Performance of Centargo: A Novel Piston-Based Injection System for High Throughput in CE CT

PURPOSE: To compare an investigational device (MEDRAD(®) Centargo CT Injection System, “Centargo”) to the currently available MEDRAD(®) Stellant CT Injection System (“Stellant”), in terms of efficiency, injector performance, and user satisfaction. PATIENTS AND METHODS: A total of 425 patients at two...

Descripción completa

Detalles Bibliográficos
Autores principales: Kemper, Corey A, Mihl, Casper, Martens, Bibi, McDermott, Michael C, Hendriks, Babs M F
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8994651/
https://www.ncbi.nlm.nih.gov/pubmed/35411194
http://dx.doi.org/10.2147/MDER.S353221
_version_ 1784684150448455680
author Kemper, Corey A
Mihl, Casper
Martens, Bibi
McDermott, Michael C
Hendriks, Babs M F
author_facet Kemper, Corey A
Mihl, Casper
Martens, Bibi
McDermott, Michael C
Hendriks, Babs M F
author_sort Kemper, Corey A
collection PubMed
description PURPOSE: To compare an investigational device (MEDRAD(®) Centargo CT Injection System, “Centargo”) to the currently available MEDRAD(®) Stellant CT Injection System (“Stellant”), in terms of efficiency, injector performance, and user satisfaction. PATIENTS AND METHODS: A total of 425 patients at two sites were enrolled; 198 patients in phase one, a randomized study (98 Stellant and 100 Centargo). The second observational phase included 227 patients who were injected with Centargo. Phase one recorded times for setup, disassembly, and patient changeovers. Demographic data, subjective image quality, and injection parameters were collected. Phase two assessed usability via a questionnaire provided to all end-users of both systems (radiographers). RESULTS: Patient changeover times were statistically significantly faster with Centargo (15.4s ± 8.7s vs 53.7s ± 19.6s, p < 0.001). Centargo day-setup times were similar to Stellant (138.1s ± 92s vs 151.8s ± 30.6s, p = 0.33) and end-of-day-disassembly times were significantly slower (60.6s ± 27s vs 17.1s ± 12.9s, p < 0.001). Based on four different scenarios modelling patient throughput, the projected time savings with Centargo over Stellant was 40–63%, with the highest efficiency improvements for higher throughputs and the use of larger contrast medium bottles. Both Centargo and Stellant usability averaged between “Very Easy” and “Easy” in all responses to the questionnaire. There were no instances of interrupted injections due to communication loss or detected air and no insufficient images due to injector performance. No safety issues were identified. CONCLUSION: Centargo was able to demonstrate improved efficiency as compared to Stellant while maintaining injector performance and high usability scores.
format Online
Article
Text
id pubmed-8994651
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Dove
record_format MEDLINE/PubMed
spelling pubmed-89946512022-04-10 Performance of Centargo: A Novel Piston-Based Injection System for High Throughput in CE CT Kemper, Corey A Mihl, Casper Martens, Bibi McDermott, Michael C Hendriks, Babs M F Med Devices (Auckl) Original Research PURPOSE: To compare an investigational device (MEDRAD(®) Centargo CT Injection System, “Centargo”) to the currently available MEDRAD(®) Stellant CT Injection System (“Stellant”), in terms of efficiency, injector performance, and user satisfaction. PATIENTS AND METHODS: A total of 425 patients at two sites were enrolled; 198 patients in phase one, a randomized study (98 Stellant and 100 Centargo). The second observational phase included 227 patients who were injected with Centargo. Phase one recorded times for setup, disassembly, and patient changeovers. Demographic data, subjective image quality, and injection parameters were collected. Phase two assessed usability via a questionnaire provided to all end-users of both systems (radiographers). RESULTS: Patient changeover times were statistically significantly faster with Centargo (15.4s ± 8.7s vs 53.7s ± 19.6s, p < 0.001). Centargo day-setup times were similar to Stellant (138.1s ± 92s vs 151.8s ± 30.6s, p = 0.33) and end-of-day-disassembly times were significantly slower (60.6s ± 27s vs 17.1s ± 12.9s, p < 0.001). Based on four different scenarios modelling patient throughput, the projected time savings with Centargo over Stellant was 40–63%, with the highest efficiency improvements for higher throughputs and the use of larger contrast medium bottles. Both Centargo and Stellant usability averaged between “Very Easy” and “Easy” in all responses to the questionnaire. There were no instances of interrupted injections due to communication loss or detected air and no insufficient images due to injector performance. No safety issues were identified. CONCLUSION: Centargo was able to demonstrate improved efficiency as compared to Stellant while maintaining injector performance and high usability scores. Dove 2022-04-05 /pmc/articles/PMC8994651/ /pubmed/35411194 http://dx.doi.org/10.2147/MDER.S353221 Text en © 2022 Kemper et al. https://creativecommons.org/licenses/by-nc/3.0/This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/ (https://creativecommons.org/licenses/by-nc/3.0/) ). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php).
spellingShingle Original Research
Kemper, Corey A
Mihl, Casper
Martens, Bibi
McDermott, Michael C
Hendriks, Babs M F
Performance of Centargo: A Novel Piston-Based Injection System for High Throughput in CE CT
title Performance of Centargo: A Novel Piston-Based Injection System for High Throughput in CE CT
title_full Performance of Centargo: A Novel Piston-Based Injection System for High Throughput in CE CT
title_fullStr Performance of Centargo: A Novel Piston-Based Injection System for High Throughput in CE CT
title_full_unstemmed Performance of Centargo: A Novel Piston-Based Injection System for High Throughput in CE CT
title_short Performance of Centargo: A Novel Piston-Based Injection System for High Throughput in CE CT
title_sort performance of centargo: a novel piston-based injection system for high throughput in ce ct
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8994651/
https://www.ncbi.nlm.nih.gov/pubmed/35411194
http://dx.doi.org/10.2147/MDER.S353221
work_keys_str_mv AT kempercoreya performanceofcentargoanovelpistonbasedinjectionsystemforhighthroughputincect
AT mihlcasper performanceofcentargoanovelpistonbasedinjectionsystemforhighthroughputincect
AT martensbibi performanceofcentargoanovelpistonbasedinjectionsystemforhighthroughputincect
AT mcdermottmichaelc performanceofcentargoanovelpistonbasedinjectionsystemforhighthroughputincect
AT hendriksbabsmf performanceofcentargoanovelpistonbasedinjectionsystemforhighthroughputincect