Evaluation of the Speed, Accuracy and Precision of the QuickMIC Rapid Antibiotic Susceptibility Testing Assay With Gram-Negative Bacteria in a Clinical Setting

The rapidly changing landscape of antimicrobial resistance poses a challenge for empirical antibiotic therapy in severely ill patients and highlights the need for fast antibiotic susceptibility diagnostics to guide treatment. Traditional methods for antibiotic susceptibility testing (AST) of bacteri...

Descripción completa

Detalles Bibliográficos
Autores principales: Malmberg, Christer, Torpner, Jessie, Fernberg, Jenny, Öhrn, Håkan, Ångström, Jonas, Johansson, Cecilia, Tängdén, Thomas, Kreuger, Johan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8984463/
https://www.ncbi.nlm.nih.gov/pubmed/35402290
http://dx.doi.org/10.3389/fcimb.2022.758262
_version_ 1784682194179981312
author Malmberg, Christer
Torpner, Jessie
Fernberg, Jenny
Öhrn, Håkan
Ångström, Jonas
Johansson, Cecilia
Tängdén, Thomas
Kreuger, Johan
author_facet Malmberg, Christer
Torpner, Jessie
Fernberg, Jenny
Öhrn, Håkan
Ångström, Jonas
Johansson, Cecilia
Tängdén, Thomas
Kreuger, Johan
author_sort Malmberg, Christer
collection PubMed
description The rapidly changing landscape of antimicrobial resistance poses a challenge for empirical antibiotic therapy in severely ill patients and highlights the need for fast antibiotic susceptibility diagnostics to guide treatment. Traditional methods for antibiotic susceptibility testing (AST) of bacteria such as broth microdilution (BMD) or the disc diffusion method (DDM) are comparatively slow and show high variability. Rapid AST methods under development often trade speed for resolution, sometimes only measuring responses at a single antibiotic concentration. QuickMIC is a recently developed lab-on-a-chip system for rapid AST. Here we evaluate the performance of the QuickMIC method with regard to speed, precision and accuracy in comparison to traditional diagnostic methods. 151 blood cultures of clinical Gram-negative isolates with a high frequency of drug resistance were tested using the QuickMIC system and compared with BMD for 12 antibiotics. To investigate sample turnaround time and method functionality in a clinical setting, another 41 clinical blood culture samples were acquired from the Uppsala University Hospital and analyzed on site in the clinical laboratory with the QuickMIC system, and compared with DDM for 8 antibiotics routinely used in the clinical laboratory. The overall essential agreement between MIC values obtained by QuickMIC and BMD was 83.4%, with an average time to result of 3 h 2 min (SD: 24.8 min) for the QuickMIC method. For the clinical dataset, the categorical agreement between QuickMIC and DDM was 96.8%, whereas essential and categorical agreement against BMD was 91.0% and 96.7%, respectively, and the total turnaround time as compared to routine diagnostics was shown to be reduced by 40% (33 h vs. 55 h). Interexperiment variability was low (average SD: 44.6% from target MIC) compared to the acceptable standard of ±1 log(2) unit (i.e. -50% to +100% deviation from target MIC) in BMD. We conclude that the QuickMIC method can provide rapid and accurate AST, and may be especially valuable in settings with high resistance rates, and for antibiotics where wildtype and antibiotic-resistant bacteria have MIC distributions that are close or overlapping.
format Online
Article
Text
id pubmed-8984463
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-89844632022-04-07 Evaluation of the Speed, Accuracy and Precision of the QuickMIC Rapid Antibiotic Susceptibility Testing Assay With Gram-Negative Bacteria in a Clinical Setting Malmberg, Christer Torpner, Jessie Fernberg, Jenny Öhrn, Håkan Ångström, Jonas Johansson, Cecilia Tängdén, Thomas Kreuger, Johan Front Cell Infect Microbiol Cellular and Infection Microbiology The rapidly changing landscape of antimicrobial resistance poses a challenge for empirical antibiotic therapy in severely ill patients and highlights the need for fast antibiotic susceptibility diagnostics to guide treatment. Traditional methods for antibiotic susceptibility testing (AST) of bacteria such as broth microdilution (BMD) or the disc diffusion method (DDM) are comparatively slow and show high variability. Rapid AST methods under development often trade speed for resolution, sometimes only measuring responses at a single antibiotic concentration. QuickMIC is a recently developed lab-on-a-chip system for rapid AST. Here we evaluate the performance of the QuickMIC method with regard to speed, precision and accuracy in comparison to traditional diagnostic methods. 