Using a roller pump for establishing extra-corporal membrane oxygenation (ECMO) – technical considerations for times of crisis
OBJECTIVE: The COVID-19 pandemic requires thinking about alternatives to establish ECMO when often-limited hardware resources are exhausted. Heart-lung-machines may potentially be used for ECMO but contain roller pumps as compared to centrifugal pumps in ECMO-circuits. We here tested roller pumps as...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9019428/ https://www.ncbi.nlm.nih.gov/pubmed/33637027 http://dx.doi.org/10.1177/0267659121996182 |
_version_ | 1784689273636651008 |
---|---|
author | Kaluza, Mirko May, Benjamin Doenst, Torsten |
author_facet | Kaluza, Mirko May, Benjamin Doenst, Torsten |
author_sort | Kaluza, Mirko |
collection | PubMed |
description | OBJECTIVE: The COVID-19 pandemic requires thinking about alternatives to establish ECMO when often-limited hardware resources are exhausted. Heart-lung-machines may potentially be used for ECMO but contain roller pumps as compared to centrifugal pumps in ECMO-circuits. We here tested roller pumps as rescue pump for ECMO-establishment. METHODS: We set up in vitro circuits on roller pumps from C5 heart-lung-machine with 5 l/minutes flow. In two series, we placed either PVC or silicon tubing for an ECMO circuit into the roller pump. We assessed the mechanical stress on the tubing (aiming to run the pump for at least 1 week), measured the temperature increase generated by the friction and assessed flow characteristics and its measurement in simulated situations resembling tube kinking and suction. RESULTS: The roller pumps led to expected and unexpected adverse events. PVC tubing burst between 36 and 78 hours, while silicon tubing lasted for at least 7 days. At 7 days, the silicone tubing showed significant signs of roller pump wear visible on the outside. The inside, however, was free of surface irregularities. Using these tubings in a roller pump led to a remarkable increase in circuit temperature (PVC: +12.0°C, silicone +2.9°C). Kinking or suction on the device caused the expected dramatic flow reduction (as assessed by direct measurement) while the roller pump display continued to show the preset flow. The roller pump is therefore not able to reliably determine the true flow rate. CONCLUSION: Roller pumps with silicone tubing but not PVC tubing may be used for running ECMO circuits. Silicone tubing may endure the roller pump shear forces for up to 1 week. Thus, repeated tubing repositioning may be a solution. Circuit heating and substantial limitations in flow detection should increase attention if clinical use in situations of crisis is considered. |
format | Online Article Text |
id | pubmed-9019428 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-90194282022-04-21 Using a roller pump for establishing extra-corporal membrane oxygenation (ECMO) – technical considerations for times of crisis Kaluza, Mirko May, Benjamin Doenst, Torsten Perfusion Original Papers OBJECTIVE: The COVID-19 pandemic requires thinking about alternatives to establish ECMO when often-limited hardware resources are exhausted. Heart-lung-machines may potentially be used for ECMO but contain roller pumps as compared to centrifugal pumps in ECMO-circuits. We here tested roller pumps as rescue pump for ECMO-establishment. METHODS: We set up in vitro circuits on roller pumps from C5 heart-lung-machine with 5 l/minutes flow. In two series, we placed either PVC or silicon tubing for an ECMO circuit into the roller pump. We assessed the mechanical stress on the tubing (aiming to run the pump for at least 1 week), measured the temperature increase generated by the friction and assessed flow characteristics and its measurement in simulated situations resembling tube kinking and suction. RESULTS: The roller pumps led to expected and unexpected adverse events. PVC tubing burst between 36 and 78 hours, while silicon tubing lasted for at least 7 days. At 7 days, the silicone tubing showed significant signs of roller pump wear visible on the outside. The inside, however, was free of surface irregularities. Using these tubings in a roller pump led to a remarkable increase in circuit temperature (PVC: +12.0°C, silicone +2.9°C). Kinking or suction on the device caused the expected dramatic flow reduction (as assessed by direct measurement) while the roller pump display continued to show the preset flow. The roller pump is therefore not able to reliably determine the true flow rate. CONCLUSION: Roller pumps with silicone tubing but not PVC tubing may be used for running ECMO circuits. Silicone tubing may endure the roller pump shear forces for up to 1 week. Thus, repeated tubing repositioning may be a solution. Circuit heating and substantial limitations in flow detection should increase attention if clinical use in situations of crisis is considered. SAGE Publications 2021-02-26 2022-04 /pmc/articles/PMC9019428/ /pubmed/33637027 http://dx.doi.org/10.1177/0267659121996182 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/This article is distributed under the terms of the Creative Commons Attribution 4.0 License (https://creativecommons.org/licenses/by/4.0/) which permits any use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Original Papers Kaluza, Mirko May, Benjamin Doenst, Torsten Using a roller pump for establishing extra-corporal membrane oxygenation (ECMO) – technical considerations for times of crisis |
title | Using a roller pump for establishing extra-corporal membrane oxygenation (ECMO) – technical considerations for times of crisis |
title_full | Using a roller pump for establishing extra-corporal membrane oxygenation (ECMO) – technical considerations for times of crisis |
title_fullStr | Using a roller pump for establishing extra-corporal membrane oxygenation (ECMO) – technical considerations for times of crisis |
title_full_unstemmed | Using a roller pump for establishing extra-corporal membrane oxygenation (ECMO) – technical considerations for times of crisis |
title_short | Using a roller pump for establishing extra-corporal membrane oxygenation (ECMO) – technical considerations for times of crisis |
title_sort | using a roller pump for establishing extra-corporal membrane oxygenation (ecmo) – technical considerations for times of crisis |
topic | Original Papers |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9019428/ https://www.ncbi.nlm.nih.gov/pubmed/33637027 http://dx.doi.org/10.1177/0267659121996182 |
work_keys_str_mv | AT kaluzamirko usingarollerpumpforestablishingextracorporalmembraneoxygenationecmotechnicalconsiderationsfortimesofcrisis AT maybenjamin usingarollerpumpforestablishingextracorporalmembraneoxygenationecmotechnicalconsiderationsfortimesofcrisis AT doensttorsten usingarollerpumpforestablishingextracorporalmembraneoxygenationecmotechnicalconsiderationsfortimesofcrisis |