Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT

OBJECTIVES: Photon-counting detector CT (PCD-CT) promises a leap in spatial resolution due to smaller detector pixel sizes than implemented in energy-integrating detector CTs (EID-CT). Our objective was to compare the visualization of smallest bone details between PCD-CT and EID-CT using a mouse as...

Descripción completa

Detalles Bibliográficos
Autores principales: Bette, Stefanie J., Braun, Franziska M., Haerting, Mark, Decker, Josua A., Luitjens, Jan H., Scheurig-Muenkler, Christian, Kroencke, Thomas J., Schwarz, Florian
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9038873/
https://www.ncbi.nlm.nih.gov/pubmed/34936011
http://dx.doi.org/10.1007/s00330-021-08441-4
_version_ 1784693999906324480
author Bette, Stefanie J.
Braun, Franziska M.
Haerting, Mark
Decker, Josua A.
Luitjens, Jan H.
Scheurig-Muenkler, Christian
Kroencke, Thomas J.
Schwarz, Florian
author_facet Bette, Stefanie J.
Braun, Franziska M.
Haerting, Mark
Decker, Josua A.
Luitjens, Jan H.
Scheurig-Muenkler, Christian
Kroencke, Thomas J.
Schwarz, Florian
author_sort Bette, Stefanie J.
collection PubMed
description OBJECTIVES: Photon-counting detector CT (PCD-CT) promises a leap in spatial resolution due to smaller detector pixel sizes than implemented in energy-integrating detector CTs (EID-CT). Our objective was to compare the visualization of smallest bone details between PCD-CT and EID-CT using a mouse as a specimen. MATERIALS AND METHODS: Two euthanized mice were scanned at a 20-slice EID-CT and a dual-source PCD-CT in single-pixel mode at various CTDI(Vol) values. Image noise and signal-to-noise ratio (SNR) were evaluated using repeated ROI measurements. Edge sharpness of bones was compared by the maximal slope within CT value plots along sampling lines intersecting predefined bones of the spine. Two readers evaluated bone detail visualization at four regions of the spine on a three-point Likert scale at various CTDI(Vol)’s. Two radiologists selected the series with better detail visualization among each of 20 SNR-matched pairs of EID-CT and PCD-CT series. RESULTS: In CTDI(Vol)-matched scans, PCD-CT series showed significantly lower image noise (Noise(CTDI=5 mGy): 16.27 ± 1.39 vs. 23.46 ± 0.96 HU, p < 0.01), higher SNR (SNR(CTDI=5 mGy): 20.57 ± 1.89 vs. 14.00 ± 0.66, p < 0.01), and higher edge sharpness (Edge Slope(lumbar spine): 981 ± 160 vs. 608 ± 146 HU/mm, p < 0.01) than EID-CT series. Two radiologists considered the delineation of bone details as feasible at consistently lower CTDI(Vol) values at PCD-CT than at EID-CT. In comparison of SNR-matched reconstructions, PCD-CT series were still considered superior in almost all cases. CONCLUSIONS: In this head-to-head comparison, PCD-CT showed superior objective and subjective image quality characteristics over EID-CT for the delineation of tiniest bone details. Even in SNR-matched pairs (acquired at different CTDI(Vol)’s), PCD-CT was strongly preferred by radiologists. KEY POINTS: • In dose-matched scans, photon-counting detector CT series showed significantly less image noise, higher signal-to-noise ratio, and higher edge sharpness than energy-integrating detector CT series. • Human observers considered the delineation of tiny bone details as feasible at much lower dose levels in photon-counting detector CT than in energy-integrating detector CT. • In direct comparison of series matched for signal-to-noise ratio, photon-counting detector CT series were considered superior in almost all cases. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s00330-021-08441-4.
format Online
Article
Text
id pubmed-9038873
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-90388732022-05-07 Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT Bette, Stefanie J. Braun, Franziska M. Haerting, Mark Decker, Josua A. Luitjens, Jan H. Scheurig-Muenkler, Christian Kroencke, Thomas J. Schwarz, Florian Eur Radiol Computed Tomography OBJECTIVES: Photon-counting detector CT (PCD-CT) promises a leap in spatial resolution due to smaller detector pixel sizes than implemented in energy-integrating detector CTs (EID-CT). Our objective was to compare the visualization of smallest bone details between PCD-CT and EID-CT using a mouse as a specimen. MATERIALS AND METHODS: Two euthanized mice were scanned at a 20-slice EID-CT and a dual-source PCD-CT in single-pixel mode at various CTDI(Vol) values. Image noise and signal-to-noise ratio (SNR) were evaluated using repeated ROI measurements. Edge sharpness of bones was compared by the maximal slope within CT value plots along sampling lines intersecting predefined bones of the spine. Two readers evaluated bone detail visualization at four regions of the spine on a three-point Likert scale at various CTDI(Vol)’s. Two radiologists selected the series with better detail visualization among each of 20 SNR-matched pairs of EID-CT and PCD-CT series. RESULTS: In CTDI(Vol)-matched scans, PCD-CT series showed significantly lower image noise (Noise(CTDI=5 mGy): 16.27 ± 1.39 vs. 23.46 ± 0.96 HU, p < 0.01), higher SNR (SNR(CTDI=5 mGy): 20.57 ± 1.89 vs. 14.00 ± 0.66, p < 0.01), and higher edge sharpness (Edge Slope(lumbar spine): 981 ± 160 vs. 608 ± 146 HU/mm, p < 0.01) than EID-CT series. Two radiologists considered the delineation of bone details as feasible at consistently lower CTDI(Vol) values at PCD-CT than at EID-CT. In comparison of SNR-matched reconstructions, PCD-CT series were still considered superior in almost all cases. CONCLUSIONS: In this head-to-head comparison, PCD-CT showed superior objective and subjective image quality characteristics over EID-CT for the delineation of tiniest bone details. Even in SNR-matched pairs (acquired at different CTDI(Vol)’s), PCD-CT was strongly preferred by radiologists. KEY POINTS: • In dose-matched scans, photon-counting detector CT series showed significantly less image noise, higher signal-to-noise ratio, and higher edge sharpness than energy-integrating detector CT series. • Human observers considered the delineation of tiny bone details as feasible at much lower dose levels in photon-counting detector CT than in energy-integrating detector CT. • In direct comparison of series matched for signal-to-noise ratio, photon-counting detector CT series were considered superior in almost all cases. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s00330-021-08441-4. Springer Berlin Heidelberg 2021-12-22 2022 /pmc/articles/PMC9038873/ /pubmed/34936011 http://dx.doi.org/10.1007/s00330-021-08441-4 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Computed Tomography
Bette, Stefanie J.
Braun, Franziska M.
Haerting, Mark
Decker, Josua A.
Luitjens, Jan H.
Scheurig-Muenkler, Christian
Kroencke, Thomas J.
Schwarz, Florian
Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT
title Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT
title_full Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT
title_fullStr Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT
title_full_unstemmed Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT
title_short Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT
title_sort visualization of bone details in a novel photon-counting dual-source ct scanner—comparison with energy-integrating ct
topic Computed Tomography
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9038873/
https://www.ncbi.nlm.nih.gov/pubmed/34936011
http://dx.doi.org/10.1007/s00330-021-08441-4
work_keys_str_mv AT bettestefaniej visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct
AT braunfranziskam visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct
AT haertingmark visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct
AT deckerjosuaa visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct
AT luitjensjanh visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct
AT scheurigmuenklerchristian visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct
AT kroenckethomasj visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct
AT schwarzflorian visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct