Cargando…

EARL compliance measurements on the biograph vision Quadra PET/CT system with a long axial field of view

BACKGROUND: Our aim was to determine sets of reconstruction parameters for the Biograph Vision Quadra (Siemens Healthineers) PET/CT system that result in quantitative images compliant with the European Association of Nuclear Medicine Research Ltd. (EARL) criteria. Using the Biograph Vision 600 (Siem...

Descripción completa

Detalles Bibliográficos
Autores principales: Prenosil, George A., Hentschel, Michael, Weitzel, Thilo, Sari, Hasan, Shi, Kuangyu, Afshar-Oromieh, Ali, Rominger, Axel
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8994003/
https://www.ncbi.nlm.nih.gov/pubmed/35394263
http://dx.doi.org/10.1186/s40658-022-00455-1
_version_ 1784684018822807552
author Prenosil, George A.
Hentschel, Michael
Weitzel, Thilo
Sari, Hasan
Shi, Kuangyu
Afshar-Oromieh, Ali
Rominger, Axel
author_facet Prenosil, George A.
Hentschel, Michael
Weitzel, Thilo
Sari, Hasan
Shi, Kuangyu
Afshar-Oromieh, Ali
Rominger, Axel
author_sort Prenosil, George A.
collection PubMed
description BACKGROUND: Our aim was to determine sets of reconstruction parameters for the Biograph Vision Quadra (Siemens Healthineers) PET/CT system that result in quantitative images compliant with the European Association of Nuclear Medicine Research Ltd. (EARL) criteria. Using the Biograph Vision 600 (Siemens Healthineers) PET/CT technology but extending the axial field of view to 106 cm, gives the Vision Quadra currently an around fivefold higher sensitivity over the Vision 600 with otherwise comparable spatial resolution. Therefore, we also investigated how the number of incident positron decays—i.e., exposure—affects EARL compliance. This will allow estimating a minimal acquisition time or a minimal applied dose in clinical scans while retaining data comparability. METHODS: We measured activity recovery curves on a NEMA IEC body phantom filled with an aqueous (18)F solution and a sphere to background ratio of 10–1 according to the latest EARL guidelines. Reconstructing 3570 image sets with varying OSEM PSF iterations, post-reconstruction Gaussian filter full width at half maximum (FWHM), and varying exposure from 59 kDecays/ml (= 3 s frame duration) to 59.2 MDecays/ml (= 1 h), allowed us to determine sets of parameters to achieve compliance with the current EARL 1 and EARL 2 standards. Recovery coefficients (RCs) were calculated for the metrics RC(max), RC(mean), and RC(peak), and the respective recovery curves were analyzed for monotonicity. The background’s coefficient of variation (COV) was also calculated. RESULTS: Using 6 iterations, 5 subsets and 7.8 mm Gauss filtering resulted in optimal EARL1 compliance and recovery curve monotonicity in all analyzed frames, except in the 3 s frames. Most robust EARL2 compliance and monotonicity were achieved with 2 iterations, 5 subsets, and 3.6 mm Gauss FWHM in frames with durations between 30 s and 10 min. RC(peak) only impeded EARL2 compliance in the 10 s and 3 s frames. CONCLUSIONS: While EARL1 compliance was robust over most exposure ranges, EARL2 compliance required exposures between 1.2 MDecays/ml to 11.5 MDecays/ml. The Biograph Vision Quadra’s high sensitivity makes frames as short as 10 s feasible for comparable quantitative images. Lowering EARL2 RC(max) limits closer to unity would possibly even permit shorter frames.
format Online
Article
Text
id pubmed-8994003
