Heartmate 3 implantation in small patients: CT-guided chest diameter assessment

In recent years, the Heartmate 3 (HM3) has largely replaced the use of other intracorporeal left ventricular assist devices in the adult field. Because the HM3 is larger than the Heartware Ventricular Assist Device, the general consensus was that for small patients, the Heartware Ventricular Assist...

Descripción completa

Detalles Bibliográficos
Autores principales: Van Puyvelde, Joeri, Jacobs, Steven, Vlasselaers, Dirk, Meyns, Bart
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9070508/
https://www.ncbi.nlm.nih.gov/pubmed/35134163
http://dx.doi.org/10.1093/icvts/ivac009
Descripción
Sumario:In recent years, the Heartmate 3 (HM3) has largely replaced the use of other intracorporeal left ventricular assist devices in the adult field. Because the HM3 is larger than the Heartware Ventricular Assist Device, the general consensus was that for small patients, the Heartware Ventricular Assist Device was the most appropriate implantable device option. Our goal was to describe our experiences with the successful implantation of the HM3 in 2 children, aged 9 and 11. We report on the chest cavity dimensions, as measured on computed tomography, that can be used to assess the feasibility of HM3 implantation in small patients.