Thermoluminescent Characteristics of GR-200, TLD-700H and TLD-100 for Low Dose Measurement: Linearity, Repeatability, Dose Rate and Photon Energy Dependence
BACKGROUND: The dose values obtained from procedures of diagnostic radiology are relatively low. To accurately and precisely measure the dose values in this dose range, it is necessary to know the characteristics of dosimeters. OBJECTIVE: The aim of this study was to evaluate several thermoluminesce...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Shiraz University of Medical Sciences
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8995761/ https://www.ncbi.nlm.nih.gov/pubmed/35433523 http://dx.doi.org/10.31661/jbpe.v0i0.1247 |
_version_ | 1784684351103959040 |
---|---|
author | Alizadeh, Mojgan Mohseni, Mehran Farhood, Bagher Aliasgharzadeh, Akbar Moradi, Habiballah Ramazani-Moghaddam-Arani, Ahmad |
author_facet | Alizadeh, Mojgan Mohseni, Mehran Farhood, Bagher Aliasgharzadeh, Akbar Moradi, Habiballah Ramazani-Moghaddam-Arani, Ahmad |
author_sort | Alizadeh, Mojgan |
collection | PubMed |
description | BACKGROUND: The dose values obtained from procedures of diagnostic radiology are relatively low. To accurately and precisely measure the dose values in this dose range, it is necessary to know the characteristics of dosimeters. OBJECTIVE: The aim of this study was to evaluate several thermoluminescent characteristics of GR-200, TLD-700H and TLD-100 for low dose measurement. MATERIAL AND METHODS: In this experimental study, linearity, repeatability, dose rate and photon energy dependence of different TLD materials were investigated in a 0.05-10 mGy range dose. It is noteworthy that the data obtained from TLD-100 were considered as reference and the data obtained from two other types of TLDs were compared with them. RESULTS: For all three types of TLD materials, there are linear relations between absorbed dose values to TLDs and their responses. TLD-100 and TLD-700H have very low sensitivity than GR-200. For GR-200 and TLD-100, the coefficients of variation values (%) are 3.00% and 2.01%, respectively, that these values are within the tolerance limit (<7.5%). However, this value for TLD-700H is 10.85% which it is more than the reported tolerance limit. Furthermore, remarkable effects of dose rate and photon energy dependence on the responses of GR-200 are not observed in a 0.5-4 mGy dose range; nevertheless, remarkable effects of dose rate and photon energy dependence on the responses of TLD-100 and TLD-700H are found in this dose range. CONCLUSION: The evaluated thermoluminescent characteristics for GR-200 are better than two other types of TLDs (TLD-100 and TLD-700H) for low dose values. |
format | Online Article Text |
id | pubmed-8995761 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Shiraz University of Medical Sciences |
record_format | MEDLINE/PubMed |
spelling | pubmed-89957612022-04-15 Thermoluminescent Characteristics of GR-200, TLD-700H and TLD-100 for Low Dose Measurement: Linearity, Repeatability, Dose Rate and Photon Energy Dependence Alizadeh, Mojgan Mohseni, Mehran Farhood, Bagher Aliasgharzadeh, Akbar Moradi, Habiballah Ramazani-Moghaddam-Arani, Ahmad J Biomed Phys Eng Original Article BACKGROUND: The dose values obtained from procedures of diagnostic radiology are relatively low. To accurately and precisely measure the dose values in this dose range, it is necessary to know the characteristics of dosimeters. OBJECTIVE: The aim of this study was to evaluate several thermoluminescent characteristics of GR-200, TLD-700H and TLD-100 for low dose measurement. MATERIAL AND METHODS: In this experimental study, linearity, repeatability, dose rate and photon energy dependence of different TLD materials were investigated in a 0.05-10 mGy range dose. It is noteworthy that the data