Temperature Changes during Implant Osteotomy Preparations in Fresh Human Cadaver Tibiae, Comparing Straight with Tapered Drills †
The success of osseointegration depends on many factors. With temperatures beyond a 47 °C threshold over 1 min, bone survival may be impaired. The purpose of the study was to evaluate, in fresh human cadaver tibiae, the temperature changes during osteotomy preparations using two straight and two tap...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8999768/ https://www.ncbi.nlm.nih.gov/pubmed/35407704 http://dx.doi.org/10.3390/ma15072369 |
_version_ | 1784685269201453056 |
---|---|
author | Soldatos, Nikolaos Nelson-Rabe, Laura Palanker, Nathan Angelov, Nikola Romanos, Georgios Weltman, Robin |
author_facet | Soldatos, Nikolaos Nelson-Rabe, Laura Palanker, Nathan Angelov, Nikola Romanos, Georgios Weltman, Robin |
author_sort | Soldatos, Nikolaos |
collection | PubMed |
description | The success of osseointegration depends on many factors. With temperatures beyond a 47 °C threshold over 1 min, bone survival may be impaired. The purpose of the study was to evaluate, in fresh human cadaver tibiae, the temperature changes during osteotomy preparations using two straight and two tapered implant systems’ drills, external irrigation, and varying revolutions per minute (RPM). The tibiae from a fresh female cadaver were harvested bilaterally. Two tapered and two straight design drills were assessed. Two-hundred and forty osteotomies were prepared at 6 mm depth following the drill sequence of the manufacturers’ protocol for each drilling speed. Difference in temperature (ΔΤ) was calculated by subtracting the baseline from the maximum temperature (ΔT = T(max) − T(base)). Drill design and drill diameter, as independent variables or synergistically, had a significant effect on ΔΤ. Tapered drills: As the drill diameter increased, ΔΤ increased at all RPM. Straight drills: As the drill diameter increased, ΔΤ remained constant or slightly decreased at all RPM. Drill diameter and design had a significant effect on ΔΤ in human tibiae, which never exceeded the critical threshold of 47 °C. Tapered drills caused significantly greater heat production compared to straight drills. |
format | Online Article Text |
id | pubmed-8999768 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-89997682022-04-12 Temperature Changes during Implant Osteotomy Preparations in Fresh Human Cadaver Tibiae, Comparing Straight with Tapered Drills † Soldatos, Nikolaos Nelson-Rabe, Laura Palanker, Nathan Angelov, Nikola Romanos, Georgios Weltman, Robin Materials (Basel) Article The success of osseointegration depends on many factors. With temperatures beyond a 47 °C threshold over 1 min, bone survival may be impaired. The purpose of the study was to evaluate, in fresh human cadaver tibiae, the temperature changes during osteotomy preparations using two straight and two tapered implant systems’ drills, external irrigation, and varying revolutions per minute (RPM). The tibiae from a fresh female cadaver were harvested bilaterally. Two tapered and two straight design drills were assessed. Two-hundred and forty osteotomies were prepared at 6 mm depth following the drill sequence of the manufacturers’ protocol for each drilling speed. Difference in temperature (ΔΤ) was calculated by subtracting the baseline from the maximum temperature (ΔT = T(max) − T(base)). Drill design and drill diameter, as independent variables or synergistically, had a significant effect on ΔΤ. Tapered drills: As the drill diameter increased, ΔΤ increased at all RPM. Straight drills: As the drill diameter increased, ΔΤ remained constant or slightly decreased at all RPM. Drill diameter and design had a significant effect on ΔΤ in human tibiae, which never exceeded the critical threshold of 47 °C. Tapered drills caused significantly greater heat production compared to straight drills. MDPI 2022-03-23 /pmc/articles/PMC8999768/ /pubmed/35407704 http://dx.doi.org/10.3390/ma15072369 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Soldatos, Nikolaos Nelson-Rabe, Laura Palanker, Nathan Angelov, Nikola Romanos, Georgios Weltman, Robin Temperature Changes during Implant Osteotomy Preparations in Fresh Human Cadaver Tibiae, Comparing Straight with Tapered Drills † |
title | Temperature Changes during Implant Osteotomy Preparations in Fresh Human Cadaver Tibiae, Comparing Straight with Tapered Drills † |
title_full | Temperature Changes during Implant Osteotomy Preparations in Fresh Human Cadaver Tibiae, Comparing Straight with Tapered Drills † |
title_fullStr | Temperature Changes during Implant Osteotomy Preparations in Fresh Human Cadaver Tibiae, Comparing Straight with Tapered Drills † |
title_full_unstemmed | Temperature Changes during Implant Osteotomy Preparations in Fresh Human Cadaver Tibiae, Comparing Straight with Tapered Drills † |
title_short | Temperature Changes during Implant Osteotomy Preparations in Fresh Human Cadaver Tibiae, Comparing Straight with Tapered Drills † |
title_sort | temperature changes during implant osteotomy preparations in fresh human cadaver tibiae, comparing straight with tapered drills † |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8999768/ https://www.ncbi.nlm.nih.gov/pubmed/35407704 http://dx.doi.org/10.3390/ma15072369 |
work_keys_str_mv | AT soldatosnikolaos temperaturechangesduringimplantosteotomypreparationsinfreshhumancadavertibiaecomparingstraightwithtapereddrills AT nelsonrabelaura temperaturechangesduringimplantosteotomypreparationsinfreshhumancadavertibiaecomparingstraightwithtapereddrills AT palankernathan temperaturechangesduringimplantosteotomypreparationsinfreshhumancadavertibiaecomparingstraightwithtapereddrills AT angelovnikola temperaturechangesduringimplantosteotomypreparationsinfreshhumancadavertibiaecomparingstraightwithtapereddrills AT romanosgeorgios temperaturechangesduringimplantosteotomypreparationsinfreshhumancadavertibiaecomparingstraightwithtapereddrills AT weltmanrobin temperaturechangesduringimplantosteotomypreparationsinfreshhumancadavertibiaecomparingstraightwithtapereddrills |