Heart Rate Measurement Accuracy of Fitbit Charge 4 and Samsung Galaxy Watch Active2: Device Evaluation Study

BACKGROUND: Fitness trackers and smart watches are frequently used to collect data in longitudinal medical studies. They allow continuous recording in real-life settings, potentially revealing previously uncaptured variabilities of biophysiological parameters and diseases. Adequate device accuracy i...

Descripción completa

Detalles Bibliográficos
Autores principales: Nissen, Michael, Slim, Syrine, Jäger, Katharina, Flaucher, Madeleine, Huebner, Hanna, Danzberger, Nina, Fasching, Peter A, Beckmann, Matthias W, Gradl, Stefan, Eskofier, Bjoern M
Formato: Online Artículo Texto
Lenguaje:English
Publicado: JMIR Publications 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8924780/
https://www.ncbi.nlm.nih.gov/pubmed/35230250
http://dx.doi.org/10.2196/33635
_version_ 1784669932473024512
author Nissen, Michael
Slim, Syrine
Jäger, Katharina
Flaucher, Madeleine
Huebner, Hanna
Danzberger, Nina
Fasching, Peter A
Beckmann, Matthias W
Gradl, Stefan
Eskofier, Bjoern M
author_facet Nissen, Michael
Slim, Syrine
Jäger, Katharina
Flaucher, Madeleine
Huebner, Hanna
Danzberger, Nina
Fasching, Peter A
Beckmann, Matthias W
Gradl, Stefan
Eskofier, Bjoern M
author_sort Nissen, Michael
collection PubMed
description BACKGROUND: Fitness trackers and smart watches are frequently used to collect data in longitudinal medical studies. They allow continuous recording in real-life settings, potentially revealing previously uncaptured variabilities of biophysiological parameters and diseases. Adequate device accuracy is a prerequisite for meaningful research. OBJECTIVE: This study aims to assess the heart rate recording accuracy in two previously unvalidated devices: Fitbit Charge 4 and Samsung Galaxy Watch Active2. METHODS: Participants performed a study protocol comprising 5 resting and sedentary, 2 low-intensity, and 3 high-intensity exercise phases, lasting an average of 19 minutes 27 seconds. Participants wore two wearables simultaneously during all activities: Fitbit Charge 4 and Samsung Galaxy Watch Active2. Reference heart rate data were recorded using a medically certified Holter electrocardiogram. The data of the reference and evaluated devices were synchronized and compared at 1-second intervals. The mean, mean absolute error, mean absolute percentage error, Lin concordance correlation coefficient, Pearson correlation coefficient, and Bland-Altman plots were analyzed. RESULTS: A total of 23 healthy adults (mean age 24.2, SD 4.6 years) participated in our study. Overall, and across all activities, the Fitbit Charge 4 slightly underestimated the heart rate, whereas the Samsung Galaxy Watch Active2 overestimated it (−1.66 beats per minute [bpm]/3.84 bpm). The Fitbit Charge 4 achieved a lower mean absolute error during resting and sedentary activities (seated rest: 7.8 vs 9.4; typing: 8.1 vs 11.6; laying down [left]: 7.2 vs 9.4; laying down [back]: 6.0 vs 8.6; and walking slowly: 6.8 vs 7.7 bpm), whereas the Samsung Galaxy Watch Active2 performed better during and after low- and high-intensity activities (standing up: 12.3 vs 9.0; walking fast: 6.1 vs 5.8; stairs: 8.8 vs 6.9; squats: 15.7 vs 6.1; resting: 9.6 vs 5.6 bpm). CONCLUSIONS: Device accuracy varied with activity. Overall, both devices achieved a mean absolute percentage error of just <10%. Thus, they were considered to produce valid results based on the limits established by previous work in the field. Neither device reached sufficient accuracy during seated rest or keyboard typing. Thus, both devices may be eligible for use in respective studies; however, researchers should consider their individual study requirements.
format Online
Article
Text
id pubmed-8924780
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher JMIR Publications
record_format MEDLINE/PubMed
spelling pubmed-89247802022-03-17 Heart Rate Measurement Accuracy of Fitbit Charge 4 and Samsung Galaxy Watch Active2: Device Evaluation Study Nissen, Michael Slim, Syrine Jäger, Katharina Flaucher, Madeleine Huebner, Hanna Danzberger, Nina Fasching, Peter A Beckmann, Matthias W Gradl, Stefan Eskofier, Bjoern M JMIR Form Res Original Paper BACKGROUND: Fitness trackers and smart watches are frequently used to collect data in longitudinal medical studies. They allow continuous recording in real-life settings, potentially revealing previously uncaptured variabilities of biophysiological parameters and diseases. Adequate device accuracy is a prerequisite for meaningful research. OBJECTIVE: This study aims to assess the heart rate recording accuracy in two previously unvalidated devices: Fitbit Charge 4 and Samsung Galaxy Watch Active2. METHODS: Participants performed a study protocol comprising 