obtained from TLD-100 were considered as reference and the data obtained from two other types of TLDs were compared with them. RESULTS: For all three types of TLD materials, there are linear relations between absorbed dose values to TLDs and their responses. TLD-100 and TLD-700H have very low sensitivity than GR-200. For GR-200 and TLD-100, the coefficients of variation values (%) are 3.00% and 2.01%, respectively, that these values are within the tolerance limit (<7.5%). However, this value for TLD-700H is 10.85% which it is more than the reported tolerance limit. Furthermore, remarkable effects of dose rate and photon energy dependence on the responses of GR-200 are not observed in a 0.5-4 mGy dose range; nevertheless, remarkable effects of dose rate and photon energy dependence on the responses of TLD-100 and TLD-700H are found in this dose range. CONCLUSION: The evaluated thermoluminescent characteristics for GR-200 are better than two other types of TLDs (TLD-100 and TLD-700H) for low dose values. Shiraz University of Medical Sciences 2022-04-01 /pmc/articles/PMC8995761/ /pubmed/35433523 http://dx.doi.org/10.31661/jbpe.v0i0.1247 Text en Copyright: © Journal of Biomedical Physics and Engineering https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 Unported License, ( http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) ) which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Alizadeh, Mojgan Mohseni, Mehran Farhood, Bagher Aliasgharzadeh, Akbar Moradi, Habiballah Ramazani-Moghaddam-Arani, Ahmad Thermoluminescent Characteristics of GR-200, TLD-700H and TLD-100 for Low Dose Measurement: Linearity, Repeatability, Dose Rate and Photon Energy Dependence |
title | Thermoluminescent Characteristics of GR-200, TLD-700H and TLD-100 for Low Dose Measurement: Linearity, Repeatability, Dose Rate and Photon Energy Dependence |
title_full | Thermoluminescent Characteristics of GR-200, TLD-700H and TLD-100 for Low Dose Measurement: Linearity, Repeatability, Dose Rate and Photon Energy Dependence |
title_fullStr | Thermoluminescent Characteristics of GR-200, TLD-700H and TLD-100 for Low Dose Measurement: Linearity, Repeatability, Dose Rate and Photon Energy Dependence |
title_full_unstemmed | Thermoluminescent Characteristics of GR-200, TLD-700H and TLD-100 for Low Dose Measurement: Linearity, Repeatability, Dose Rate and Photon Energy Dependence |
title_short | Thermoluminescent Characteristics of GR-200, TLD-700H and TLD-100 for Low Dose Measurement: Linearity, Repeatability, Dose Rate and Photon Energy Dependence |
title_sort | thermoluminescent characteristics of gr-200, tld-700h and tld-100 for low dose measurement: linearity, repeatability, dose rate and photon energy dependence |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8995761/ https://www.ncbi.nlm.nih.gov/pubmed/35433523 http://dx.doi.org/10.31661/jbpe.v0i0.1247 |
work_keys_str_mv | AT alizadehmojgan thermoluminescentcharacteristicsofgr200tld700handtld100forlowdosemeasurementlinearityrepeatabilitydoserateandphotonenergydependence AT mohsenimehran thermoluminescentcharacteristicsofgr200tld700handtld100forlowdosemeasurementlinearityrepeatabilitydoserateandphotonenergydependence AT farhoodbagher thermoluminescentcharacteristicsofgr200tld700handtld100forlowdosemeasurementlinearityrepeatabilitydoserateandphotonenergydependence AT aliasgharzadehakbar thermoluminescentcharacteristicsofgr200tld700handtld100forlowdosemeasurementlinearityrepeatabilitydoserateandphotonenergydependence AT moradihabiballah thermoluminescentcharacteristicsofgr200tld700handtld100forlowdosemeasurementlinearityrepeatabilitydoserateandphotonenergydependence AT ramazanimoghaddamaraniahmad thermoluminescentcharacteristicsofgr200tld700handtld100forlowdosemeasurementlinearityrepeatabilitydoserateandphotonenergydependence |