The image quality and diagnostic accuracy of T1-mapping-based synthetic late gadolinium enhancement imaging: comparison with conventional late gadolinium enhancement imaging in real-life clinical situation

BACKGROUNDS: Synthetic late gadolinium enhancement (LGE) images are less sensitive to inversion time (TI) and robust to motion artifact, because it is generated retrospectively by post-contrast T1-mapping images. To explore the clinical applicability of synthetic LGE, we investigated the image quali...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Suji, Kim, Panki, Im, Dong Jin, Suh, Young Joo, Hong, Yoo Jin, Choi, Byoung Wook, Kim, Young Jin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9008937/
https://www.ncbi.nlm.nih.gov/pubmed/35418081
http://dx.doi.org/10.1186/s12968-022-00857-1
_version_ 1784687170247720960
author Lee, Suji
Kim, Panki
Im, Dong Jin
Suh, Young Joo
Hong, Yoo Jin
Choi, Byoung Wook
Kim, Young Jin
author_facet Lee, Suji
Kim, Panki
Im, Dong Jin
Suh, Young Joo
Hong, Yoo Jin
Choi, Byoung Wook
Kim, Young Jin
author_sort Lee, Suji
collection PubMed
description BACKGROUNDS: Synthetic late gadolinium enhancement (LGE) images are less sensitive to inversion time (TI) and robust to motion artifact, because it is generated retrospectively by post-contrast T1-mapping images. To explore the clinical applicability of synthetic LGE, we investigated the image quality and diagnostic accuracy of synthetic LGE images, in comparison to that of conventional LGE for various disease groups. METHOD AND MATERIALS: From July to November 2019, a total of 98 patients who underwent cardiovascular magnetic resonance imaging (CMR), including LGE and T1-mapping sequences, with suspicion of myocardial abnormality were retrospectively included. Synthetic magnitude inversion-recovery (IR) and phase-sensitive IR (PSIR) images were generated through calculations based on the post-contrast T1-mapping sequence. Three cardiothoracic radiologists independently analyzed the image quality of conventional and synthetic LGE images on an ordinal scale with per-segment basis and the image qualities were compared with chi-square test. The agreement of LGE detection was analyzed on per-patient and per-segment basis with Cohen’s kappa test. In addition, the LGE area and percentage were semi-quantitatively analyzed for LGE positive ischemic (n = 14) and hypertrophic cardiomyopathy (n = 13) subgroups by two cardiothoracic radiologists. The difference of quantified LGE area and percentage between conventional and synthetic LGE images were assessed with Mann–Whitney U-test and the inter-reader agreement was assessed with Bland–Altman analysis. RESULTS: The image quality of synthetic images was significantly better than conventional images in both magnitude IR and PSIR through all three observers (P < 0.001, all). The agreements of per-patient and per-segment LGE detection rates were excellent (kappa = 0.815–0.864). The semi-quantitative analysis showed no significant difference in the LGE area and percentage between conventional and synthetic LGE images. In the inter-reader agreement showed only small systematic differences in both magnitude IR and PSIR and synthetic LGE images showed smaller systematic biases compared to conventional LGE images. CONCLUSION: Compared to conventional LGE images, synthetic LGE images have better image quality in real-life clinical situation. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12968-022-00857-1.
format Online
Article
Text
id pubmed-9008937
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-90089372022-04-15 The image quality and diagnostic accuracy of T1-mapping-based synthetic late gadolinium enhancement imaging: comparison with conventional late gadolinium enhancement imaging in real-life clinical situation Lee, Suji Kim, Panki Im, Dong Jin Suh, Young Joo Hong, Yoo Jin Choi, Byoung Wook Kim, Young Jin J Cardiovasc Magn Reson Research BACKGROUNDS: Synthetic late gadolinium enhancement (LGE) images are less sensitive to inversion time (TI) and robust to motion artifact, because it is generated retrospectively by post-contrast T1-mapping images. To explore the clinical applicability of synthetic LGE, we investigated the image quality and diagnostic accuracy of synthetic LGE images, in comparison to that of conventional LGE for various disease groups. METHOD AND MATERIALS: From July to November 2019, a total of 98 patients who underwent cardiovascular magnetic resonance imaging (CMR), including LGE and T1-mapping sequences, with suspicion of myocardial abnormality were retrospectively included. Synthetic magnitude inversion-recovery (IR) and phase-sensitive IR (PSIR) images were generated through calculations based on the post-contrast T1-mapping sequence. Three cardiothoracic radiologists independently analyzed the image quality of conventional and synthetic LGE images on an ordinal scale with per-segment basis and the image qualities were compared with chi-square test. The agreement of LGE detection was analyzed on per-patient and per-segment basis with Cohen’s kappa test. In addition, the LGE area and percentage were semi-quantitatively analyzed for LGE positive ischemic (n = 14) and hypertrophic cardiomyopathy (n = 13) subgroups by two cardiothoracic radiologists. The difference of quantified LGE area and percentage between conventional and synthetic LGE images were assessed with Mann–Whitney U-test and the inter-reader agreement was assessed with Bland–Altman analysis. RESULTS: The image quality of synthetic images was significantly better than conventional images in both magnitude IR and PSIR through all three observers (P < 0.001, all). The agreements of per-patient and per-segment LGE detection rates were excellent (kappa = 0.815–0.864). The semi-quantitative analysis showed no significant difference in the LGE area and percentage between conventional and synthetic LGE images. In the inter-reader agreement showed only small systematic differences in both magnitude IR and PSIR and synthetic LGE images showed smaller systematic biases compared to conventional LGE images. CONCLUSION: Compared to conventional LGE images, synthetic LGE images have better image quality in real-life clinical situation. