Validation of the A&D UA‐651 Plus/UA‐651SL Plus automated sphygmomanometer according to the ISO 81060–2, 2018 and ISO 81060–2 Amendment 1, 2020, which resulted in the currently pending Amendment to Amendment 1

In 2020 the ANSI/AAMI/ISO published Amendment 1 to the 81060–2 Standard. The purpose was to try to address issues relating to the distribution of subjects with respect to limb size ranges. It also required a new plot showing the limb circumferences on the x‐axis and the device minus reference errors...

Descripción completa

Detalles Bibliográficos
Autor principal: Alpert, Bruce S.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8989751/
https://www.ncbi.nlm.nih.gov/pubmed/35312160
http://dx.doi.org/10.1111/jch.14464
_version_ 1784683238993690624
author Alpert, Bruce S.
author_facet Alpert, Bruce S.
author_sort Alpert, Bruce S.
collection PubMed
description In 2020 the ANSI/AAMI/ISO published Amendment 1 to the 81060–2 Standard. The purpose was to try to address issues relating to the distribution of subjects with respect to limb size ranges. It also required a new plot showing the limb circumferences on the x‐axis and the device minus reference errors on the y‐axis (Alpert plot). The study reported here showed inadequacies of the Amendment if significantly overlapping cuffs were to be tested. The innovative approach led to the testing of 135 subjects. Requirements for blood pressure (BP), arm circumference, and gender were fulfilled. The standard same arm sequential protocol was used. Criterion 1 calculations gave a mean ± standard deviation (SD) device minus manual BP values of 0.22 ± 7.90 mmHg for systolic BP and ‐0.68 ± 7.36 mmHg for diastolic BP. The maximum allowed mean value for either measurement is 5.0 mmHg. The SD values for Criterion 2 were 6.03 mmHg for systolic BP (maximum allowed 6.95) and 6.47 mmHg for diastolic BP (maximum allowed 6.90). All results passed the Standard requirements. This study demonstrated inadequacies of Amendment 1 and led to the development of an Amendment 2, still in the process of finalization. The new Amendment corrects the issues with significantly overlapping cuffs and “wide‐range” cuffs. The A&D UA‐651 Plus/UA‐651SL Plus BP monitors and the five cuffs not only passed the requirements of 81060–2:2018 and Amendment 1 but also a revised and more rigorous protocol with more subjects involved. These devices and cuffs can be used with confidence.
format Online
Article
Text
id pubmed-8989751
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-89897512022-04-13 Validation of the A&D UA‐651 Plus/UA‐651SL Plus automated sphygmomanometer according to the ISO 81060–2, 2018 and ISO 81060–2 Amendment 1, 2020, which resulted in the currently pending Amendment to Amendment 1 Alpert, Bruce S. J Clin Hypertens (Greenwich) Short Research Article In 2020 the ANSI/AAMI/ISO published Amendment 1 to the 81060–2 Standard. The purpose was to try to address issues relating to the distribution of subjects with respect to limb size ranges. It also required a new plot showing the limb circumferences on the x‐axis and the device minus reference errors on the y‐axis (Alpert plot). The study reported here showed inadequacies of the Amendment if significantly overlapping cuffs were to be tested. The innovative approach led to the testing of 135 subjects. Requirements for blood pressure (BP), arm circumference, and gender were fulfilled. The standard same arm sequential protocol was used. Criterion 1 calculations gave a mean ± standard deviation (SD) device minus manual BP values of 0.22 ± 7.90 mmHg for systolic BP and ‐0.68 ± 7.36 mmHg for diastolic BP. The maximum allowed mean value for either measurement is 5.0 mmHg. The SD values for Criterion 2 were 6.03 mmHg for systolic BP (maximum allowed 6.95) and 6.47 mmHg for diastolic BP (maximum allowed 6.90). All results passed the Standard requirements. This study demonstrated inadequacies of Amendment 1 and led to the development of an Amendment 2, still in the process of finalization. The new Amendment corrects the issues with significantly overlapping cuffs and “wide‐range” cuffs. The A&D UA‐651 Plus/UA‐651SL Plus BP monitors and the five cuffs not only passed the requirements of 81060–2:2018 and Amendment 1 but also a revised and more rigorous protocol with more subjects involved. These devices and cuffs can be used with confidence. John Wiley and Sons Inc. 2022-03-21 /pmc/articles/PMC8989751/ /pubmed/35312160 http://dx.doi.org/10.1111/jch.14464 Text en © 2022 The Authors. The Journal of Clinical Hypertension published by Wiley Periodicals LLC https://creativecommons.org/licenses/by-nc/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
spellingShingle Short Research Article
Alpert, Bruce S.
Validation of the A&D UA‐651 Plus/UA‐651SL Plus automated sphygmomanometer according to the ISO 81060–2, 2018 and ISO 81060–2 Amendment 1, 2020, which resulted in the currently pending Amendment to Amendment 1
title Validation of the A&D UA‐651 Plus/UA‐651SL Plus automated sphygmomanometer according to the ISO 81060–2, 2018 and ISO 81060–2 Amendment 1, 2020, which resulted in the currently pending Amendment to Amendment 1
title_full Validation of the A&D UA‐651 Plus/UA‐651SL Plus automated sphygmomanometer according to the ISO 81060–2, 2018 and ISO 81060–2 Amendment 1, 2020, which resulted in the currently pending Amendment to Amendment 1
title_fullStr Validation of the A&D UA‐651 Plus/UA‐651SL Plus automated sphygmomanometer according to the ISO 81060–2, 2018 and ISO 81060–2 Amendment 1, 2020, which resulted in the currently pending Amendment to Amendment 1
title_full_unstemmed Validation of the A&D UA‐651 Plus/UA‐651SL Plus automated sphygmomanometer according to the ISO 81060–2, 2018 and ISO 81060–2 Amendment 1, 2020, which resulted in the currently pending Amendment to Amendment 1
title_short Validation of the A&D UA‐651 Plus/UA‐651SL Plus automated sphygmomanometer according to the ISO 81060–2, 2018 and ISO 81060–2 Amendment 1, 2020, which resulted in the currently pending Amendment to Amendment 1
title_sort validation of the a&d ua‐651 plus/ua‐651sl plus automated sphygmomanometer according to the iso 81060–2, 2018 and iso 81060–2 amendment 1, 2020, which resulted in the currently pending amendment to amendment 1
topic Short Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8989751/
https://www.ncbi.nlm.nih.gov/pubmed/35312160
http://dx.doi.org/10.1111/jch.14464
work_keys_str_mv AT alpertbruces validationoftheadua651plusua651slplusautomatedsphygmomanometeraccordingtotheiso8106022018andiso810602amendment12020whichresultedinthecurrentlypendingamendmenttoamendment1