Evaluation of ocular biometry in primary angle-closure disease with two swept source optical coherence tomography devices

PURPOSE: To investigate agreement between 2 swept source OCT biometers, IOL Master700 and Anterion, in various ocular biometry and intraocular lens (IOL) calculations of primary angle-closure disease (PACD). SETTING: Rajavithi Hospital, Bangkok, Thailand. DESIGN: Prospective comparative study. METHO...

Descripción completa

Detalles Bibliográficos
Autores principales: Wanichwecharungruang, Boonsong, Amornpetchsathaporn, Anyarak, Wongwijitsook, Wisakorn, Kongsomboon, Kittipong, Chantra, Somporn
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8936461/
https://www.ncbi.nlm.nih.gov/pubmed/35312733
http://dx.doi.org/10.1371/journal.pone.0265844
Descripción
Sumario:PURPOSE: To investigate agreement between 2 swept source OCT biometers, IOL Master700 and Anterion, in various ocular biometry and intraocular lens (IOL) calculations of primary angle-closure disease (PACD). SETTING: Rajavithi Hospital, Bangkok, Thailand. DESIGN: Prospective comparative study. METHODS: This study conducted in a tertiary eye care center involving biometric measurements obtained with 2 devices in phakic eye with diagnosis of PACD. Mean difference and intraclass correlation coefficient (ICC) with confidence limits were assessed, and calculations of estimated residual refraction of the IOL were analysed using Barrett’s formula. RESULTS: Sixty-nine eyes from 45 PACD patients were enrolled for the study. Excellent agreement of various parameters was revealed, with ICC (confidence limits) of K1 = 0.953 (0.861–0.979), K2 = 0.950 (0.778–0.98), ACD = 0.932 (0.529–0.978), WTW = 0.775 (0.477–0.888), and LT = 0.947 (0.905–0.97). Mean difference of axial length (AL) was -0.01 ± 0.02 mm with ICC of 1.000. IOL calculation was assessed with Barrett’s formula, and Bland-Altman plot showed excellent agreement in the results of the 2 devices for the IOL power and estimated post-operative residual refraction (EPR). CONCLUSIONS: Mean differences of biometric parameters, obtained with IOL Master700 and Anterion, were small, and ICC showed excellent concordance. No clinical relevance in calculation of IOL power was found, and the two devices appeared to be comparably effective in clinical practice.