Nuclear Fragmentation Imaging for Carbon-Ion Radiation Therapy Monitoring: an In Silico Study
PURPOSE: This article presents an in vivo imaging technique based on nuclear fragmentation of carbon ions in irradiated tissues for potential real-time monitoring of carbon-ion radiation therapy (CIRT) treatment delivery and quality assurance purposes in clinical settings. MATERIALS AND METHODS: A p...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Particle Therapy Co-operative Group
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9009459/ https://www.ncbi.nlm.nih.gov/pubmed/35530183 http://dx.doi.org/10.14338/IJPT-20-00040.1 |
_version_ | 1784687272040333312 |
---|---|
author | Bey, Anissa Ma, Jiasen Furutani, Keith M. Herman, Michael G. Johnson, Jedediah E. Foote, Robert L. Beltran, Chris J. |
author_facet | Bey, Anissa Ma, Jiasen Furutani, Keith M. Herman, Michael G. Johnson, Jedediah E. Foote, Robert L. Beltran, Chris J. |
author_sort | Bey, Anissa |
collection | PubMed |
description | PURPOSE: This article presents an in vivo imaging technique based on nuclear fragmentation of carbon ions in irradiated tissues for potential real-time monitoring of carbon-ion radiation therapy (CIRT) treatment delivery and quality assurance purposes in clinical settings. MATERIALS AND METHODS: A proof-of-concept imaging and monitoring system (IMS) was devised to implement the technique. Monte Carlo simulations were performed for a prospective pencil-beam scanning CIRT nozzle. The development IMS benchmark considered a 5×5-cm(2) pixelated charged-particle detector stack positioned downstream from a target phantom and list-mode data acquisition. The abundance and production origins, that is, vertices, of the detected fragments were studied. Fragment trajectories were approximated by straight lines and a beam back-projection algorithm was built to reconstruct the vertices. The spatial distribution of the vertices was then used to determine plan relevant markers. RESULTS: The IMS technique was applied for a simulated CIRT case, a primary brain tumor. Four treatment plan monitoring markers were conclusively recovered: a depth dose distribution correlated profile, ion beam range, treatment target boundaries, and the beam spot position. Promising millimeter-scale (3-mm, ≤10% uncertainty) beam range and submillimeter (≤0.6-mm precision for shifts <3 cm) beam spot position verification accuracies were obtained for typical therapeutic energies between 150 and 290 MeV/u. CONCLUSIONS: This work demonstrated a viable online monitoring technique for CIRT treatment delivery. The method's strong advantage is that it requires few signal inputs (position and timing), which can be simultaneously acquired with readily available technology. Future investigations will probe the technique's applicability to motion-sensitive organ sites and patient tissue heterogeneities. In-beam measurements with candidate detector-acquisition systems are ultimately essential to validate the IMS benchmark performance and subsequent deployment in the clinic. |
format | Online Article Text |
id | pubmed-9009459 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | The Particle Therapy Co-operative Group |
record_format | MEDLINE/PubMed |
spelling | pubmed-90094592022-05-05 Nuclear Fragmentation Imaging for Carbon-Ion Radiation Therapy Monitoring: an In Silico Study Bey, Anissa Ma, Jiasen Furutani, Keith M. Herman, Michael G. Johnson, Jedediah E. Foote, Robert L. Beltran, Chris J. Int J Part Ther Original Articles PURPOSE: This article presents an in vivo imaging technique based on nuclear fragmentation of carbon ions in irradiated tissues for potential real-time monitoring of carbon-ion radiation therapy (CIRT) treatment delivery and quality assurance purposes in clinical settings. MATERIALS AND METHODS: A proof-of-concept imaging and monitoring system (IMS) was devised to implement the technique. Monte Carlo simulations were performed for a prospective pencil-beam scanning CIRT nozzle. The development IMS benchmark considered a 5×5-cm(2) pixelated charged-particle detector stack positioned downstream from a target phantom and list-mode data acquisition. The abundance and production origins, that is, vertices, of the detected fragments were studied. Fragment trajectories were approximated by straight lines and a beam back-projection algorithm was built to reconstruct the vertices. The spatial distribution of the vertices was then used to determine plan relevant markers. RESULTS: The IMS technique was applied for a simulated CIRT case, a primary brain tumor. Four treatment plan monitoring markers were conclusively recovered: a depth dose distribution correlated profile, ion beam range, treatment target boundaries, and the beam spot position. Promising millimeter-scale (3-mm, ≤10% uncertainty) beam range and submillimeter (≤0.6-mm precision for shifts <3 cm) beam spot position verification accuracies were obtained for typical therapeutic energies between 150 and 290 MeV/u. CONCLUSIONS: This work demonstrated a viable online monitoring technique for CIRT treatment delivery. The method's strong advantage is that it requires few signal inputs (position and timing), which can be simultaneously acquired with readily available technology. Future investigations will probe the technique's applicability to motion-sensitive organ sites and patient tissue heterogeneities. In-beam measurements with candidate detector-acquisition systems are ultimately essential to validate the IMS benchmark performance and subsequent deployment in the clinic. The Particle Therapy Co-operative Group 2021-09-01 /pmc/articles/PMC9009459/ /pubmed/35530183 http://dx.doi.org/10.14338/IJPT-20-00040.1 Text en ©Copyright 2021 The Author(s) https://creativecommons.org/licenses/by/4.0/Distributed under Creative Commons CC-BY. Open Access |
spellingShingle | Original Articles Bey, Anissa Ma, Jiasen Furutani, Keith M. Herman, Michael G. Johnson, Jedediah E. Foote, Robert L. Beltran, Chris J. Nuclear Fragmentation Imaging for Carbon-Ion Radiation Therapy Monitoring: an In Silico Study |
title | Nuclear Fragmentation Imaging for Carbon-Ion Radiation Therapy Monitoring: an In Silico Study |
title_full | Nuclear Fragmentation Imaging for Carbon-Ion Radiation Therapy Monitoring: an In Silico Study |
title_fullStr | Nuclear Fragmentation Imaging for Carbon-Ion Radiation Therapy Monitoring: an In Silico Study |
title_full_unstemmed | Nuclear Fragmentation Imaging for Carbon-Ion Radiation Therapy Monitoring: an In Silico Study |
title_short | Nuclear Fragmentation Imaging for Carbon-Ion Radiation Therapy Monitoring: an In Silico Study |
title_sort | nuclear fragmentation imaging for carbon-ion radiation therapy monitoring: an in silico study |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9009459/ https://www.ncbi.nlm.nih.gov/pubmed/35530183 http://dx.doi.org/10.14338/IJPT-20-00040.1 |
work_keys_str_mv | AT beyanissa nuclearfragmentationimagingforcarbonionradiationtherapymonitoringaninsilicostudy AT majiasen nuclearfragmentationimagingforcarbonionradiationtherapymonitoringaninsilicostudy AT furutanikeithm nuclearfragmentationimagingforcarbonionradiationtherapymonitoringaninsilicostudy AT hermanmichaelg nuclearfragmentationimagingforcarbonionradiationtherapymonitoringaninsilicostudy AT johnsonjedediahe nuclearfragmentationimagingforcarbonionradiationtherapymonitoringaninsilicostudy AT footerobertl nuclearfragmentationimagingforcarbonionradiationtherapymonitoringaninsilicostudy AT beltranchrisj nuclearfragmentationimagingforcarbonionradiationtherapymonitoringaninsilicostudy |