Device for retrieval of vena cava filter with combination of a multi-loop snare and Amplatz catheter
A 25-year-old woman diagnosed as having a non-massive pulmonary embolism and deep vein thrombosis from the right superficial femoral to the right common iliac vein was treated by deployment of a DENALI(®) Vena Cava Filter. Filter retrieval was attempted 6 months later using a BARD snare retrieval ki...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8943558/ https://www.ncbi.nlm.nih.gov/pubmed/35341096 http://dx.doi.org/10.1177/2050313X221086102 |
_version_ | 1784673544602386432 |
---|---|
author | Matsushita, Masato Takano, Masamichi Miyauchi, Yasushi |
author_facet | Matsushita, Masato Takano, Masamichi Miyauchi, Yasushi |
author_sort | Matsushita, Masato |
collection | PubMed |
description | A 25-year-old woman diagnosed as having a non-massive pulmonary embolism and deep vein thrombosis from the right superficial femoral to the right common iliac vein was treated by deployment of a DENALI(®) Vena Cava Filter. Filter retrieval was attempted 6 months later using a BARD snare retrieval kit. However, the conventional technique was unsuccessful because of a tilt and attachment of the filter head to the vessel wall. Hence, we passed an EN Snare Endovascular Snare System through an Amplatz Left 1 guiding catheter to successfully penetrate the attached portion of the filter hook and vessel wall. This combination provided controlled direction and a strong backup force that helped capture and retrieve the filter. This technique could be an alternative method to retrieve inferior vena cava filter with severe tilt and tight attachment to the vena cava wall when the conventional technique is unsuccessful. |
format | Online Article Text |
id | pubmed-8943558 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-89435582022-03-25 Device for retrieval of vena cava filter with combination of a multi-loop snare and Amplatz catheter Matsushita, Masato Takano, Masamichi Miyauchi, Yasushi SAGE Open Med Case Rep Case Report A 25-year-old woman diagnosed as having a non-massive pulmonary embolism and deep vein thrombosis from the right superficial femoral to the right common iliac vein was treated by deployment of a DENALI(®) Vena Cava Filter. Filter retrieval was attempted 6 months later using a BARD snare retrieval kit. However, the conventional technique was unsuccessful because of a tilt and attachment of the filter head to the vessel wall. Hence, we passed an EN Snare Endovascular Snare System through an Amplatz Left 1 guiding catheter to successfully penetrate the attached portion of the filter hook and vessel wall. This combination provided controlled direction and a strong backup force that helped capture and retrieve the filter. This technique could be an alternative method to retrieve inferior vena cava filter with severe tilt and tight attachment to the vena cava wall when the conventional technique is unsuccessful. SAGE Publications 2022-03-22 /pmc/articles/PMC8943558/ /pubmed/35341096 http://dx.doi.org/10.1177/2050313X221086102 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by-nc/4.0/This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access page (https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Case Report Matsushita, Masato Takano, Masamichi Miyauchi, Yasushi Device for retrieval of vena cava filter with combination of a multi-loop snare and Amplatz catheter |
title | Device for retrieval of vena cava filter with combination of a multi-loop
snare and Amplatz catheter |
title_full | Device for retrieval of vena cava filter with combination of a multi-loop
snare and Amplatz catheter |
title_fullStr | Device for retrieval of vena cava filter with combination of a multi-loop
snare and Amplatz catheter |
title_full_unstemmed | Device for retrieval of vena cava filter with combination of a multi-loop
snare and Amplatz catheter |
title_short | Device for retrieval of vena cava filter with combination of a multi-loop
snare and Amplatz catheter |
title_sort | device for retrieval of vena cava filter with combination of a multi-loop
snare and amplatz catheter |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8943558/ https://www.ncbi.nlm.nih.gov/pubmed/35341096 http://dx.doi.org/10.1177/2050313X221086102 |
work_keys_str_mv | AT matsushitamasato deviceforretrievalofvenacavafilterwithcombinationofamultiloopsnareandamplatzcatheter AT takanomasamichi deviceforretrievalofvenacavafilterwithcombinationofamultiloopsnareandamplatzcatheter AT miyauchiyasushi deviceforretrievalofvenacavafilterwithcombinationofamultiloopsnareandamplatzcatheter |