151 blood cultures of clinical Gram-negative isolates with a high frequency of drug resistance were tested using the QuickMIC system and compared with BMD for 12 antibiotics. To investigate sample turnaround time and method functionality in a clinical setting, another 41 clinical blood culture samples were acquired from the Uppsala University Hospital and analyzed on site in the clinical laboratory with the QuickMIC system, and compared with DDM for 8 antibiotics routinely used in the clinical laboratory. The overall essential agreement between MIC values obtained by QuickMIC and BMD was 83.4%, with an average time to result of 3 h 2 min (SD: 24.8 min) for the QuickMIC method. For the clinical dataset, the categorical agreement between QuickMIC and DDM was 96.8%, whereas essential and categorical agreement against BMD was 91.0% and 96.7%, respectively, and the total turnaround time as compared to routine diagnostics was shown to be reduced by 40% (33 h vs. 55 h). Interexperiment variability was low (average SD: 44.6% from target MIC) compared to the acceptable standard of ±1 log(2) unit (i.e. -50% to +100% deviation from target MIC) in BMD. We conclude that the QuickMIC method can provide rapid and accurate AST, and may be especially valuable in settings with high resistance rates, and for antibiotics where wildtype and antibiotic-resistant bacteria have MIC distributions that are close or overlapping. Frontiers Media S.A. 2022-03-23 /pmc/articles/PMC8984463/ /pubmed/35402290 http://dx.doi.org/10.3389/fcimb.2022.758262 Text en Copyright © 2022 Malmberg, Torpner, Fernberg, Öhrn, Ångström, Johansson, Tängdén and Kreuger https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Cellular and Infection Microbiology
Malmberg, Christer
Torpner, Jessie
Fernberg, Jenny
Öhrn, Håkan
Ångström, Jonas
Johansson, Cecilia
Tängdén, Thomas
Kreuger, Johan
Evaluation of the Speed, Accuracy and Precision of the QuickMIC Rapid Antibiotic Susceptibility Testing Assay With Gram-Negative Bacteria in a Clinical Setting
title Evaluation of the Speed, Accuracy and Precision of the QuickMIC Rapid Antibiotic Susceptibility Testing Assay With Gram-Negative Bacteria in a Clinical Setting
title_full Evaluation of the Speed, Accuracy and Precision of the QuickMIC Rapid Antibiotic Susceptibility Testing Assay With Gram-Negative Bacteria in a Clinical Setting
title_fullStr Evaluation of the Speed, Accuracy and Precision of the QuickMIC Rapid Antibiotic Susceptibility Testing Assay With Gram-Negative Bacteria in a Clinical Setting
title_full_unstemmed Evaluation of the Speed, Accuracy and Precision of the QuickMIC Rapid Antibiotic Susceptibility Testing Assay With Gram-Negative Bacteria in a Clinical Setting
title_short Evaluation of the Speed, Accuracy and Precision of the QuickMIC Rapid Antibiotic Susceptibility Testing Assay With Gram-Negative Bacteria in a Clinical Setting
title_sort evaluation of the speed, accuracy and precision of the quickmic rapid antibiotic susceptibility testing assay with gram-negative bacteria in a clinical setting
topic Cellular and Infection Microbiology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8984463/
https://www.ncbi.nlm.nih.gov/pubmed/35402290
http://dx.doi.org/10.3389/fcimb.2022.758262
work_keys_str_mv AT malmbergchrister evaluationofthespeedaccuracyandprecisionofthequickmicrapidantibioticsusceptibilitytestingassaywithgramnegativebacteriainaclinicalsetting
AT torpnerjessie evaluationofthespeedaccuracyandprecisionofthequickmicrapidantibioticsusceptibilitytestingassaywithgramnegativebacteriainaclinicalsetting
AT fernbergjenny evaluationofthespeedaccuracyandprecisionofthequickmicrapidantibioticsusceptibilitytestingassaywithgramnegativebacteriainaclinicalsetting
AT ohrnhakan evaluationofthespeedaccuracyandprecisionofthequickmicrapidantibioticsusceptibilitytestingassaywithgramnegativebacteriainaclinicalsetting
AT angstromjonas evaluationofthespeedaccuracyandprecisionofthequickmicrapidantibioticsusceptibilitytestingassaywithgramnegativebacteriainaclinicalsetting
AT johanssoncecilia evaluationofthespeedaccuracyandprecisionofthequickmicrapidantibioticsusceptibilitytestingassaywithgramnegativebacteriainaclinicalsetting
AT tangdenthomas evaluationofthespeedaccuracyandprecisionofthequickmicrapidantibioticsusceptibilitytestingassaywithgramnegativebacteriainaclinicalsetting
AT kreugerjohan evaluationofthespeedaccuracyandprecisionofthequickmicrapidantibioticsusceptibilitytestingassaywithgramnegativebacteriainaclinicalsetting