Cargando…

The BIOMONITOR III Injectable Cardiac Monitor: Clinical Experience with a Novel Injectable Cardiac Monitor

Background: Injectable cardiac monitors (ICMs) are leadless subcutaneous devices for long-term monitoring of arrhythmias. The BIOTRONIK BIOMONITOR III is a novel ICM with a miniaturized profile, long sensing vector, and simplified implantation technique. Methods: R-wave amplitude was recorded immedi...

Descripción completa

Detalles Bibliográficos
Autores principales: Reinsch, Nico, Füting, Anna, Höwel, Dennis, Neven, Kars
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8954265/
https://www.ncbi.nlm.nih.gov/pubmed/35329960
http://dx.doi.org/10.3390/jcm11061634
_version_ 1784676052068466688
author Reinsch, Nico
Füting, Anna
Höwel, Dennis
Neven, Kars
author_facet Reinsch, Nico
Füting, Anna
Höwel, Dennis
Neven, Kars
author_sort Reinsch, Nico
collection PubMed
description Background: Injectable cardiac monitors (ICMs) are leadless subcutaneous devices for long-term monitoring of arrhythmias. The BIOTRONIK BIOMONITOR III is a novel ICM with a miniaturized profile, long sensing vector, and simplified implantation technique. Methods: R-wave amplitude was recorded immediately after implantation, the day after implantation, and after 3 months. Follow-up was scheduled after 3 months or after an event. All data from the ICM were retrieved. The anatomical position of the ICM was determined post-implantation and after 3 months. A patient questionnaire was conducted after 3 months. Results: In 36 patients (mean age 67 ± 13 years; 40% male) an ICM was inserted. Six patients were not included in the final analysis. The median time from skin cut to wound closure was 6 [IQR 5–7] minutes. Mean R-wave amplitude increased over time (0.73 ± 32 mV vs. 0.78 ± 0.38 mV vs. 0.81 ± 0.39 mV; p = ns). Three months after implantation, the ICM was in an anatomically stable position. In 14 (47%) patients, true episodes were detected. False arrhythmia alerts were detected in 13 (43%) patients. The total number of false detections was low, and the patient satisfaction rate was high. Conclusion: Implantation of the novel BIOMONITOR III is fast and uncomplicated; its sensing characteristics are excellent and improve over time, and patient satisfaction is high.
format Online
Article
Text
id pubmed-8954265
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-89542652022-03-26 The BIOMONITOR III Injectable Cardiac Monitor: Clinical Experience with a Novel Injectable Cardiac Monitor Reinsch, Nico Füting, Anna Höwel, Dennis Neven, Kars J Clin Med Article Background: Injectable cardiac monitors (ICMs) are leadless subcutaneous devices for long-term monitoring of arrhythmias. The BIOTRONIK BIOMONITOR III is a novel ICM with a miniaturized profile, long sensing vector, and simplified implantation technique. Methods: R-wave amplitude was recorded immediately after implantation, the day after implantation, and after 3 months. Follow-up was scheduled after 3 months or after an event. All data from the ICM were retrieved. The anatomical position of the ICM was determined post-implantation and after 3 months. A patient questionnaire was conducted after 3 months. Results: In 36 patients (mean age 67 ± 13 years; 40% male) an ICM was inserted. Six patients were not included in the final analysis. The median time from skin cut to wound closure was 6 [IQR 5–7] minutes. Mean R-wave amplitude increased over time (0.73 ± 32 mV vs. 0.78 ± 0.38 mV vs. 0.81 ± 0.39 mV; p = ns). Three months after implantation, the ICM was in an anatomically stable position. In 14 (47%) patients, true episodes were detected. False arrhythmia alerts were detected in 13 (43%) patients. The total number of false detections was low, and the patient satisfaction rate was high. Conclusion: Implantation of the novel BIOMONITOR III is fast and uncomplicated; its sensing characteristics are excellent and improve over time, and patient satisfaction is high. MDPI 2022-03-16 /pmc/articles/PMC8954265/ /pubmed/35329960 http://dx.doi.org/10.3390/jcm11061634 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Reinsch, Nico
Füting, Anna
Höwel, Dennis
Neven, Kars
The BIOMONITOR III Injectable Cardiac Monitor: Clinical Experience with a Novel Injectable Cardiac Monitor
title The BIOMONITOR III Injectable Cardiac Monitor: Clinical Experience with a Novel Injectable Cardiac Monitor
title_full The BIOMONITOR III Injectable Cardiac Monitor: Clinical Experience with a Novel Injectable Cardiac Monitor
title_fullStr The BIOMONITOR III Injectable Cardiac Monitor: Clinical Experience with a Novel Injectable Cardiac Monitor
title_full_unstemmed The BIOMONITOR III Injectable Cardiac Monitor: Clinical Experience with a Novel Injectable Cardiac Monitor
title_short The BIOMONITOR III Injectable Cardiac Monitor: Clinical Experience with a Novel Injectable Cardiac Monitor
title_sort biomonitor iii injectable cardiac monitor: clinical experience with a novel injectable cardiac monitor
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8954265/
https://www.ncbi.nlm.nih.gov/pubmed/35329960
http://dx.doi.org/10.3390/jcm11061634
work_keys_str_mv AT reinschnico thebiomonitoriiiinjectablecardiacmonitorclinicalexperiencewithanovelinjectablecardiacmonitor
AT futinganna thebiomonitoriiiinjectablecardiacmonitorclinicalexperiencewithanovelinjectablecardiacmonitor
AT howeldennis thebiomonitoriiiinjectablecardiacmonitorclinicalexperiencewithanovelinjectablecardiacmonitor
AT nevenkars thebiomonitoriiiinjectablecardiacmonitorclinicalexperiencewithanovelinjectablecardiacmonitor
AT reinschnico biomonitoriiiinjectablecardiacmonitorclinicalexperiencewithanovelinjectablecardiacmonitor
AT futinganna biomonitoriiiinjectablecardiacmonitorclinicalexperiencewithanovelinjectablecardiacmonitor
AT howeldennis biomonitoriiiinjectablecardiacmonitorclinicalexperiencewithanovelinjectablecardiacmonitor
AT nevenkars biomonitoriiiinjectablecardiacmonitorclinicalexperiencewithanovelinjectablecardiacmonitor