CRISIS ventilator: A 3D printed option for ventilator surge in mass respiratory pandemics
BACKGROUND: The COVID-19 pandemic revealed flaws in the stockpiling and distribution of ventilators. In this study, we assessed the durability, sterilizability, and performance of a 3D-printed ventilator. METHODS: SLS-printed devices were dropped from 1.83 m and autoclaved before evaluation on a COV...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8915621/ https://www.ncbi.nlm.nih.gov/pubmed/35379482 http://dx.doi.org/10.1016/j.amjsurg.2022.02.064 |
Sumario: | BACKGROUND: The COVID-19 pandemic revealed flaws in the stockpiling and distribution of ventilators. In this study, we assessed the durability, sterilizability, and performance of a 3D-printed ventilator. METHODS: SLS-printed devices were dropped from 1.83 m and autoclaved before evaluation on a COVID-19 simulated patient. The respiratory performance of an extrusion-printed device was studied using a variable compliance model. Ranges of sustainable respiratory rates were evaluated as a function of tidal volume. RESULTS: Autoclaving and dropping the device did not negatively impact minute ventilation or PIP for sustained ventilation. Equivalence was significant across all measures except for comparing the autoclaved and dropped with p = 0.06. Extrusion produced ventilators achieved minute ventilation ranging from 4.1 to 12.2 L/min for all simulated compliances; there was an inverse correlation between tidal volume and respiratory rate. CONCLUSION: The CRISIS ventilator is a durable, sterilizable, and reusable 3D-printed ventilator using off-the-shelf materials which could be employed variety of adult lung diseases. Further in-vivo testing is needed. |
---|