Impact of a PMMA tube on performances of a Vereos PET/CT system adapted for BSL-3 environment according to the NEMA NU2-2012 standard
INTRODUCTION: A Vereos PET/CT device was adapted to be compatible with the experimentation in large animals within BSL-3 environment. The aim of this study was to investigate the impact of this modification on the performance according to NEMA NU2-2012 standard. METHODS: Spatial resolution, sensitiv...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8938744/ https://www.ncbi.nlm.nih.gov/pubmed/35316846 http://dx.doi.org/10.1186/s40658-022-00450-6 |
_version_ | 1784672611521789952 |
---|---|
author | Kahlaoui, Nidhal Naninck, Thibaut Le Grand, Roger Chapon, Catherine |
author_facet | Kahlaoui, Nidhal Naninck, Thibaut Le Grand, Roger Chapon, Catherine |
author_sort | Kahlaoui, Nidhal |
collection | PubMed |
description | INTRODUCTION: A Vereos PET/CT device was adapted to be compatible with the experimentation in large animals within BSL-3 environment. The aim of this study was to investigate the impact of this modification on the performance according to NEMA NU2-2012 standard. METHODS: Spatial resolution, sensitivity, count rate performance, accuracies of corrections and image quality were assessed using the NEMA NU2-2012 standards before and after installation of a transparent poly-methyl methacrylate tube of 8 mm thickness, 680 mm diameter and 2800 mm long inside the tunnel of the system. In addition, CT performance tests were performed according to manufacturer standard procedure. RESULTS: Although the presence of the tube led to a slight decrease in sensitivity, performance measurements were in accordance with manufacturer preconisation ranges and comparable to previous performance published data. CONCLUSION: Modifications of Vereos PET/CT system allowing its use in BSL-3 conditions did not affect significantly its performance according to NEMA NU2-2012 standard. KEY POINTS: Question. Does a BSL-3 compatible modification alter Philips Vereos PET/CT performances according to NEMA NU2-2012 standards? Pertinent findings. Our Vereos PET/CT system was modified by a wall separating BSL-1 and BSL-3 sides and an 8 mm thickness PMMA tube inserted into the bore of the camera in order to extend the BSL-3 containment along the bed movement. The performances of our modified system according to NEMA NU2-2012 standards were not significantly impacted by the modifications and were in accordance with the values prescribed by the manufacturer. Implications for patients care. Our clinical PET/CT device was modified for human infectious diseases studies in Non-Human Primates. This unusual set up may then provide truly transposable data from preclinical studies into clinical application in infected patients. |
format | Online Article Text |
id | pubmed-8938744 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Springer International Publishing |
record_format | MEDLINE/PubMed |
spelling | pubmed-89387442022-03-22 Impact of a PMMA tube on performances of a Vereos PET/CT system adapted for BSL-3 environment according to the NEMA NU2-2012 standard Kahlaoui, Nidhal Naninck, Thibaut Le Grand, Roger Chapon, Catherine EJNMMI Phys Original Research INTRODUCTION: A Vereos PET/CT device was adapted to be compatible with the experimentation in large animals within BSL-3 environment. The aim of this study was to investigate the impact of this modification on the performance according to NEMA NU2-2012 standard. METHODS: Spatial resolution, sensitivity, count rate performance, accuracies of corrections and image quality were assessed using the NEMA NU2-2012 standards before and after installation of a transparent poly-methyl methacrylate tube of 8 mm thickness, 680 mm diameter and 2800 mm long inside the tunnel of the system. In addition, CT performance tests were performed according to manufacturer standard procedure. RESULTS: Although the presence of the tube led to a slight decrease in sensitivity, performance measurements were in accordance with manufacturer preconisation ranges and comparable to previous performance published data. CONCLUSION: Modifications of Vereos PET/CT system allowing its use in BSL-3 conditions did not affect significantly its performance according to NEMA NU2-2012 standard. KEY POINTS: Question. Does a BSL-3 compatible modification alter Philips Vereos PET/CT performances according to NEMA NU2-2012 standards? Pertinent findings. Our Vereos PET/CT system was modified by a wall separating BSL-1 and BSL-3 sides and an 8 mm thickness PMMA tube inserted into the bore of the camera in order to extend the BSL-3 containment along the bed movement. The performances of our modified system according to NEMA NU2-2012 standards were not significantly impacted by the modifications and were in accordance with the values prescribed by the manufacturer. Implications for patients care. Our clinical PET/CT device was modified for human infectious diseases studies in Non-Human Primates. This unusual set up may then provide truly transposable data from preclinical studies into clinical application in infected patients. Springer International Publishing 2022-03-22 /pmc/articles/PMC8938744/ /pubmed/35316846 http://dx.doi.org/10.1186/s40658-022-00450-6 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Original Research Kahlaoui, Nidhal Naninck, Thibaut Le Grand, Roger Chapon, Catherine Impact of a PMMA tube on performances of a Vereos PET/CT system adapted for BSL-3 environment according to the NEMA NU2-2012 standard |
title | Impact of a PMMA tube on performances of a Vereos PET/CT system adapted for BSL-3 environment according to the NEMA NU2-2012 standard |
title_full | Impact of a PMMA tube on performances of a Vereos PET/CT system adapted for BSL-3 environment according to the NEMA NU2-2012 standard |
title_fullStr | Impact of a PMMA tube on performances of a Vereos PET/CT system adapted for BSL-3 environment according to the NEMA NU2-2012 standard |
title_full_unstemmed | Impact of a PMMA tube on performances of a Vereos PET/CT system adapted for BSL-3 environment according to the NEMA NU2-2012 standard |
title_short | Impact of a PMMA tube on performances of a Vereos PET/CT system adapted for BSL-3 environment according to the NEMA NU2-2012 standard |
title_sort | impact of a pmma tube on performances of a vereos pet/ct system adapted for bsl-3 environment according to the nema nu2-2012 standard |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8938744/ https://www.ncbi.nlm.nih.gov/pubmed/35316846 http://dx.doi.org/10.1186/s40658-022-00450-6 |
work_keys_str_mv | AT kahlaouinidhal impactofapmmatubeonperformancesofavereospetctsystemadaptedforbsl3environmentaccordingtothenemanu22012standard AT naninckthibaut impactofapmmatubeonperformancesofavereospetctsystemadaptedforbsl3environmentaccordingtothenemanu22012standard AT legrandroger impactofapmmatubeonperformancesofavereospetctsystemadaptedforbsl3environmentaccordingtothenemanu22012standard AT chaponcatherine impactofapmmatubeonperformancesofavereospetctsystemadaptedforbsl3environmentaccordingtothenemanu22012standard |