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12968-022-00857-1. BioMed Central 2022-04-14 /pmc/articles/PMC9008937/ /pubmed/35418081 http://dx.doi.org/10.1186/s12968-022-00857-1 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
spellingShingle Research
Lee, Suji
Kim, Panki
Im, Dong Jin
Suh, Young Joo
Hong, Yoo Jin
Choi, Byoung Wook
Kim, Young Jin
The image quality and diagnostic accuracy of T1-mapping-based synthetic late gadolinium enhancement imaging: comparison with conventional late gadolinium enhancement imaging in real-life clinical situation
title The image quality and diagnostic accuracy of T1-mapping-based synthetic late gadolinium enhancement imaging: comparison with conventional late gadolinium enhancement imaging in real-life clinical situation
title_full The image quality and diagnostic accuracy of T1-mapping-based synthetic late gadolinium enhancement imaging: comparison with conventional late gadolinium enhancement imaging in real-life clinical situation
title_fullStr The image quality and diagnostic accuracy of T1-mapping-based synthetic late gadolinium enhancement imaging: comparison with conventional late gadolinium enhancement imaging in real-life clinical situation
title_full_unstemmed The image quality and diagnostic accuracy of T1-mapping-based synthetic late gadolinium enhancement imaging: comparison with conventional late gadolinium enhancement imaging in real-life clinical situation
title_short The image quality and diagnostic accuracy of T1-mapping-based synthetic late gadolinium enhancement imaging: comparison with conventional late gadolinium enhancement imaging in real-life clinical situation
title_sort image quality and diagnostic accuracy of t1-mapping-based synthetic late gadolinium enhancement imaging: comparison with conventional late gadolinium enhancement imaging in real-life clinical situation
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9008937/
https://www.ncbi.nlm.nih.gov/pubmed/35418081
http://dx.doi.org/10.1186/s12968-022-00857-1
work_keys_str_mv AT leesuji theimagequalityanddiagnosticaccuracyoft1mappingbasedsyntheticlategadoliniumenhancementimagingcomparisonwithconventionallategadoliniumenhancementimaginginreallifeclinicalsituation
AT kimpanki theimagequalityanddiagnosticaccuracyoft1mappingbasedsyntheticlategadoliniumenhancementimagingcomparisonwithconventionallategadoliniumenhancementimaginginreallifeclinicalsituation
AT imdongjin theimagequalityanddiagnosticaccuracyoft1mappingbasedsyntheticlategadoliniumenhancementimagingcomparisonwithconventionallategadoliniumenhancementimaginginreallifeclinicalsituation
AT suhyoungjoo theimagequalityanddiagnosticaccuracyoft1mappingbasedsyntheticlategadoliniumenhancementimagingcomparisonwithconventionallategadoliniumenhancementimaginginreallifeclinicalsituation
AT hongyoojin theimagequalityanddiagnosticaccuracyoft1mappingbasedsyntheticlategadoliniumenhancementimagingcomparisonwithconventionallategadoliniumenhancementimaginginreallifeclinicalsituation
AT choibyoungwook theimagequalityanddiagnosticaccuracyoft1mappingbasedsyntheticlategadoliniumenhancementimagingcomparisonwithconventionallategadoliniumenhancementimaginginreallifeclinicalsituation
AT kimyoungjin theimagequalityanddiagnosticaccuracyoft1mappingbasedsyntheticlategadoliniumenhancementimagingcomparisonwithconventionallategadoliniumenhancementimaginginreallifeclinicalsituation
AT leesuji imagequalityanddiagnosticaccuracyoft1mappingbasedsyntheticlategadoliniumenhancementimagingcomparisonwithconventionallategadoliniumenhancementimaginginreallifeclinicalsituation
AT kimpanki imagequalityanddiagnosticaccuracyoft1mappingbasedsyntheticlategadoliniumenhancementimagingcomparisonwithconventionallategadoliniumenhancementimaginginreallifeclinicalsituation
AT imdongjin imagequalityanddiagnosticaccuracyoft1mappingbasedsyntheticlategadoliniumenhancementimagingcomparisonwithconventionallategadoliniumenhancementimaginginreallifeclinicalsituation
AT suhyoungjoo imagequalityanddiagnosticaccuracyoft1mappingbasedsyntheticlategadoliniumenhancementimagingcomparisonwithconventionallategadoliniumenhancementimaginginreallifeclinicalsituation
AT hongyoojin imagequalityanddiagnosticaccuracyoft1mappingbasedsyntheticlategadoliniumenhancementimagingcomparisonwithconventionallategadoliniumenhancementimaginginreallifeclinicalsituation
AT choibyoungwook imagequalityanddiagnosticaccuracyoft1mappingbasedsyntheticlategadoliniumenhancementimagingcomparisonwithconventionallategadoliniumenhancementimaginginreallifeclinicalsituation
AT kimyoungjin imagequalityanddiagnosticaccuracyoft1mappingbasedsyntheticlategadoliniumenhancementimagingcomparisonwithconventionallategadoliniumenhancementimaginginreallifeclinicalsituation