Cargando…
Performance evaluation study of a commercially available smart patient-controlled analgesia pump with the microbalance method and an infusion analyzer
BACKGROUND: Patient-controlled analgesia (PCA) has been widely used as an effective medical treatment for pain and for postoperative analgesia. However, improper dose errors in intravenous (IV) administration of narcotic analgesics from a PCA infusion pump can cause patient harm. Furthermore, opioid...
Autores principales: | Park, Jinsoo, Jung, Bongsu |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Dental Society of Anesthsiology
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8995673/ https://www.ncbi.nlm.nih.gov/pubmed/35449784 http://dx.doi.org/10.17245/jdapm.2022.22.2.129 |
Ejemplares similares
-
Efficacy evaluation of syringe pump developed for continuous drug infusion
por: Jung, Bongsu, et al.
Publicado: (2016) -
A Comparative Pulse Accuracy Study of Two Commercially Available Patch Insulin Infusion Pumps
por: Bowen, Jenna L, et al.
Publicado: (2016) -
Smart syringe pumps for drug infusion during dental intravenous sedation
por: Seo, Kwang-Suk, et al.
Publicado: (2016) -
Performance Evaluation of the i-SmartCare 10 Analyzer and Method Comparison of Six Point-of-Care Blood Gas Analyzers
por: Kim, Sang-Mi, et al.
Publicado: (2022) -
Monitoring fluid intake by commercially available smart water bottles
por: Cohen, Rachel, et al.
Publicado: (2022)