Cargando…
COVID-19 in the Clinic: Human Testing of an Aerosol Containment Mask for Endoscopic Clinic Procedures
OBJECTIVE: To create an aerosol containment mask (ACM) for common otolaryngologic endoscopic procedures that also provides nanoparticle-level protection to patients. STUDY DESIGN: Prospective feasibility study . SETTING: In-person testing with a novel ACM. METHODS: The mask was designed in Solidwork...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8978456/ https://www.ncbi.nlm.nih.gov/pubmed/34311614 http://dx.doi.org/10.1177/01945998211029184 |
Sumario: | OBJECTIVE: To create an aerosol containment mask (ACM) for common otolaryngologic endoscopic procedures that also provides nanoparticle-level protection to patients. STUDY DESIGN: Prospective feasibility study . SETTING: In-person testing with a novel ACM. METHODS: The mask was designed in Solidworks and 3D printed. Measurements were made on 10 healthy volunteers who wore the ACM while reading the Rainbow Passage repeatedly and performing a forced cough or sneeze at 5-second intervals over 1 minute with an endoscope in place. RESULTS: There was a large variation in the number of aerosol particles generated among the volunteers. Only the sneeze task showed a significant increase compared with normal breathing in the 0.3-µm particle size when compared with a 1-tailed t test (P = .013). Both the 0.5-µm and 2.5-µm particle sizes showed significant increases for all tasks, while the 2 largest particle sizes, 5 and 10 µm, showed no significant increase (both P < .01). With the suction off, 3 of 30 events (2 sneeze events and 1 cough event) had increases in particle counts, both inside and outside the mask. With the suction on, 2 of 30 events had an increase in particle counts outside the mask without a corresponding increase in particle counts inside the mask. Therefore, these fluctuations in particle counts were determined to be due to random fluctuation in room particle levels. CONCLUSION: ACM will accommodate rigid and flexible endoscopes plus instruments and may prevent the leakage of patient-generated aerosols, thus avoiding contamination of the room and protecting health care workers from airborne contagions. LEVEL OF EVIDENCE: 2 |
---|