Unreliable Automated Complete Blood Count Results: Causes, Recognition, and Resolution

Automated hematology analyzers generate accurate complete blood counts (CBC) results on nearly all specimens. However, every laboratory encounters, at times, some specimens that yield no or inaccurate result(s) for one or more CBC parameters even when the analyzer is functioning properly and the man...

Descripción completa

Detalles Bibliográficos
Autores principales: Gulati, Gene, Uppal, Guldeep, Gong, Jerald
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Society for Laboratory Medicine 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9057813/
https://www.ncbi.nlm.nih.gov/pubmed/35470271
http://dx.doi.org/10.3343/alm.2022.42.5.515
_version_ 1784697986731737088
author Gulati, Gene
Uppal, Guldeep
Gong, Jerald
author_facet Gulati, Gene
Uppal, Guldeep
Gong, Jerald
author_sort Gulati, Gene
collection PubMed
description Automated hematology analyzers generate accurate complete blood counts (CBC) results on nearly all specimens. However, every laboratory encounters, at times, some specimens that yield no or inaccurate result(s) for one or more CBC parameters even when the analyzer is functioning properly and the manufacturer’s instructions are followed to the letter. Inaccurate results, which may adversely affect patient care, are clinically unreliable and require the attention of laboratory professionals. Laboratory professionals must recognize unreliable results, determine the possible cause(s), and be acquainted with the ways to obtain reliable results on such specimens. We present a concise overview of the known causes of unreliable automated CBC results, ways to recognize them, and means commonly utilized to obtain reliable results. Some examples of unreliable automated CBC results are also illustrated. Pertinent analyzer-specific information can be found in the manufacturers’ operating manuals.
format Online
Article
Text
id pubmed-9057813
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Korean Society for Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-90578132022-09-01 Unreliable Automated Complete Blood Count Results: Causes, Recognition, and Resolution Gulati, Gene Uppal, Guldeep Gong, Jerald Ann Lab Med Review Article Automated hematology analyzers generate accurate complete blood counts (CBC) results on nearly all specimens. However, every laboratory encounters, at times, some specimens that yield no or inaccurate result(s) for one or more CBC parameters even when the analyzer is functioning properly and the manufacturer’s instructions are followed to the letter. Inaccurate results, which may adversely affect patient care, are clinically unreliable and require the attention of laboratory professionals. Laboratory professionals must recognize unreliable results, determine the possible cause(s), and be acquainted with the ways to obtain reliable results on such specimens. We present a concise overview of the known causes of unreliable automated CBC results, ways to recognize them, and means commonly utilized to obtain reliable results. Some examples of unreliable automated CBC results are also illustrated. Pertinent analyzer-specific information can be found in the manufacturers’ operating manuals. Korean Society for Laboratory Medicine 2022-09-01 2022-09-01 /pmc/articles/PMC9057813/ /pubmed/35470271 http://dx.doi.org/10.3343/alm.2022.42.5.515 Text en © Korean Society for Laboratory Medicine https://creativecommons.org/licenses/by-nc/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0 (https://creativecommons.org/licenses/by-nc/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Review Article
Gulati, Gene
Uppal, Guldeep
Gong, Jerald
Unreliable Automated Complete Blood Count Results: Causes, Recognition, and Resolution
title Unreliable Automated Complete Blood Count Results: Causes, Recognition, and Resolution
title_full Unreliable Automated Complete Blood Count Results: Causes, Recognition, and Resolution
title_fullStr Unreliable Automated Complete Blood Count Results: Causes, Recognition, and Resolution
title_full_unstemmed Unreliable Automated Complete Blood Count Results: Causes, Recognition, and Resolution
title_short Unreliable Automated Complete Blood Count Results: Causes, Recognition, and Resolution
title_sort unreliable automated complete blood count results: causes, recognition, and resolution
topic Review Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9057813/
https://www.ncbi.nlm.nih.gov/pubmed/35470271
http://dx.doi.org/10.3343/alm.2022.42.5.515
work_keys_str_mv AT gulatigene unreliableautomatedcompletebloodcountresultscausesrecognitionandresolution
AT uppalguldeep unreliableautomatedcompletebloodcountresultscausesrecognitionandresolution
AT gongjerald unreliableautomatedcompletebloodcountresultscausesrecognitionandresolution