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-89940032022-04-22 EARL compliance measurements on the biograph vision Quadra PET/CT system with a long axial field of view Prenosil, George A. Hentschel, Michael Weitzel, Thilo Sari, Hasan Shi, Kuangyu Afshar-Oromieh, Ali Rominger, Axel EJNMMI Phys Original Research BACKGROUND: Our aim was to determine sets of reconstruction parameters for the Biograph Vision Quadra (Siemens Healthineers) PET/CT system that result in quantitative images compliant with the European Association of Nuclear Medicine Research Ltd. (EARL) criteria. Using the Biograph Vision 600 (Siemens Healthineers) PET/CT technology but extending the axial field of view to 106 cm, gives the Vision Quadra currently an around fivefold higher sensitivity over the Vision 600 with otherwise comparable spatial resolution. Therefore, we also investigated how the number of incident positron decays—i.e., exposure—affects EARL compliance. This will allow estimating a minimal acquisition time or a minimal applied dose in clinical scans while retaining data comparability. METHODS: We measured activity recovery curves on a NEMA IEC body phantom filled with an aqueous (18)F solution and a sphere to background ratio of 10–1 according to the latest EARL guidelines. Reconstructing 3570 image sets with varying OSEM PSF iterations, post-reconstruction Gaussian filter full width at half maximum (FWHM), and varying exposure from 59 kDecays/ml (= 3 s frame duration) to 59.2 MDecays/ml (= 1 h), allowed us to determine sets of parameters to achieve compliance with the current EARL 1 and EARL 2 standards. Recovery coefficients (RCs) were calculated for the metrics RC(max), RC(mean), and RC(peak), and the respective recovery curves were analyzed for monotonicity. The background’s coefficient of variation (COV) was also calculated. RESULTS: Using 6 iterations, 5 subsets and 7.8 mm Gauss filtering resulted in optimal EARL1 compliance and recovery curve monotonicity in all analyzed frames, except in the 3 s frames. Most robust EARL2 compliance and monotonicity were achieved with 2 iterations, 5 subsets, and 3.6 mm Gauss FWHM in frames with durations between 30 s and 10 min. RC(peak) only impeded EARL2 compliance in the 10 s and 3 s frames. CONCLUSIONS: While EARL1 compliance was robust over most exposure ranges, EARL2 compliance required exposures between 1.2 MDecays/ml to 11.5 MDecays/ml. The Biograph Vision Quadra’s high sensitivity makes frames as short as 10 s feasible for comparable quantitative images. Lowering EARL2 RC(max) limits closer to unity would possibly even permit shorter frames. Springer International Publishing 2022-04-08 /pmc/articles/PMC8994003/ /pubmed/35394263 http://dx.doi.org/10.1186/s40658-022-00455-1 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Research
Prenosil, George A.
Hentschel, Michael
Weitzel, Thilo
Sari, Hasan
Shi, Kuangyu
Afshar-Oromieh, Ali
Rominger, Axel
EARL compliance measurements on the biograph vision Quadra PET/CT system with a long axial field of view
title EARL compliance measurements on the biograph vision Quadra PET/CT system with a long axial field of view
title_full EARL compliance measurements on the biograph vision Quadra PET/CT system with a long axial field of view
title_fullStr EARL compliance measurements on the biograph vision Quadra PET/CT system with a long axial field of view
title_full_unstemmed EARL compliance measurements on the biograph vision Quadra PET/CT system with a long axial field of view
title_short EARL compliance measurements on the biograph vision Quadra PET/CT system with a long axial field of view
title_sort earl compliance measurements on the biograph vision quadra pet/ct system with a long axial field of view
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8994003/
https://www.ncbi.nlm.nih.gov/pubmed/35394263
http://dx.doi.org/10.1186/s40658-022-00455-1
work_keys_str_mv AT prenosilgeorgea earlcompliancemeasurementsonthebiographvisionquadrapetctsystemwithalongaxialfieldofview
AT hentschelmichael earlcompliancemeasurementsonthebiographvisionquadrapetctsystemwithalongaxialfieldofview
AT weitzelthilo earlcompliancemeasurementsonthebiographvisionquadrapetctsystemwithalongaxialfieldofview
AT sarihasan earlcompliancemeasurementsonthebiographvisionquadrapetctsystemwithalongaxialfieldofview
AT shikuangyu earlcompliancemeasurementsonthebiographvisionquadrapetctsystemwithalongaxialfieldofview
AT afsharoromiehali earlcompliancemeasurementsonthebiographvisionquadrapetctsystemwithalongaxialfieldofview
AT romingeraxel earlcompliancemeasurementsonthebiographvisionquadrapetctsystemwithalongaxialfieldofview