Cargando…
Comparative quantification of focal and diffuse visual field loss by the SPARK Precision threshold algorithm and SITA
PURPOSE: The aims of this paper were to examine focal and diffuse visual field loss in terms of threshold agreement between the widely used SITA Standard Humphrey Field Analyser (HFA) threshold algorithm with the SPARK Precision algorithm (Oculus Twinfield 2). METHODS: A total of 39 treated glaucoma...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9061659/ https://www.ncbi.nlm.nih.gov/pubmed/34962591 http://dx.doi.org/10.1007/s00417-021-05430-7 |
_version_ | 1784698773176320000 |
---|---|
author | Foo, S. K. Cubbidge, R. P. Heitmar, R. |
author_facet | Foo, S. K. Cubbidge, R. P. Heitmar, R. |
author_sort | Foo, S. K. |
collection | PubMed |
description | PURPOSE: The aims of this paper were to examine focal and diffuse visual field loss in terms of threshold agreement between the widely used SITA Standard Humphrey Field Analyser (HFA) threshold algorithm with the SPARK Precision algorithm (Oculus Twinfield 2). METHODS: A total of 39 treated glaucoma patients (34 primary open angle and 5 primary angle closure glaucoma) and 31 cataract patients without glaucoma were tested in succession with the Oculus Twinfield 2 (Oculus Optikgeräte GmbH, Wetzlar, Germany) using the SPARK Precision algorithm and with the HFA 3 (Carl Zeiss Meditec, Dublin, CA) using the 30–2 SITA Standard algorithm. RESULTS: SPARK Precision required around half the testing time of SITA Standard. There was a good correlation between the MS of the two threshold algorithms but MD and PSD were significantly less severe with SPARK Precision in both glaucoma (focal field loss) and cataract (diffuse field loss) groups (p < 0.001). There was poor agreement for all global indices (MS, MD and PSD) between the two algorithms and there was a significant proportional bias of MD in the glaucoma group and PSD in both glaucoma and cataract groups. The pointwise sensitivity analysis yielded higher threshold estimates in SPARK Precision than in SITA Standard in the nasal field. Classification of glaucoma severity using AGIS was significantly lower with SPARK Precision compared to SITA Standard (p < 0.001). CONCLUSION: SITA renders deeper defects than SPARK. Compared to the SITA Standard threshold algorithm, SPARK Precision cannot quantify early glaucomatous field loss. This may be due to the mathematical linear interpolation of threshold sensitivity or deeper scotomas due to the plateau effect caused by the reduced dynamic range of the Twinfield 2 perimeter. Although not of clinical significance in early glaucoma, the plateau effect may hinder the long-term follow-up of patients during disease progression. |
format | Online Article Text |
id | pubmed-9061659 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Springer Berlin Heidelberg |
record_format | MEDLINE/PubMed |
spelling | pubmed-90616592022-05-07 Comparative quantification of focal and diffuse visual field loss by the SPARK Precision threshold algorithm and SITA Foo, S. K. Cubbidge, R. P. Heitmar, R. Graefes Arch Clin Exp Ophthalmol Glaucoma PURPOSE: The aims of this paper were to examine focal and diffuse visual field loss in terms of threshold agreement between the widely used SITA Standard Humphrey Field Analyser (HFA) threshold algorithm with the SPARK Precision algorithm (Oculus Twinfield 2). METHODS: A total of 39 treated glaucoma patients (34 primary open angle and 5 primary angle closure glaucoma) and 31 cataract patients without glaucoma were tested in succession with the Oculus Twinfield 2 (Oculus Optikgeräte GmbH, Wetzlar, Germany) using the SPARK Precision algorithm and with the HFA 3 (Carl Zeiss Meditec, Dublin, CA) using the 30–2 SITA Standard algorithm. RESULTS: SPARK Precision required around half the testing time of SITA Standard. There was a good correlation between the MS of the two threshold algorithms but MD and PSD were significantly less severe with SPARK Precision in both glaucoma (focal field loss) and cataract (diffuse field loss) groups (p < 0.001). There was poor agreement for all global indices (MS, MD and PSD) between the two algorithms and there was a significant proportional bias of MD in the glaucoma group and PSD in both glaucoma and cataract groups. The pointwise sensitivity analysis yielded higher threshold estimates in SPARK Precision than in SITA Standard in the nasal field. Classification of glaucoma severity using AGIS was significantly lower with SPARK Precision compared to SITA Standard (p < 0.001). CONCLUSION: SITA renders deeper defects than SPARK. Compared to the SITA Standard threshold algorithm, SPARK Precision cannot quantify early glaucomatous field loss. This may be due to the mathematical linear interpolation of threshold sensitivity or deeper scotomas due to the plateau effect caused by the reduced dynamic range of the Twinfield 2 perimeter. Although not of clinical significance in early glaucoma, the plateau effect may hinder the long-term follow-up of patients during disease progression. Springer Berlin Heidelberg 2021-12-28 2022 /pmc/articles/PMC9061659/ /pubmed/34962591 http://dx.doi.org/10.1007/s00417-021-05430-7 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Glaucoma Foo, S. K. Cubbidge, R. P. Heitmar, R. Comparative quantification of focal and diffuse visual field loss by the SPARK Precision threshold algorithm and SITA |
title | Comparative quantification of focal and diffuse visual field loss by the SPARK Precision threshold algorithm and SITA |
title_full | Comparative quantification of focal and diffuse visual field loss by the SPARK Precision threshold algorithm and SITA |
title_fullStr | Comparative quantification of focal and diffuse visual field loss by the SPARK Precision threshold algorithm and SITA |
title_full_unstemmed | Comparative quantification of focal and diffuse visual field loss by the SPARK Precision threshold algorithm and SITA |
title_short | Comparative quantification of focal and diffuse visual field loss by the SPARK Precision threshold algorithm and SITA |
title_sort | comparative quantification of focal and diffuse visual field loss by the spark precision threshold algorithm and sita |
topic | Glaucoma |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9061659/ https://www.ncbi.nlm.nih.gov/pubmed/34962591 http://dx.doi.org/10.1007/s00417-021-05430-7 |
work_keys_str_mv | AT foosk comparativequantificationoffocalanddiffusevisualfieldlossbythesparkprecisionthresholdalgorithmandsita AT cubbidgerp comparativequantificationoffocalanddiffusevisualfieldlossbythesparkprecisionthresholdalgorithmandsita AT heitmarr comparativequantificationoffocalanddiffusevisualfieldlossbythesparkprecisionthresholdalgorithmandsita |