Quantitative Blood Flow Measurements in the Common Carotid Artery: A Comparative Study of High-Frame-Rate Ultrasound Vector Flow Imaging, Pulsed Wave Doppler, and Phase Contrast Magnetic Resonance Imaging
V Flow is commercially developed by high-frame-rate ultrasound vector flow imaging. Compared to conventional color Doppler, V Flow is angle-independent and is capable of measuring both the magnitude and the direction of blood flow velocities. This paper aims to investigate the differences between V...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8947594/ https://www.ncbi.nlm.nih.gov/pubmed/35328242 http://dx.doi.org/10.3390/diagnostics12030690 |
_version_ | 1784674477204832256 |
---|---|
author | Du, Yigang Ding, Haiyan He, Le Yiu, Billy Y. S. Deng, Linsong Yu, Alfred C. H. Zhu, Lei |
author_facet | Du, Yigang Ding, Haiyan He, Le Yiu, Billy Y. S. Deng, Linsong Yu, Alfred C. H. Zhu, Lei |
author_sort | Du, Yigang |
collection | PubMed |
description | V Flow is commercially developed by high-frame-rate ultrasound vector flow imaging. Compared to conventional color Doppler, V Flow is angle-independent and is capable of measuring both the magnitude and the direction of blood flow velocities. This paper aims to investigate the differences between V Flow and pulsed wave Doppler (PW) relative to phase contrast magnetic resonance imaging (PC-MRI), for the quantitative measurements of blood flow in common carotid arteries (CCA) and, consequently, to evaluate the accuracy of the new technique, V Flow. Sixty-four CCAs were measured using V Flow, PW, and PC-MRI. The maximum velocities, time-averaged mean (TAMEAN) velocities, and volume flow were measured using different imaging technologies. The mean error with standard deviation (Std), the median of absolute errors, and the r-values between V Flow and PC-MRI results for the maximum velocity, the TAMEAN velocity, and the volume flow measurements are {9.40% ± 14.91%; 11.84%; 0.84}, {21.52% ± 14.46%; 19.28%; 0.86}, and {−2.80% ± 14.01%; 10.38%; 0.7}, respectively, and are {53.44% ± 29.68%; 49.79%; 0.74}, {27.83% ± 31.60%; 23.83; 0.71}, and {21.01% ± 29.64%; 25.48%; 0.34}, respectively, for those between PW and PC-MRI. The boxplot, linear regression and Bland–Altman plots were performed for each comparison, which illustrated that the results measured via V Flow rather than via PW agreed more closely with those measured via PC-MRI. |
format | Online Article Text |
id | pubmed-8947594 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-89475942022-03-25 Quantitative Blood Flow Measurements in the Common Carotid Artery: A Comparative Study of High-Frame-Rate Ultrasound Vector Flow Imaging, Pulsed Wave Doppler, and Phase Contrast Magnetic Resonance Imaging Du, Yigang Ding, Haiyan He, Le Yiu, Billy Y. S. Deng, Linsong Yu, Alfred C. H. Zhu, Lei Diagnostics (Basel) Article V Flow is commercially developed by high-frame-rate ultrasound vector flow imaging. Compared to conventional color Doppler, V Flow is angle-independent and is capable of measuring both the magnitude and the direction of blood flow velocities. This paper aims to investigate the differences between V Flow and pulsed wave Doppler (PW) relative to phase contrast magnetic resonance imaging (PC-MRI), for the quantitative measurements of blood flow in common carotid arteries (CCA) and, consequently, to evaluate the accuracy of the new technique, V Flow. Sixty-four CCAs were measured using V Flow, PW, and PC-MRI. The maximum velocities, time-averaged mean (TAMEAN) velocities, and volume flow were measured using different imaging technologies. The mean error with standard deviation (Std), the median of absolute errors, and the r-values between V Flow and PC-MRI results for the maximum velocity, the TAMEAN velocity, and