Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT
OBJECTIVES: Photon-counting detector CT (PCD-CT) promises a leap in spatial resolution due to smaller detector pixel sizes than implemented in energy-integrating detector CTs (EID-CT). Our objective was to compare the visualization of smallest bone details between PCD-CT and EID-CT using a mouse as...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9038873/ https://www.ncbi.nlm.nih.gov/pubmed/34936011 http://dx.doi.org/10.1007/s00330-021-08441-4 |
_version_ | 1784693999906324480 |
---|---|
author | Bette, Stefanie J. Braun, Franziska M. Haerting, Mark Decker, Josua A. Luitjens, Jan H. Scheurig-Muenkler, Christian Kroencke, Thomas J. Schwarz, Florian |
author_facet | Bette, Stefanie J. Braun, Franziska M. Haerting, Mark Decker, Josua A. Luitjens, Jan H. Scheurig-Muenkler, Christian Kroencke, Thomas J. Schwarz, Florian |
author_sort | Bette, Stefanie J. |
collection | PubMed |
description | OBJECTIVES: Photon-counting detector CT (PCD-CT) promises a leap in spatial resolution due to smaller detector pixel sizes than implemented in energy-integrating detector CTs (EID-CT). Our objective was to compare the visualization of smallest bone details between PCD-CT and EID-CT using a mouse as a specimen. MATERIALS AND METHODS: Two euthanized mice were scanned at a 20-slice EID-CT and a dual-source PCD-CT in single-pixel mode at various CTDI(Vol) values. Image noise and signal-to-noise ratio (SNR) were evaluated using repeated ROI measurements. Edge sharpness of bones was compared by the maximal slope within CT value plots along sampling lines intersecting predefined bones of the spine. Two readers evaluated bone detail visualization at four regions of the spine on a three-point Likert scale at various CTDI(Vol)’s. Two radiologists selected the series with better detail visualization among each of 20 SNR-matched pairs of EID-CT and PCD-CT series. RESULTS: In CTDI(Vol)-matched scans, PCD-CT series showed significantly lower image noise (Noise(CTDI=5 mGy): 16.27 ± 1.39 vs. 23.46 ± 0.96 HU, p < 0.01), higher SNR (SNR(CTDI=5 mGy): 20.57 ± 1.89 vs. 14.00 ± 0.66, p < 0.01), and higher edge sharpness (Edge Slope(lumbar spine): 981 ± 160 vs. 608 ± 146 HU/mm, p < 0.01) than EID-CT series. Two radiologists considered the delineation of bone details as feasible at consistently lower CTDI(Vol) values at PCD-CT than at EID-CT. In comparison of SNR-matched reconstructions, PCD-CT series were still considered superior in almost all cases. CONCLUSIONS: In this head-to-head comparison, PCD-CT showed superior objective and subjective image quality characteristics over EID-CT for the delineation of tiniest bone details. Even in SNR-matched pairs (acquired at different CTDI(Vol)’s), PCD-CT was strongly preferred by radiologists. KEY POINTS: • In dose-matched scans, photon-counting detector CT series showed significantly less image noise, higher signal-to-noise ratio, and higher edge sharpness than energy-integrating detector CT series. • Human observers considered the delineation of tiny bone details as feasible at much lower dose levels in photon-counting detector CT than in energy-integrating detector CT. • In direct comparison of series matched for signal-to-noise ratio, photon-counting detector CT series were considered superior in almost all cases. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s00330-021-08441-4. |
format | Online Article Text |
id | pubmed-9038873 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Springer Berlin Heidelberg |
record_format | MEDLINE/PubMed |
spelling | pubmed-90388732022-05-07 Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT Bette, Stefanie J. Braun, Franziska M. Haerting, Mark Decker, Josua A. Luitjens, Jan H. Scheurig-Muenkler, Christian Kroencke, Thomas J. Schwarz, Florian Eur Radiol Computed Tomography OBJECTIVES: Photon-counting detector CT (PCD-CT) promises a leap in spatial resolution due to smaller detector pixel sizes than implemented in energy-integrating detector CTs (EID-CT). Our objective was to compare the visualization of smallest bone details between PCD-CT and EID-CT using a mouse as a specimen. MATERIALS AND METHODS: Two euthanized mice were scanned at a 20-slice EID-CT and a dual-source PCD-CT in single-pixel mode at various CTDI(Vol) values. Image noise and signal-to-noise ratio (SNR) were evaluated using repeated ROI measurements. Edge sharpness of bones was compared by the maximal slope within CT value plots along sampling lines intersecting predefined bones of the spine. Two readers evaluated bone detail visualization at four regions of the spine on a three-point Likert scale at various CTDI(Vol)’s. Two radiologists selected the series with better detail visualization among each of 20 SNR-matched pairs of EID-CT and PCD-CT series. RESULTS: In CTDI(Vol)-matched scans, PCD-CT series showed significantly lower image noise (Noise(CTDI=5 mGy): 16.27 ± 1.39 vs. 23.46 ± 0.96 HU, p < 0.01), higher SNR (SNR(CTDI=5 mGy): 20.57 ± 1.89 vs. 14.00 ± 0.66, p < 0.01), and higher edge sharpness (Edge Slope(lumbar spine): 981 ± 160 vs. 608 ± 146 HU/mm, p < 0.01) than EID-CT series. Two radiologists considered the delineation of bone details as feasible at consistently lower CTDI(Vol) values at PCD-CT than at EID-CT. In comparison of SNR-matched reconstructions, PCD-CT series were still considered superior in almost all cases. CONCLUSIONS: In this head-to-head comparison, PCD-CT showed superior objective and subjective image quality characteristics over EID-CT for the delineation of tiniest bone details. Even in SNR-matched pairs (acquired at different CTDI(Vol)’s), PCD-CT was strongly preferred by radiologists. KEY POINTS: • In dose-matched scans, photon-counting detector CT series showed significantly less image noise, higher signal-to-noise ratio, and higher edge sharpness than energy-integrating detector CT series. • Human observers considered the delineation of tiny bone details as feasible at much lower dose levels in photon-counting detector CT than in energy-integrating detector CT. • In direct comparison of series matched for signal-to-noise ratio, photon-counting detector CT series were considered superior in almost all cases. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s00330-021-08441-4. Springer Berlin Heidelberg 2021-12-22 2022 /pmc/articles/PMC9038873/ /pubmed/34936011 http://dx.doi.org/10.1007/s00330-021-08441-4 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Computed Tomography Bette, Stefanie J. Braun, Franziska M. Haerting, Mark Decker, Josua A. Luitjens, Jan H. Scheurig-Muenkler, Christian Kroencke, Thomas J. Schwarz, Florian Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT |
title | Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT |
title_full | Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT |
title_fullStr | Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT |
title_full_unstemmed | Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT |
title_short | Visualization of bone details in a novel photon-counting dual-source CT scanner—comparison with energy-integrating CT |
title_sort | visualization of bone details in a novel photon-counting dual-source ct scanner—comparison with energy-integrating ct |
topic | Computed Tomography |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9038873/ https://www.ncbi.nlm.nih.gov/pubmed/34936011 http://dx.doi.org/10.1007/s00330-021-08441-4 |
work_keys_str_mv | AT bettestefaniej visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct AT braunfranziskam visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct AT haertingmark visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct AT deckerjosuaa visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct AT luitjensjanh visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct AT scheurigmuenklerchristian visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct AT kroenckethomasj visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct AT schwarzflorian visualizationofbonedetailsinanovelphotoncountingdualsourcectscannercomparisonwithenergyintegratingct |