Experimental Evaluation of the Elson Test Efficiency Following Central Slip Injury
PURPOSE: The purpose of this article is to explore the amount of work, quantitated by flexion and extension cycles, that is needed to obtain a positive Elson test following a central slip injury. METHODS: Thirteen frozen cadaveric fingers from individuals with an average age of 79.6 years were used....
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8991862/ https://www.ncbi.nlm.nih.gov/pubmed/35415584 http://dx.doi.org/10.1016/j.jhsg.2021.08.004 |
_version_ | 1784683659178016768 |
---|---|
author | Houston, Tyler Shipley, Tyler Bilderback, Karl Clark, Michael Barton, R. Shane Cvek, Urska Kilgore, Phillip Trutschl, Marjan Solitro, Giovanni F. |
author_facet | Houston, Tyler Shipley, Tyler Bilderback, Karl Clark, Michael Barton, R. Shane Cvek, Urska Kilgore, Phillip Trutschl, Marjan Solitro, Giovanni F. |
author_sort | Houston, Tyler |
collection | PubMed |
description | PURPOSE: The purpose of this article is to explore the amount of work, quantitated by flexion and extension cycles, that is needed to obtain a positive Elson test following a central slip injury. METHODS: Thirteen frozen cadaveric fingers from individuals with an average age of 79.6 years were used. Testing was performed by imposing sinusoidal displacement of the 2 tendons, with loads ranging from 30 N to 2 N at 1 Hz. Following transection to the central slip, each finger was cycled 1,000 times using the same protocol adopted for the control. Following 100, 200, 300, and 1,000 cycles, we measured the extension angles of the metacarpophalangeal, proximal interphalangeal, and distal interphalangeal joints from the flexed position and the distance between landmarks of the extensor apparatus and simulated an Elson test. RESULTS: In both the fingers, the range of motion of the metacarpophalangeal and distal interphalangeal joints measured in the controls remained unchanged, whereas the range of motion of the proximal interphalangeal joint was significantly reduced immediately after central slip transection. Combining both ring and middle fingers, for a displacement of 5 mm, the force measured in the control (1.05 ± 0.69 N) increased to the value of 2.36 ± 0.97 N at the 1,000th cycle. Although the middle finger has shown a significant difference in force at 100 cycles following central slip transection, 200 cycles were needed to observe a difference on the ring finger. CONCLUSIONS: In controlled conditions, there is a variation in resistance to flexion of the distal interphalangeal joint. However, the amplitude of the forces is so small that they are likely imperceptible clinically. Delayed testing should be considered to increase the sensitivity of the test or in patients experiencing pain. TYPE OF STUDY/LEVEL OF EVIDENCE: Diagnostic V. |
format | Online Article Text |
id | pubmed-8991862 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-89918622022-04-11 Experimental Evaluation of the Elson Test Efficiency Following Central Slip Injury Houston, Tyler Shipley, Tyler Bilderback, Karl Clark, Michael Barton, R. Shane Cvek, Urska Kilgore, Phillip Trutschl, Marjan Solitro, Giovanni F. J Hand Surg Glob Online Original Research PURPOSE: The purpose of this article is to explore the amount of work, quantitated by flexion and extension cycles, that is needed to obtain a positive Elson test following a central slip injury. METHODS: Thirteen frozen cadaveric fingers from individuals with an average age of 79.6 years were used. Testing was performed by imposing sinusoidal displacement of the 2 tendons, with loads ranging from 30 N to 2 N at 1 Hz. Following transection to the central slip, each finger was cycled 1,000 times using the same protocol adopted for the control. Following 100, 200, 300, and 1,000 cycles, we measured the extension angles of the metacarpophalangeal, proximal interphalangeal, and distal interphalangeal joints from the flexed position and the distance between landmarks of the extensor apparatus and simulated an Elson test. RESULTS: In both the fingers, the range of motion of the metacarpophalangeal and distal interphalangeal joints measured in the controls remained unchanged, whereas the range of motion of the proximal interphalangeal joint was significantly reduced immediately after central slip transection. Combining both ring and middle fingers, for a displacement of 5 mm, the force measured in the control (1.05 ± 0.69 N) increased to the value of 2.36 ± 0.97 N at the 1,000th cycle. Although the middle finger has shown a significant difference in force at 100 cycles following central slip transection, 200 cycles were needed to observe a difference on the ring finger. CONCLUSIONS: In controlled conditions, there is a variation in resistance to flexion of the distal interphalangeal joint. However, the amplitude of the forces is so small that they are likely imperceptible clinically. Delayed testing should be considered to increase the sensitivity of the test or in patients experiencing pain. TYPE OF STUDY/LEVEL OF EVIDENCE: Diagnostic V. Elsevier 2021-09-20 /pmc/articles/PMC8991862/ /pubmed/35415584 http://dx.doi.org/10.1016/j.jhsg.2021.08.004 Text en © 2021 The Authors https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/). |
spellingShingle | Original Research Houston, Tyler Shipley, Tyler Bilderback, Karl Clark, Michael Barton, R. Shane Cvek, Urska Kilgore, Phillip Trutschl, Marjan Solitro, Giovanni F. Experimental Evaluation of the Elson Test Efficiency Following Central Slip Injury |
title | Experimental Evaluation of the Elson Test Efficiency Following Central Slip Injury |
title_full | Experimental Evaluation of the Elson Test Efficiency Following Central Slip Injury |
title_fullStr | Experimental Evaluation of the Elson Test Efficiency Following Central Slip Injury |
title_full_unstemmed | Experimental Evaluation of the Elson Test Efficiency Following Central Slip Injury |
title_short | Experimental Evaluation of the Elson Test Efficiency Following Central Slip Injury |
title_sort | experimental evaluation of the elson test efficiency following central slip injury |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8991862/ https://www.ncbi.nlm.nih.gov/pubmed/35415584 http://dx.doi.org/10.1016/j.jhsg.2021.08.004 |
work_keys_str_mv | AT houstontyler experimentalevaluationoftheelsontestefficiencyfollowingcentralslipinjury AT shipleytyler experimentalevaluationoftheelsontestefficiencyfollowingcentralslipinjury AT bilderbackkarl experimentalevaluationoftheelsontestefficiencyfollowingcentralslipinjury AT clarkmichael experimentalevaluationoftheelsontestefficiencyfollowingcentralslipinjury AT bartonrshane experimentalevaluationoftheelsontestefficiencyfollowingcentralslipinjury AT cvekurska experimentalevaluationoftheelsontestefficiencyfollowingcentralslipinjury AT kilgorephillip experimentalevaluationoftheelsontestefficiencyfollowingcentralslipinjury AT trutschlmarjan experimentalevaluationoftheelsontestefficiencyfollowingcentralslipinjury AT solitrogiovannif experimentalevaluationoftheelsontestefficiencyfollowingcentralslipinjury |