Cargando…

The Movesense Medical Sensor Chest Belt Device as Single Channel ECG for RR Interval Detection and HRV Analysis during Resting State and Incremental Exercise: A Cross-Sectional Validation Study

The value of heart rate variability (HRV) in the fields of health, disease, and exercise science has been established through numerous investigations. The typical mobile-based HRV device simply records interbeat intervals, without differentiation between noise or arrythmia as can be done with an ele...

Descripción completa

Detalles Bibliográficos
Autores principales: Rogers, Bruce, Schaffarczyk, Marcelle, Clauß, Martina, Mourot, Laurent, Gronwald, Thomas
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8914935/
https://www.ncbi.nlm.nih.gov/pubmed/35271179
http://dx.doi.org/10.3390/s22052032
_version_ 1784667879195541504
author Rogers, Bruce
Schaffarczyk, Marcelle
Clauß, Martina
Mourot, Laurent
Gronwald, Thomas
author_facet Rogers, Bruce
Schaffarczyk, Marcelle
Clauß, Martina
Mourot, Laurent
Gronwald, Thomas
author_sort Rogers, Bruce
collection PubMed
description The value of heart rate variability (HRV) in the fields of health, disease, and exercise science has been established through numerous investigations. The typical mobile-based HRV device simply records interbeat intervals, without differentiation between noise or arrythmia as can be done with an electrocardiogram (ECG). The intent of this report is to validate a new single channel ECG device, the Movesense Medical sensor, against a conventional 12 channel ECG. A heterogeneous group of 21 participants performed an incremental cycling ramp to failure with measurements of HRV, before (PRE), during (EX), and after (POST). Results showed excellent correlations between devices for linear indexes with Pearson’s r between 0.98 to 1.0 for meanRR, SDNN, RMSSD, and 0.95 to 0.97 for the non-linear index DFA a1 during PRE, EX, and POST. There was no significant difference in device specific meanRR during PRE and POST. Bland–Altman analysis showed high agreement between devices (PRE and POST: meanRR bias of 0.0 and 0.4 ms, LOA of 1.9 to −1.8 ms and 2.3 to −1.5; EX: meanRR bias of 11.2 to 6.0 ms; LOA of 29.8 to −7.4 ms during low intensity exercise and 8.5 to 3.5 ms during high intensity exercise). The Movesense Medical device can be used in lieu of a reference ECG for the calculation of HRV with the potential to differentiate noise from atrial fibrillation and represents a significant advance in both a HR and HRV recording device in a chest belt form factor for lab-based or remote field-application.
format Online
Article
Text
id pubmed-8914935
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-89149352022-03-12 The Movesense Medical Sensor Chest Belt Device as Single Channel ECG for RR Interval Detection and HRV Analysis during Resting State and Incremental Exercise: A Cross-Sectional Validation Study Rogers, Bruce Schaffarczyk, Marcelle Clauß, Martina Mourot, Laurent Gronwald, Thomas Sensors (Basel) Article The value of heart rate variability (HRV) in the fields of health, disease, and exercise science has been established through numerous investigations. The typical mobile-based HRV device simply records interbeat intervals, without differentiation between noise or arrythmia as can be done with an electrocardiogram (ECG). The intent of this report is to validate a new single channel ECG device, the Movesense Medical sensor, against a conventional 12 channel ECG. A heterogeneous group of 21 participants performed an incremental cycling ramp to failure with measurements of HRV, before (PRE), during (EX), and after (POST). Results showed excellent correlations between devices for linear indexes with Pearson’s r between 0.98 to 1.0 for meanRR, SDNN, RMSSD, and 0.95 to 0.97 for the non-linear index DFA a1 during PRE, EX, and POST. There was no significant difference in device specific meanRR during PRE and POST. Bland–Altman analysis showed high agreement between devices (PRE and POST: meanRR bias of 0.0 and 0.4 ms, LOA of 1.9 to −1.8 ms and 2.3 to −1.5; EX: meanRR bias of 11.2 to 6.0 ms; LOA of 29.8 to −7.4 ms during low intensity exercise and 8.5 to 3.5 ms during high intensity exercise). The Movesense Medical device can be used in lieu of a reference ECG for the calculation of HRV with the potential to differentiate noise from atrial fibrillation and represents a significant advance in both a HR and HRV recording device in a chest belt form factor for lab-based or remote field-application. MDPI 2022-03-05 /pmc/articles/PMC8914935/ /pubmed/35271179 http://dx.doi.org/10.3390/s22052032 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Rogers, Bruce