5 resting and sedentary, 2 low-intensity, and 3 high-intensity exercise phases, lasting an average of 19 minutes 27 seconds. Participants wore two wearables simultaneously during all activities: Fitbit Charge 4 and Samsung Galaxy Watch Active2. Reference heart rate data were recorded using a medically certified Holter electrocardiogram. The data of the reference and evaluated devices were synchronized and compared at 1-second intervals. The mean, mean absolute error, mean absolute percentage error, Lin concordance correlation coefficient, Pearson correlation coefficient, and Bland-Altman plots were analyzed. RESULTS: A total of 23 healthy adults (mean age 24.2, SD 4.6 years) participated in our study. Overall, and across all activities, the Fitbit Charge 4 slightly underestimated the heart rate, whereas the Samsung Galaxy Watch Active2 overestimated it (−1.66 beats per minute [bpm]/3.84 bpm). The Fitbit Charge 4 achieved a lower mean absolute error during resting and sedentary activities (seated rest: 7.8 vs 9.4; typing: 8.1 vs 11.6; laying down [left]: 7.2 vs 9.4; laying down [back]: 6.0 vs 8.6; and walking slowly: 6.8 vs 7.7 bpm), whereas the Samsung Galaxy Watch Active2 performed better during and after low- and high-intensity activities (standing up: 12.3 vs 9.0; walking fast: 6.1 vs 5.8; stairs: 8.8 vs 6.9; squats: 15.7 vs 6.1; resting: 9.6 vs 5.6 bpm). CONCLUSIONS: Device accuracy varied with activity. Overall, both devices achieved a mean absolute percentage error of just <10%. Thus, they were considered to produce valid results based on the limits established by previous work in the field. Neither device reached sufficient accuracy during seated rest or keyboard typing. Thus, both devices may be eligible for use in respective studies; however, researchers should consider their individual study requirements. JMIR Publications 2022-03-01 /pmc/articles/PMC8924780/ /pubmed/35230250 http://dx.doi.org/10.2196/33635 Text en ©Michael Nissen, Syrine Slim, Katharina Jäger, Madeleine Flaucher, Hanna Huebner, Nina Danzberger, Peter A Fasching, Matthias W Beckmann, Stefan Gradl, Bjoern M Eskofier. Originally published in JMIR Formative Research (https://formative.jmir.org), 01.03.2022. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Formative Research, is properly cited. The complete bibliographic information, a link to the original publication on https://formative.jmir.org, as well as this copyright and license information must be included.
spellingShingle Original Paper
Nissen, Michael
Slim, Syrine
Jäger, Katharina
Flaucher, Madeleine
Huebner, Hanna
Danzberger, Nina
Fasching, Peter A
Beckmann, Matthias W
Gradl, Stefan
Eskofier, Bjoern M
Heart Rate Measurement Accuracy of Fitbit Charge 4 and Samsung Galaxy Watch Active2: Device Evaluation Study
title Heart Rate Measurement Accuracy of Fitbit Charge 4 and Samsung Galaxy Watch Active2: Device Evaluation Study
title_full Heart Rate Measurement Accuracy of Fitbit Charge 4 and Samsung Galaxy Watch Active2: Device Evaluation Study
title_fullStr Heart Rate Measurement Accuracy of Fitbit Charge 4 and Samsung Galaxy Watch Active2: Device Evaluation Study
title_full_unstemmed Heart Rate Measurement Accuracy of Fitbit Charge 4 and Samsung Galaxy Watch Active2: Device Evaluation Study
title_short Heart Rate Measurement Accuracy of Fitbit Charge 4 and Samsung Galaxy Watch Active2: Device Evaluation Study
title_sort heart rate measurement accuracy of fitbit charge 4 and samsung galaxy watch active2: device evaluation study
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8924780/
https://www.ncbi.nlm.nih.gov/pubmed/35230250
http://dx.doi.org/10.2196/33635
work_keys_str_mv AT nissenmichael heartratemeasurementaccuracyoffitbitcharge4andsamsunggalaxywatchactive2deviceevaluationstudy
AT slimsyrine heartratemeasurementaccuracyoffitbitcharge4andsamsunggalaxywatchactive2deviceevaluationstudy
AT jagerkatharina heartratemeasurementaccuracyoffitbitcharge4andsamsunggalaxywatchactive2deviceevaluationstudy
AT flauchermadeleine heartratemeasurementaccuracyoffitbitcharge4andsamsunggalaxywatchactive2deviceevaluationstudy
AT huebnerhanna heartratemeasurementaccuracyoffitbitcharge4andsamsunggalaxywatchactive2deviceevaluationstudy
AT danzbergernina heartratemeasurementaccuracyoffitbitcharge4andsamsunggalaxywatchactive2deviceevaluationstudy
AT faschingpetera heartratemeasurementaccuracyoffitbitcharge4andsamsunggalaxywatchactive2deviceevaluationstudy
AT beckmannmatthiasw heartratemeasurementaccuracyoffitbitcharge4andsamsunggalaxywatchactive2deviceevaluationstudy
AT gradlstefan heartratemeasurementaccuracyoffitbitcharge4andsamsunggalaxywatchactive2deviceevaluationstudy
AT eskofierbjoernm heartratemeasurementaccuracyoffitbitcharge4andsamsunggalaxywatchactive2deviceevaluationstudy