Modified electron beam output calibration based on IAEA Technical Report Series 398
PURPOSE: The recently worldwide standard measurement of electron beam reference dosimetry include the International Atomic Energy Agency (IAEA) Technical Report Series (TRS)‐398 and Association of Physicists in Medicine (AAPM) Task Group (TG)‐51 protocols. Muir et al. have modified calibration metho...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8992941/ https://www.ncbi.nlm.nih.gov/pubmed/35226389 http://dx.doi.org/10.1002/acm2.13573 |
_version_ | 1784683808543473664 |
---|---|
author | Pawiro, Supriyanto Ardjo Mahfirotin, Dwi Aprilia Assegab, Muhamad Iqbal Wibowo, Wahyu Edy |
author_facet | Pawiro, Supriyanto Ardjo Mahfirotin, Dwi Aprilia Assegab, Muhamad Iqbal Wibowo, Wahyu Edy |
author_sort | Pawiro, Supriyanto Ardjo |
collection | PubMed |
description | PURPOSE: The recently worldwide standard measurement of electron beam reference dosimetry include the International Atomic Energy Agency (IAEA) Technical Report Series (TRS)‐398 and Association of Physicists in Medicine (AAPM) Task Group (TG)‐51 protocols. Muir et al. have modified calibration methods for electron beam calibration based on AAPM TG‐51. They found that the use of cylindrical chambers at low energy gave acceptable results. In this study, we propose and report a modified calibration for electron beam based on IAEA TRS‐398, the standard reference dosimetry protocol worldwide. METHODS: This work was carried out with energies of 6, 8, 10, 12, and 15 MeV. The electron beam is generated from Elektra Synergy Platform and Versa HD linear accelerator. The charge readings were measured with PTW 30013, IBA CC13, Exradin A1Sl, and Exradin A11 chambers connected to the electrometer. The dose calculation uses an equation of modified calibration for electron beam using the updated [Formula: see text] factor in previous work. The absorbed dose to water for electron beam is expressed in dose per monitor unit (cGy/MU). Thus, we compared dose per monitor unit (D/MU) calculation using a modified calibration to TRS‐398. RESULTS: In this work, we have succeeded in implementing the modified calibration of electron beam based on TRS‐398 by applying a cylindrical chamber in all energy beams and using the updated [Formula: see text] factor. The ratio of the absorbed dose to water between original and modified calibration protocols of TRS‐398 (D (w)) for the cylindrical chamber was 1.002 on the Elekta Synergy Platform and 1.000 on the Versa HD while for the parallel‐plate chamber it was 1.013 on the Elekta Synergy Platform and 1.014 on the Versa HD. Based on these results, both the cylindrical and parallel‐plate chambers are still within the tolerance limit allowed by the TRS‐398 protocol, which is ±2%. Therefore, modified calibration based on TRS‐398 gives acceptable results and is simpler to use clinically. |
format | Online Article Text |
id | pubmed-8992941 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-89929412022-04-13 Modified electron beam output calibration based on IAEA Technical Report Series 398 Pawiro, Supriyanto Ardjo Mahfirotin, Dwi Aprilia Assegab, Muhamad Iqbal Wibowo, Wahyu Edy J Appl Clin Med Phys Radiation Measurements PURPOSE: The recently worldwide standard measurement of electron beam reference dosimetry include the International Atomic Energy Agency (IAEA) Technical Report Series (TRS)‐398 and Association of Physicists in Medicine (AAPM) Task Group (TG)‐51 protocols. Muir et al. have modified calibration methods for electron beam calibration based on AAPM TG‐51. They found that the use of cylindrical chambers at low energy gave acceptable results. In this study, we propose and report a modified calibration for electron beam based on IAEA TRS‐398, the standard reference dosimetry protocol worldwide. METHODS: This work was carried out with energies of 6, 8, 10, 12, and 15 MeV. The electron beam is generated from Elektra Synergy Platform and Versa HD linear accelerator. The charge readings were measured with PTW 30013, IBA CC13, Exradin A1Sl, and Exradin A11 chambers connected to the electrometer. The dose calculation uses an equation of modified calibration for electron beam using the updated [Formula: see text] factor in previous work. The absorbed dose to water for electron beam is expressed in dose per monitor unit (cGy/MU). Thus, we compared dose per monitor unit (D/MU) calculation using a modified calibration to TRS‐398. RESULTS: In this work, we have succeeded in implementing the modified calibration of electron beam based on TRS‐398 by applying a cylindrical chamber in all energy beams and using the updated [Formula: see text] factor. The ratio of the absorbed dose to water between original and modified calibration protocols of TRS‐398 (D (w)) for the cylindrical chamber was 1.002 on the Elekta Synergy Platform and 1.000 on the Versa HD while for the parallel‐plate chamber it was 1.013 on the Elekta Synergy Platform and 1.014 on the Versa HD. Based on these results, both the cylindrical and parallel‐plate chambers are still within the tolerance limit allowed by the TRS‐398 protocol, which is ±2%. Therefore, modified calibration based on TRS‐398 gives acceptable results and is simpler to use clinically. John Wiley and Sons Inc. 2022-02-28 /pmc/articles/PMC8992941/ /pubmed/35226389 http://dx.doi.org/10.1002/acm2.13573 Text en © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Measurements Pawiro, Supriyanto Ardjo Mahfirotin, Dwi Aprilia Assegab, Muhamad Iqbal Wibowo, Wahyu Edy Modified electron beam output calibration based on IAEA Technical Report Series 398 |
title | Modified electron beam output calibration based on IAEA Technical Report Series 398 |
title_full | Modified electron beam output calibration based on IAEA Technical Report Series 398 |
title_fullStr | Modified electron beam output calibration based on IAEA Technical Report Series 398 |
title_full_unstemmed | Modified electron beam output calibration based on IAEA Technical Report Series 398 |
title_short | Modified electron beam output calibration based on IAEA Technical Report Series 398 |
title_sort | modified electron beam output calibration based on iaea technical report series 398 |
topic | Radiation Measurements |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8992941/ https://www.ncbi.nlm.nih.gov/pubmed/35226389 http://dx.doi.org/10.1002/acm2.13573 |
work_keys_str_mv | AT pawirosupriyantoardjo modifiedelectronbeamoutputcalibrationbasedoniaeatechnicalreportseries398 AT mahfirotindwiaprilia modifiedelectronbeamoutputcalibrationbasedoniaeatechnicalreportseries398 AT assegabmuhamadiqbal modifiedelectronbeamoutputcalibrationbasedoniaeatechnicalreportseries398 AT wibowowahyuedy modifiedelectronbeamoutputcalibrationbasedoniaeatechnicalreportseries398 |