Quantitative Blood Flow Measurements in the Common Carotid Artery: A Comparative Study of High-Frame-Rate Ultrasound Vector Flow Imaging, Pulsed Wave Doppler, and Phase Contrast Magnetic Resonance Imaging

V Flow is commercially developed by high-frame-rate ultrasound vector flow imaging. Compared to conventional color Doppler, V Flow is angle-independent and is capable of measuring both the magnitude and the direction of blood flow velocities. This paper aims to investigate the differences between V...

Descripción completa

Detalles Bibliográficos
Autores principales: Du, Yigang, Ding, Haiyan, He, Le, Yiu, Billy Y. S., Deng, Linsong, Yu, Alfred C. H., Zhu, Lei
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8947594/
https://www.ncbi.nlm.nih.gov/pubmed/35328242
http://dx.doi.org/10.3390/diagnostics12030690
Descripción
Sumario:V Flow is commercially developed by high-frame-rate ultrasound vector flow imaging. Compared to conventional color Doppler, V Flow is angle-independent and is capable of measuring both the magnitude and the direction of blood flow velocities. This paper aims to investigate the differences between V Flow and pulsed wave Doppler (PW) relative to phase contrast magnetic resonance imaging (PC-MRI), for the quantitative measurements of blood flow in common carotid arteries (CCA) and, consequently, to evaluate the accuracy of the new technique, V Flow. Sixty-four CCAs were measured using V Flow, PW, and PC-MRI. The maximum velocities, time-averaged mean (TAMEAN) velocities, and volume flow were measured using different imaging technologies. The mean error with standard deviation (Std), the median of absolute errors, and the r-values between V Flow and PC-MRI results for the maximum velocity, the TAMEAN velocity, and the volume flow measurements are {9.40% ± 14.91%; 11.84%; 0.84}, {21.52% ± 14.46%; 19.28%; 0.86}, and {−2.80% ± 14.01%; 10.38%; 0.7}, respectively, and are {53.44% ± 29.68%; 49.79%; 0.74}, {27.83% ± 31.60%; 23.83; 0.71}, and {21.01% ± 29.64%; 25.48%; 0.34}, respectively, for those between PW and PC-MRI. The boxplot, linear regression and Bland–Altman plots were performed for each comparison, which illustrated that the results measured via V Flow rather than via PW agreed more closely with those measured via PC-MRI.