the volume flow measurements are {9.40% ± 14.91%; 11.84%; 0.84}, {21.52% ± 14.46%; 19.28%; 0.86}, and {−2.80% ± 14.01%; 10.38%; 0.7}, respectively, and are {53.44% ± 29.68%; 49.79%; 0.74}, {27.83% ± 31.60%; 23.83; 0.71}, and {21.01% ± 29.64%; 25.48%; 0.34}, respectively, for those between PW and PC-MRI. The boxplot, linear regression and Bland–Altman plots were performed for each comparison, which illustrated that the results measured via V Flow rather than via PW agreed more closely with those measured via PC-MRI. MDPI 2022-03-11 /pmc/articles/PMC8947594/ /pubmed/35328242 http://dx.doi.org/10.3390/diagnostics12030690 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Du, Yigang Ding, Haiyan He, Le Yiu, Billy Y. S. Deng, Linsong Yu, Alfred C. H. Zhu, Lei Quantitative Blood Flow Measurements in the Common Carotid Artery: A Comparative Study of High-Frame-Rate Ultrasound Vector Flow Imaging, Pulsed Wave Doppler, and Phase Contrast Magnetic Resonance Imaging |
title | Quantitative Blood Flow Measurements in the Common Carotid Artery: A Comparative Study of High-Frame-Rate Ultrasound Vector Flow Imaging, Pulsed Wave Doppler, and Phase Contrast Magnetic Resonance Imaging |
title_full | Quantitative Blood Flow Measurements in the Common Carotid Artery: A Comparative Study of High-Frame-Rate Ultrasound Vector Flow Imaging, Pulsed Wave Doppler, and Phase Contrast Magnetic Resonance Imaging |
title_fullStr | Quantitative Blood Flow Measurements in the Common Carotid Artery: A Comparative Study of High-Frame-Rate Ultrasound Vector Flow Imaging, Pulsed Wave Doppler, and Phase Contrast Magnetic Resonance Imaging |
title_full_unstemmed | Quantitative Blood Flow Measurements in the Common Carotid Artery: A Comparative Study of High-Frame-Rate Ultrasound Vector Flow Imaging, Pulsed Wave Doppler, and Phase Contrast Magnetic Resonance Imaging |
title_short | Quantitative Blood Flow Measurements in the Common Carotid Artery: A Comparative Study of High-Frame-Rate Ultrasound Vector Flow Imaging, Pulsed Wave Doppler, and Phase Contrast Magnetic Resonance Imaging |
title_sort | quantitative blood flow measurements in the common carotid artery: a comparative study of high-frame-rate ultrasound vector flow imaging, pulsed wave doppler, and phase contrast magnetic resonance imaging |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8947594/ https://www.ncbi.nlm.nih.gov/pubmed/35328242 http://dx.doi.org/10.3390/diagnostics12030690 |
work_keys_str_mv | AT duyigang quantitativebloodflowmeasurementsinthecommoncarotidarteryacomparativestudyofhighframerateultrasoundvectorflowimagingpulsedwavedopplerandphasecontrastmagneticresonanceimaging AT dinghaiyan quantitativebloodflowmeasurementsinthecommoncarotidarteryacomparativestudyofhighframerateultrasoundvectorflowimagingpulsedwavedopplerandphasecontrastmagneticresonanceimaging AT hele quantitativebloodflowmeasurementsinthecommoncarotidarteryacomparativestudyofhighframerateultrasoundvectorflowimagingpulsedwavedopplerandphasecontrastmagneticresonanceimaging AT yiubillyys quantitativebloodflowmeasurementsinthecommoncarotidarteryacomparativestudyofhighframerateultrasoundvectorflowimagingpulsedwavedopplerandphasecontrastmagneticresonanceimaging AT denglinsong quantitativebloodflowmeasurementsinthecommoncarotidarteryacomparativestudyofhighframerateultrasoundvectorflowimagingpulsedwavedopplerandphasecontrastmagneticresonanceimaging AT yualfredch quantitativebloodflowmeasurementsinthecommoncarotidarteryacomparativestudyofhighframerateultrasoundvectorflowimagingpulsedwavedopplerandphasecontrastmagneticresonanceimaging AT zhulei quantitativebloodflowmeasurementsinthecommoncarotidarteryacomparativestudyofhighframerateultrasoundvectorflowimagingpulsedwavedopplerandphasecontrastmagneticresonanceimaging |