An Alternative Strategy for Screening and Confirmation of 330 Pesticides in Ground- and Surface Water Using Liquid Chromatography Tandem Mass Spectrometry
The presence of pesticide residues in water is a huge worldwide concern. In this paper we described the development and validation of a new liquid chromatography tandem mass spectrometric (LC-MS/MS) method for both screening and quantification of pesticides in water samples. In the sample preparatio...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8950376/ https://www.ncbi.nlm.nih.gov/pubmed/35335236 http://dx.doi.org/10.3390/molecules27061872 |
Sumario: | The presence of pesticide residues in water is a huge worldwide concern. In this paper we described the development and validation of a new liquid chromatography tandem mass spectrometric (LC-MS/MS) method for both screening and quantification of pesticides in water samples. In the sample preparation stage, the samples were buffered to pH 7.0 and pre-concentrated on polymeric-based cartridges via solid-phase extraction (SPE). Highly sensitive detection was carried out with mobile phases containing only 5 mM ammonium formate (pH of 6.8) as an eluent additive and using only positive ionization mode in MS/MS instrument. Hence, only 200-fold sample enrichment was required to set a screening detection limit (SDL) and reporting limit (RL) of 10 ng/L. The confirmatory method was validated at 10 and 100 ng/L spiking levels. The apparent recoveries obtained from the matrix-matched calibration (5–500 ng/L) were within the acceptable range (60–120%), also the precision (relative standard deviation, RSD) was not higher than 20%. During the development, 480 pesticides were tested and 330 compounds fulfilled the requirements of validation. The method was successfully applied to proficiency test samples to evaluate its accuracy. Moreover, the method robustness test was carried out using higher sample volume (500 mL) followed by automated SPE enrichment. Finally, the method was used to analyze 20 real samples, in which some compounds were detected around 10 ng/L, but never exceeded the assay maximum level. |
---|