“Catch, Flip, and Remove” Technique for Retrieval of a Hypermobile Detached Leadless Pacemaker: A Case Report and Literature Review
Leadless pacemaker (LP) detachment is a rare but life-threatening complication that may occur during implantation. While different snaring techniques have been described to remove partially or completely detached LPs, there are currently no reports of snaring a hypermobile LP that travels between di...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MediaSphere Medical
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9023022/ https://www.ncbi.nlm.nih.gov/pubmed/35474861 http://dx.doi.org/10.19102/icrm.2022.130404 |
Sumario: | Leadless pacemaker (LP) detachment is a rare but life-threatening complication that may occur during implantation. While different snaring techniques have been described to remove partially or completely detached LPs, there are currently no reports of snaring a hypermobile LP that travels between different cardiac chambers. This report describes a technique to successfully snare a hypermobile detached LP by first “catching” onto the tines for stabilization with the help of a multi-loop snare, followed by using a second snare for the proximal retrieval feature. |
---|