Nitinol Staples for Olecranon Osteotomy Fixation, Juxtacortical Versus Inset, Effect on Biomechanical Stability
PURPOSE: Hardware prominence is a concern in the fixation of olecranon osteotomies. Staple fixation has provided low-profile secure fixation in other areas of orthopedics. Without insetting, staples still have subcutaneous prominence. This study examines whether nitinol staples, when inset into bone...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8991424/ https://www.ncbi.nlm.nih.gov/pubmed/35415560 http://dx.doi.org/10.1016/j.jhsg.2021.03.004 |
_version_ | 1784683567302836224 |
---|---|
author | Mahenthiran, Ajay Kacena-Merrell, Ethan Chen, Weinong W. Lim, Boon Him Dineen, Hannah |
author_facet | Mahenthiran, Ajay Kacena-Merrell, Ethan Chen, Weinong W. Lim, Boon Him Dineen, Hannah |
author_sort | Mahenthiran, Ajay |
collection | PubMed |
description | PURPOSE: Hardware prominence is a concern in the fixation of olecranon osteotomies. Staple fixation has provided low-profile secure fixation in other areas of orthopedics. Without insetting, staples still have subcutaneous prominence. This study examines whether nitinol staples, when inset into bone via cortical notching, in an olecranon osteotomy can provide fixation strength sufficient for daily activities. METHODS: Olecranon osteotomies were created in 8 cadaver arms and fixed with 2 nitinol staples. For inset and juxtacortical (noninset) staples, a micrometer measured the displacement between preplaced proximal and distal wires for 3 increasing loads: 0 N, 15 N, and 150 N. This measurement reflected the loss of osteotomy compression. We placed each arm in a pneumatic machine that flexed the elbow from 0° to 90° for 500 cycles at each load. We performed a 2-tailed t test (α value 0.05, β value 0.2) to evaluate for differences in the loss of compression between inset and noninset nitinol staples. RESULTS: We performed the displacement measurement procedure for both staple types at each of the 3 loads. At 0 N, the average displacement of inset was 0 mm and that of noninset was 0.02 mm. At 15 N, the average displacement of inset was 0.02 mm and that of noninset was 0.04 mm. At 150 N, the average displacement of inset was 0.05 mm and that of noninset was 0.09 mm. When comparing the displacement at the 3 force loads, there were no statistically significant differences between the staple types (P = .323). CONCLUSIONS: This study shows that inset staples do not considerably weaken osteotomy fixation with nitinol staples. Thus, nitinol staples may provide a low-profile, operatively-efficient fixation method compared with tension-band or screw-and-plate fixation methods for olecranon osteotomies. Future research can include comparing staples with plate constructs. Type of study/level of evidence: Therapeutic III. |
format | Online Article Text |
id | pubmed-8991424 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-89914242022-04-11 Nitinol Staples for Olecranon Osteotomy Fixation, Juxtacortical Versus Inset, Effect on Biomechanical Stability Mahenthiran, Ajay Kacena-Merrell, Ethan Chen, Weinong W. Lim, Boon Him Dineen, Hannah J Hand Surg Glob Online Original Research PURPOSE: Hardware prominence is a concern in the fixation of olecranon osteotomies. Staple fixation has provided low-profile secure fixation in other areas of orthopedics. Without insetting, staples still have subcutaneous prominence. This study examines whether nitinol staples, when inset into bone via cortical notching, in an olecranon osteotomy can provide fixation strength sufficient for daily activities. METHODS: Olecranon osteotomies were created in 8 cadaver arms and fixed with 2 nitinol staples. For inset and juxtacortical (noninset) staples, a micrometer measured the displacement between preplaced proximal and distal wires for 3 increasing loads: 0 N, 15 N, and 150 N. This measurement reflected the loss of osteotomy compression. We placed each arm in a pneumatic machine that flexed the elbow from 0° to 90° for 500 cycles at each load. We performed a 2-tailed t test (α value 0.05, β value 0.2) to evaluate for differences in the loss of compression between inset and noninset nitinol staples. RESULTS: We performed the displacement measurement procedure for both staple types at each of the 3 loads. At 0 N, the average displacement of inset was 0 mm and that of noninset was 0.02 mm. At 15 N, the average displacement of inset was 0.02 mm and that of noninset was 0.04 mm. At 150 N, the average displacement of inset was 0.05 mm and that of noninset was 0.09 mm. When comparing the displacement at the 3 force loads, there were no statistically significant differences between the staple types (P = .323). CONCLUSIONS: This study shows that inset staples do not considerably weaken osteotomy fixation with nitinol staples. Thus, nitinol staples may provide a low-profile, operatively-efficient fixation method compared with tension-band or screw-and-plate fixation methods for olecranon osteotomies. Future research can include comparing staples with plate constructs. Type of study/level of evidence: Therapeutic III. Elsevier 2021-04-27 /pmc/articles/PMC8991424/ /pubmed/35415560 http://dx.doi.org/10.1016/j.jhsg.2021.03.004 Text en © 2021 The Authors https://creativecommons.org/licenses/by/4.0/This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Original Research Mahenthiran, Ajay Kacena-Merrell, Ethan Chen, Weinong W. Lim, Boon Him Dineen, Hannah Nitinol Staples for Olecranon Osteotomy Fixation, Juxtacortical Versus Inset, Effect on Biomechanical Stability |
title | Nitinol Staples for Olecranon Osteotomy Fixation, Juxtacortical Versus Inset, Effect on Biomechanical Stability |
title_full | Nitinol Staples for Olecranon Osteotomy Fixation, Juxtacortical Versus Inset, Effect on Biomechanical Stability |
title_fullStr | Nitinol Staples for Olecranon Osteotomy Fixation, Juxtacortical Versus Inset, Effect on Biomechanical Stability |
title_full_unstemmed | Nitinol Staples for Olecranon Osteotomy Fixation, Juxtacortical Versus Inset, Effect on Biomechanical Stability |
title_short | Nitinol Staples for Olecranon Osteotomy Fixation, Juxtacortical Versus Inset, Effect on Biomechanical Stability |
title_sort | nitinol staples for olecranon osteotomy fixation, juxtacortical versus inset, effect on biomechanical stability |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8991424/ https://www.ncbi.nlm.nih.gov/pubmed/35415560 http://dx.doi.org/10.1016/j.jhsg.2021.03.004 |
work_keys_str_mv | AT mahenthiranajay nitinolstaplesforolecranonosteotomyfixationjuxtacorticalversusinseteffectonbiomechanicalstability AT kacenamerrellethan nitinolstaplesforolecranonosteotomyfixationjuxtacorticalversusinseteffectonbiomechanicalstability AT chenweinongw nitinolstaplesforolecranonosteotomyfixationjuxtacorticalversusinseteffectonbiomechanicalstability AT limboonhim nitinolstaplesforolecranonosteotomyfixationjuxtacorticalversusinseteffectonbiomechanicalstability AT dineenhannah nitinolstaplesforolecranonosteotomyfixationjuxtacorticalversusinseteffectonbiomechanicalstability |