Performance evaluation of the i‐Smart 300E cartridge for point‐of‐care electrolyte measurement in serum and plasma

BACKGROUND: Electrolytes are measured regularly in a variety of clinical settings because electrolyte imbalance can be life‐threatening. Although arterial blood‐gas analysis reports electrolyte levels, the result often is discrepant with results from serum and plasma samples. Since prompt and accura...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Beomki, Park, Hyung‐Doo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8993652/
https://www.ncbi.nlm.nih.gov/pubmed/35156738
http://dx.doi.org/10.1002/jcla.24295
_version_ 1784683944236548096
author Lee, Beomki
Park, Hyung‐Doo
author_facet Lee, Beomki
Park, Hyung‐Doo
author_sort Lee, Beomki
collection PubMed
description BACKGROUND: Electrolytes are measured regularly in a variety of clinical settings because electrolyte imbalance can be life‐threatening. Although arterial blood‐gas analysis reports electrolyte levels, the result often is discrepant with results from serum and plasma samples. Since prompt and accurate measurement of serum electrolyte levels could allow early treatment, point‐of‐care (POC) electrolyte analyzers would be beneficial. We evaluated a POC electrolyte analyzer cartridge based on the Clinical and Laboratory Standard Institute (CLSI) guidelines. METHODS: Precision and linearity were assessed according to the CLSI EP05‐A3 and EP06‐A guidelines, respectively. A comparison study was conducted with both serum and plasma samples according to the CLSI EP09‐A3. For serum, results from the i‐Smart 300E analyzer were compared with results from the Nova 8 and i‐Smart 30 analyzers. For plasma, results were compared among the i‐Smart 300E, Nova 8, i‐Smart 30, and Cobas c702 analyzers. RESULTS: Coefficients of variation in the precision analysis were all less than 5%. Linearity assessment demonstrated a coefficient of determination between 0.999 and 1.000 for all analytes. The comparison study showed a high Pearson's correlation coefficient greater than 0.9 for all analytes, instruments, and specimens. CONCLUSIONS: The i‐Smart 300E demonstrated good analytical performance. Its use could be beneficial in terms of both efficiency and clinical outcome in point‐of‐care testing (POCT) for electrolyte levels from serum and plasma samples.
format Online
Article
Text
id pubmed-8993652
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-89936522022-04-13 Performance evaluation of the i‐Smart 300E cartridge for point‐of‐care electrolyte measurement in serum and plasma Lee, Beomki Park, Hyung‐Doo J Clin Lab Anal Research Articles BACKGROUND: Electrolytes are measured regularly in a variety of clinical settings because electrolyte imbalance can be life‐threatening. Although arterial blood‐gas analysis reports electrolyte levels, the result often is discrepant with results from serum and plasma samples. Since prompt and accurate measurement of serum electrolyte levels could allow early treatment, point‐of‐care (POC) electrolyte analyzers would be beneficial. We evaluated a POC electrolyte analyzer cartridge based on the Clinical and Laboratory Standard Institute (CLSI) guidelines. METHODS: Precision and linearity were assessed according to the CLSI EP05‐A3 and EP06‐A guidelines, respectively. A comparison study was conducted with both serum and plasma samples according to the CLSI EP09‐A3. For serum, results from the i‐Smart 300E analyzer were compared with results from the Nova 8 and i‐Smart 30 analyzers. For plasma, results were compared among the i‐Smart 300E, Nova 8, i‐Smart 30, and Cobas c702 analyzers. RESULTS: Coefficients of variation in the precision analysis were all less than 5%. Linearity assessment demonstrated a coefficient of determination between 0.999 and 1.000 for all analytes. The comparison study showed a high Pearson's correlation coefficient greater than 0.9 for all analytes, instruments, and specimens. CONCLUSIONS: The i‐Smart 300E demonstrated good analytical performance. Its use could be beneficial in terms of both efficiency and clinical outcome in point‐of‐care testing (POCT) for electrolyte levels from serum and plasma samples. John Wiley and Sons Inc. 2022-02-14 /pmc/articles/PMC8993652/ /pubmed/35156738 http://dx.doi.org/10.1002/jcla.24295 Text en © 2022 The Authors. Journal of Clinical Laboratory Analysis published by Wiley Periodicals LLC. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc-nd/4.0/ (https://creativecommons.org/licenses/by-nc-nd/4.0/) License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made.
spellingShingle Research Articles
Lee, Beomki
Park, Hyung‐Doo
Performance evaluation of the i‐Smart 300E cartridge for point‐of‐care electrolyte measurement in serum and plasma
title Performance evaluation of the i‐Smart 300E cartridge for point‐of‐care electrolyte measurement in serum and plasma
title_full Performance evaluation of the i‐Smart 300E cartridge for point‐of‐care electrolyte measurement in serum and plasma
title_fullStr Performance evaluation of the i‐Smart 300E cartridge for point‐of‐care electrolyte measurement in serum and plasma
title_full_unstemmed Performance evaluation of the i‐Smart 300E cartridge for point‐of‐care electrolyte measurement in serum and plasma
title_short Performance evaluation of the i‐Smart 300E cartridge for point‐of‐care electrolyte measurement in serum and plasma
title_sort performance evaluation of the i‐smart 300e cartridge for point‐of‐care electrolyte measurement in serum and plasma
topic Research Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8993652/
https://www.ncbi.nlm.nih.gov/pubmed/35156738
http://dx.doi.org/10.1002/jcla.24295
work_keys_str_mv AT leebeomki performanceevaluationoftheismart300ecartridgeforpointofcareelectrolytemeasurementinserumandplasma
AT parkhyungdoo performanceevaluationoftheismart300ecartridgeforpointofcareelectrolytemeasurementinserumandplasma