Polyethylene liner motion in dual-mobility hip prostheses: static and dynamic radiostereometry in 16 patients 1 year after operation

BACKGROUND AND PURPOSE: Dual-mobility hip arthroplasty utilizes a freely rotating polyethylene acetabular liner to protect against dislocation. As liner motion has not been confirmed in vivo, we undertook this using dynamic radiostereometry (RSA). PATIENTS AND METHODS: 6 patients with Anatomical Dua...

Descripción completa

Detalles Bibliográficos
Autores principales: JØRGENSEN, Peter Bo, KAPTEIN, Bart L, SØBALLE, Kjeld, JAKOBSEN, Stig S, STILLING, Maiken
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medical Journals Sweden, on behalf of the Nordic Orthopedic Federation 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8958851/
https://www.ncbi.nlm.nih.gov/pubmed/35347340
http://dx.doi.org/10.2340/17453674.2022.2253
_version_ 1784677033584885760
author JØRGENSEN, Peter Bo
KAPTEIN, Bart L
SØBALLE, Kjeld
JAKOBSEN, Stig S
STILLING, Maiken
author_facet JØRGENSEN, Peter Bo
KAPTEIN, Bart L
SØBALLE, Kjeld
JAKOBSEN, Stig S
STILLING, Maiken
author_sort JØRGENSEN, Peter Bo
collection PubMed
description BACKGROUND AND PURPOSE: Dual-mobility hip arthroplasty utilizes a freely rotating polyethylene acetabular liner to protect against dislocation. As liner motion has not been confirmed in vivo, we undertook this using dynamic radiostereometry (RSA). PATIENTS AND METHODS: 6 patients with Anatomical Dual Mobility acetabular components were included. Markers were implanted in the liners using a drill guide. Static RSA recordings and patient-reported outcome measures were obtained postoperatively and at 1-year follow-up. Dynamic RSA recordings were obtained at 1-year follow-up during passive hip movement: abduction/external rotation, adduction/internal rotation (modified FABER–FADIR), to end-range and at 45° hip flexion. Liner and neck movements were described as anteversion, inclination, and rotation. RESULTS: Liner movement during modified FABER–FADIR was detected in 12 of 16 patients. Median (range) absolute liner movements were: anteversion 10° (5–20), inclination 6° (2–12), and rotation 11° (5–48) relative to the cup. Median absolute change in the resulting liner/neck angle (small articulation) was 28° (12–46) and in liner/cup angle (larger articulation) was 6° (4–21). Static RSA showed changes in median liner anteversion from 7° (–12 to 23) postoperatively to 10° (–3 to 16) at 1-year follow-up and inclination from 42° (35–66) postoperatively to 59° (46–80) at 1-year follow-up. Liner/neck contact was associated with high initial liner anteversion (p = 0.01). INTERPRETATION: The polyethylene liner moves over time. 1 year after surgery the liner can move with or without liner/neck contact. The majority of movement is in the smaller articulation between head and liner.
format Online
Article
Text
id pubmed-8958851
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Medical Journals Sweden, on behalf of the Nordic Orthopedic Federation
record_format MEDLINE/PubMed
spelling pubmed-89588512022-03-28 Polyethylene liner motion in dual-mobility hip prostheses: static and dynamic radiostereometry in 16 patients 1 year after operation JØRGENSEN, Peter Bo KAPTEIN, Bart L SØBALLE, Kjeld JAKOBSEN, Stig S STILLING, Maiken Acta Orthop Article BACKGROUND AND PURPOSE: Dual-mobility hip arthroplasty utilizes a freely rotating polyethylene acetabular liner to protect against dislocation. As liner motion has not been confirmed in vivo, we undertook this using dynamic radiostereometry (RSA). PATIENTS AND METHODS: 6 patients with Anatomical Dual Mobility acetabular components were included. Markers were implanted in the liners using a drill guide. Static RSA recordings and patient-reported outcome measures were obtained postoperatively and at 1-year follow-up. Dynamic RSA recordings were obtained at 1-year follow-up during passive hip movement: abduction/external rotation, adduction/internal rotation (modified FABER–FADIR), to end-range and at 45° hip flexion. Liner and neck movements were described as anteversion, inclination, and rotation. RESULTS: Liner movement during modified FABER–FADIR was detected in 12 of 16 patients. Median (range) absolute liner movements were: anteversion 10° (5–20), inclination 6° (2–12), and rotation 11° (5–48) relative to the cup. Median absolute change in the resulting liner/neck angle (small articulation) was 28° (12–46) and in liner/cup angle (larger articulation) was 6° (4–21). Static RSA showed changes in median liner anteversion from 7° (–12 to 23) postoperatively to 10° (–3 to 16) at 1-year follow-up and inclination from 42° (35–66) postoperatively to 59° (46–80) at 1-year follow-up. Liner/neck contact was associated with high initial liner anteversion (p = 0.01). INTERPRETATION: The polyethylene liner moves over time. 1 year after surgery the liner can move with or without liner/neck contact. The majority of movement is in the smaller articulation between head and liner. Medical Journals Sweden, on behalf of the Nordic Orthopedic Federation 2022-03-21 /pmc/articles/PMC8958851/ /pubmed/35347340 http://dx.doi.org/10.2340/17453674.2022.2253 Text en © 2022 The Author(s) https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (https://creativecommons.org/licenses/by-nc/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material for non-commercial purposes, provided proper attribution to the original work.
spellingShingle Article
JØRGENSEN, Peter Bo
KAPTEIN, Bart L
SØBALLE, Kjeld
JAKOBSEN, Stig S
STILLING, Maiken
Polyethylene liner motion in dual-mobility hip prostheses: static and dynamic radiostereometry in 16 patients 1 year after operation
title Polyethylene liner motion in dual-mobility hip prostheses: static and dynamic radiostereometry in 16 patients 1 year after operation
title_full Polyethylene liner motion in dual-mobility hip prostheses: static and dynamic radiostereometry in 16 patients 1 year after operation
title_fullStr Polyethylene liner motion in dual-mobility hip prostheses: static and dynamic radiostereometry in 16 patients 1 year after operation
title_full_unstemmed Polyethylene liner motion in dual-mobility hip prostheses: static and dynamic radiostereometry in 16 patients 1 year after operation
title_short Polyethylene liner motion in dual-mobility hip prostheses: static and dynamic radiostereometry in 16 patients 1 year after operation
title_sort polyethylene liner motion in dual-mobility hip prostheses: static and dynamic radiostereometry in 16 patients 1 year after operation
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8958851/
https://www.ncbi.nlm.nih.gov/pubmed/35347340
http://dx.doi.org/10.2340/17453674.2022.2253
work_keys_str_mv AT jørgensenpeterbo polyethylenelinermotionindualmobilityhipprosthesesstaticanddynamicradiostereometryin16patients1yearafteroperation
AT kapteinbartl polyethylenelinermotionindualmobilityhipprosthesesstaticanddynamicradiostereometryin16patients1yearafteroperation
AT søballekjeld polyethylenelinermotionindualmobilityhipprosthesesstaticanddynamicradiostereometryin16patients1yearafteroperation
AT jakobsenstigs polyethylenelinermotionindualmobilityhipprosthesesstaticanddynamicradiostereometryin16patients1yearafteroperation
AT stillingmaiken polyethylenelinermotionindualmobilityhipprosthesesstaticanddynamicradiostereometryin16patients1yearafteroperation