and stored urine before it can be recommended for use in control programs in Brazil. Sociedade Brasileira de Medicina Tropical - SBMT 2022-02-25 /pmc/articles/PMC8909435/ /pubmed/35239906 http://dx.doi.org/10.1590/0037-8682-0389-2021 Text en https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License |
spellingShingle | Major Article Favre, Tereza Cristina Beck, Lilian Christina Nóbrega Holsback Bezerra, Fernando Schemelzer Moraes Graeff-Teixeira, Carlos Coelho, Paulo Marcos Zech Enk, Martin Johannes Katz, Naftale Oliveira, Ricardo Riccio dos Reis, Mitermayer Galvão Pieri, Otávio Sarmento Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius |
title | Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius |
title_full | Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius |
title_fullStr | Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius |
title_full_unstemmed | Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius |
title_short | Reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of Brazil after one year of storage at -20 degrees Celsius |
title_sort | reliability of point-of-care circulating cathodic antigen assay for diagnosing schistosomiasis mansoni in urine samples from an endemic area of brazil after one year of storage at -20 degrees celsius |
topic | Major Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8909435/ https://www.ncbi.nlm.nih.gov/pubmed/35239906 http://dx.doi.org/10.1590/0037-8682-0389-2021 |
work_keys_str_mv | AT favreterezacristina reliabilityofpointofcarecirculatingcathodicantigenassayfordiagnosingschistosomiasismansoniinurinesamplesfromanendemicareaofbrazilafteroneyearofstorageat20degreescelsius AT becklilianchristinanobregaholsback reliabilityofpointofcarecirculatingcathodicantigenassayfordiagnosingschistosomiasismansoniinurinesamplesfromanendemicareaofbrazilafteroneyearofstorageat20degreescelsius AT bezerrafernandoschemelzermoraes reliabilityofpointofcarecirculatingcathodicantigenassayfordiagnosingschistosomiasismansoniinurinesamplesfromanendemicareaofbrazilafteroneyearofstorageat20degreescelsius AT graeffteixeiracarlos reliabilityofpointofcarecirculatingcathodicantigenassayfordiagnosingschistosomiasismansoniinurinesamplesfromanendemicareaofbrazilafteroneyearofstorageat20degreescelsius AT coelhopaulomarcoszech reliabilityofpointofcarecirculatingcathodicantigenassayfordiagnosingschistosomiasismansoniinurinesamplesfromanendemicareaofbrazilafteroneyearofstorageat20degreescelsius AT enkmartinjohannes reliabilityofpointofcarecirculatingcathodicantigenassayfordiagnosingschistosomiasismansoniinurinesamplesfromanendemicareaofbrazilafteroneyearofstorageat20degreescelsius AT katznaftale reliabilityofpointofcarecirculatingcathodicantigenassayfordiagnosingschistosomiasismansoniinurinesamplesfromanendemicareaofbrazilafteroneyearofstorageat20degreescelsius AT oliveiraricardoriccio reliabilityofpointofcarecirculatingcathodicantigenassayfordiagnosingschistosomiasismansoniinurinesamplesfromanendemicareaofbrazilafteroneyearofstorageat20degreescelsius AT dosreismitermayergalvao reliabilityofpointofcarecirculatingcathodicantigenassayfordiagnosingschistosomiasismansoniinurinesamplesfromanendemicareaofbrazilafteroneyearofstorageat20degreescelsius AT pieriotaviosarmento reliabilityofpointofcarecirculatingcathodicantigenassayfordiagnosingschistosomiasismansoniinurinesamplesfromanendemicareaofbrazilafteroneyearofstorageat20degreescelsius |