Schaffarczyk, Marcelle
Clauß, Martina
Mourot, Laurent
Gronwald, Thomas
The Movesense Medical Sensor Chest Belt Device as Single Channel ECG for RR Interval Detection and HRV Analysis during Resting State and Incremental Exercise: A Cross-Sectional Validation Study
title The Movesense Medical Sensor Chest Belt Device as Single Channel ECG for RR Interval Detection and HRV Analysis during Resting State and Incremental Exercise: A Cross-Sectional Validation Study
title_full The Movesense Medical Sensor Chest Belt Device as Single Channel ECG for RR Interval Detection and HRV Analysis during Resting State and Incremental Exercise: A Cross-Sectional Validation Study
title_fullStr The Movesense Medical Sensor Chest Belt Device as Single Channel ECG for RR Interval Detection and HRV Analysis during Resting State and Incremental Exercise: A Cross-Sectional Validation Study
title_full_unstemmed The Movesense Medical Sensor Chest Belt Device as Single Channel ECG for RR Interval Detection and HRV Analysis during Resting State and Incremental Exercise: A Cross-Sectional Validation Study
title_short The Movesense Medical Sensor Chest Belt Device as Single Channel ECG for RR Interval Detection and HRV Analysis during Resting State and Incremental Exercise: A Cross-Sectional Validation Study
title_sort movesense medical sensor chest belt device as single channel ecg for rr interval detection and hrv analysis during resting state and incremental exercise: a cross-sectional validation study
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8914935/
https://www.ncbi.nlm.nih.gov/pubmed/35271179
http://dx.doi.org/10.3390/s22052032
work_keys_str_mv AT rogersbruce themovesensemedicalsensorchestbeltdeviceassinglechannelecgforrrintervaldetectionandhrvanalysisduringrestingstateandincrementalexerciseacrosssectionalvalidationstudy
AT schaffarczykmarcelle themovesensemedicalsensorchestbeltdeviceassinglechannelecgforrrintervaldetectionandhrvanalysisduringrestingstateandincrementalexerciseacrosssectionalvalidationstudy
AT claußmartina themovesensemedicalsensorchestbeltdeviceassinglechannelecgforrrintervaldetectionandhrvanalysisduringrestingstateandincrementalexerciseacrosssectionalvalidationstudy
AT mourotlaurent themovesensemedicalsensorchestbeltdeviceassinglechannelecgforrrintervaldetectionandhrvanalysisduringrestingstateandincrementalexerciseacrosssectionalvalidationstudy
AT gronwaldthomas themovesensemedicalsensorchestbeltdeviceassinglechannelecgforrrintervaldetectionandhrvanalysisduringrestingstateandincrementalexerciseacrosssectionalvalidationstudy
AT rogersbruce movesensemedicalsensorchestbeltdeviceassinglechannelecgforrrintervaldetectionandhrvanalysisduringrestingstateandincrementalexerciseacrosssectionalvalidationstudy
AT schaffarczykmarcelle movesensemedicalsensorchestbeltdeviceassinglechannelecgforrrintervaldetectionandhrvanalysisduringrestingstateandincrementalexerciseacrosssectionalvalidationstudy
AT claußmartina movesensemedicalsensorchestbeltdeviceassinglechannelecgforrrintervaldetectionandhrvanalysisduringrestingstateandincrementalexerciseacrosssectionalvalidationstudy
AT mourotlaurent movesensemedicalsensorchestbeltdeviceassinglechannelecgforrrintervaldetectionandhrvanalysisduringrestingstateandincrementalexerciseacrosssectionalvalidationstudy
AT gronwaldthomas movesensemedicalsensorchestbeltdeviceassinglechannelecgforrrintervaldetectionandhrvanalysisduringrestingstateandincrementalexerciseacrosssectionalvalidationstudy