Precision of low-dose CT-based micromotion analysis technique for the assessment of early acetabular cup migration compared with gold standard RSA: a prospective study of 30 patients up to 1 year
BACKGROUND AND PURPOSE: Computed tomography micromotion analysis (CTMA) can be used to determine implant micro-movements using low-dose CT scans. By using CTMA, a non-invasive measurement of joint implant movement is enabled. We evaluated the precision of CTMA in measuring early cup migration. Stand...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medical Journals Sweden, on behalf of the Nordic Orthopedic Federation
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9047498/ https://www.ncbi.nlm.nih.gov/pubmed/35478261 http://dx.doi.org/10.2340/17453674.2022.2528 |
_version_ | 1784695740349546496 |
---|---|
author | ANGELOMENOS, Vasileios MOHADDES, Maziar ITAYEM, Raed SHAREGHI, Bita |
author_facet | ANGELOMENOS, Vasileios MOHADDES, Maziar ITAYEM, Raed SHAREGHI, Bita |
author_sort | ANGELOMENOS, Vasileios |
collection | PubMed |
description | BACKGROUND AND PURPOSE: Computed tomography micromotion analysis (CTMA) can be used to determine implant micro-movements using low-dose CT scans. By using CTMA, a non-invasive measurement of joint implant movement is enabled. We evaluated the precision of CTMA in measuring early cup migration. Standard marker-based radiostereometric analysis (RSA) was used as reference. We hypothesised that CTMA can be used as an alternative to RSA in assessing implant micromotions. PATIENTS AND METHODS: We included 30 patients undergoing total hip arthroplasty (THA). Acetabular cup migration at 1 year was measured with RSA and CTMA. To determine the precision of both methods, 20 double examinations (postoperatively) with repositioning of the patients were performed. The precision was calculated from zero by assuming that there was no motion of the prosthesis between the 2 examinations. RESULTS: The precision of RSA ranged from 0.06 to 0.15 mm for translations and 0.21° to 0.63° for rotations. Corresponding values for CTMA were 0.06 to 0.13 mm and 0.23° to 0.35°. A good level of agreement was found between the methods regarding cup migration and rotation at 1 year. INTERPRETATION: The precision of CTMA in measuring acetabular cup migration and rotation is comparable to marker-based RSA. CTMA could possibly thus be used as an alternative method to detect early implant migration. |
format | Online Article Text |
id | pubmed-9047498 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Medical Journals Sweden, on behalf of the Nordic Orthopedic Federation |
record_format | MEDLINE/PubMed |
spelling | pubmed-90474982022-04-29 Precision of low-dose CT-based micromotion analysis technique for the assessment of early acetabular cup migration compared with gold standard RSA: a prospective study of 30 patients up to 1 year ANGELOMENOS, Vasileios MOHADDES, Maziar ITAYEM, Raed SHAREGHI, Bita Acta Orthop Article BACKGROUND AND PURPOSE: Computed tomography micromotion analysis (CTMA) can be used to determine implant micro-movements using low-dose CT scans. By using CTMA, a non-invasive measurement of joint implant movement is enabled. We evaluated the precision of CTMA in measuring early cup migration. Standard marker-based radiostereometric analysis (RSA) was used as reference. We hypothesised that CTMA can be used as an alternative to RSA in assessing implant micromotions. PATIENTS AND METHODS: We included 30 patients undergoing total hip arthroplasty (THA). Acetabular cup migration at 1 year was measured with RSA and CTMA. To determine the precision of both methods, 20 double examinations (postoperatively) with repositioning of the patients were performed. The precision was calculated from zero by assuming that there was no motion of the prosthesis between the 2 examinations. RESULTS: The precision of RSA ranged from 0.06 to 0.15 mm for translations and 0.21° to 0.63° for rotations. Corresponding values for CTMA were 0.06 to 0.13 mm and 0.23° to 0.35°. A good level of agreement was found between the methods regarding cup migration and rotation at 1 year. INTERPRETATION: The precision of CTMA in measuring acetabular cup migration and rotation is comparable to marker-based RSA. CTMA could possibly thus be used as an alternative method to detect early implant migration. Medical Journals Sweden, on behalf of the Nordic Orthopedic Federation 2022-04-22 /pmc/articles/PMC9047498/ /pubmed/35478261 http://dx.doi.org/10.2340/17453674.2022.2528 Text en © 2022 The Author(s) https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (https://creativecommons.org/licenses/by-nc/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material for non-commercial purposes, provided proper attribution to the original work. |
spellingShingle | Article ANGELOMENOS, Vasileios MOHADDES, Maziar ITAYEM, Raed SHAREGHI, Bita Precision of low-dose CT-based micromotion analysis technique for the assessment of early acetabular cup migration compared with gold standard RSA: a prospective study of 30 patients up to 1 year |
title | Precision of low-dose CT-based micromotion analysis technique for the assessment of early acetabular cup migration compared with gold standard RSA: a prospective study of 30 patients up to 1 year |
title_full | Precision of low-dose CT-based micromotion analysis technique for the assessment of early acetabular cup migration compared with gold standard RSA: a prospective study of 30 patients up to 1 year |
title_fullStr | Precision of low-dose CT-based micromotion analysis technique for the assessment of early acetabular cup migration compared with gold standard RSA: a prospective study of 30 patients up to 1 year |
title_full_unstemmed | Precision of low-dose CT-based micromotion analysis technique for the assessment of early acetabular cup migration compared with gold standard RSA: a prospective study of 30 patients up to 1 year |
title_short | Precision of low-dose CT-based micromotion analysis technique for the assessment of early acetabular cup migration compared with gold standard RSA: a prospective study of 30 patients up to 1 year |
title_sort | precision of low-dose ct-based micromotion analysis technique for the assessment of early acetabular cup migration compared with gold standard rsa: a prospective study of 30 patients up to 1 year |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9047498/ https://www.ncbi.nlm.nih.gov/pubmed/35478261 http://dx.doi.org/10.2340/17453674.2022.2528 |
work_keys_str_mv | AT angelomenosvasileios precisionoflowdosectbasedmicromotionanalysistechniquefortheassessmentofearlyacetabularcupmigrationcomparedwithgoldstandardrsaaprospectivestudyof30patientsupto1year AT mohaddesmaziar precisionoflowdosectbasedmicromotionanalysistechniquefortheassessmentofearlyacetabularcupmigrationcomparedwithgoldstandardrsaaprospectivestudyof30patientsupto1year AT itayemraed precisionoflowdosectbasedmicromotionanalysistechniquefortheassessmentofearlyacetabularcupmigrationcomparedwithgoldstandardrsaaprospectivestudyof30patientsupto1year AT shareghibita precisionoflowdosectbasedmicromotionanalysistechniquefortheassessmentofearlyacetabularcupmigrationcomparedwithgoldstandardrsaaprospectivestudyof30patientsupto1year |