Level of Agreement and Correlation Between the Estimated Hemoglobin A1c Results Derived by Continuous or Conventional Glucose Monitoring Systems Compared with the Point-of-Care or Laboratory-Based Measurements: An Observational Study

INTRODUCTION: Hemoglobin A1C (HbA1c) is an important marker for diabetes care management. With the increasing use of new technologies such as continuous glucose monitoring (CGM) and point-of-care testing (POCT), patients and their physicians have been able to monitor and continuously check their blo...

Descripción completa

Detalles Bibliográficos
Autores principales: Al Hayek, Ayman A., Sobki, Samia H., Al-Saeed, Abdulghani H., Alzahrani, Wael M., Al Dawish, Mohamed A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Healthcare 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9076797/
https://www.ncbi.nlm.nih.gov/pubmed/35306640
http://dx.doi.org/10.1007/s13300-022-01240-0
_version_ 1784702003568443392
author Al Hayek, Ayman A.
Sobki, Samia H.
Al-Saeed, Abdulghani H.
Alzahrani, Wael M.
Al Dawish, Mohamed A.
author_facet Al Hayek, Ayman A.
Sobki, Samia H.
Al-Saeed, Abdulghani H.
Alzahrani, Wael M.
Al Dawish, Mohamed A.
author_sort Al Hayek, Ayman A.
collection PubMed
description INTRODUCTION: Hemoglobin A1C (HbA1c) is an important marker for diabetes care management. With the increasing use of new technologies such as continuous glucose monitoring (CGM) and point-of-care testing (POCT), patients and their physicians have been able to monitor and continuously check their blood glucose levels in an efficient and timely manner. This study aimed to investigate the level of agreement between the standard laboratory test for HbA1c (Lab-HbA1c) with point-of-care testing (POCT-HbA1c) and glucose monitoring index (GMI) derived by intermittently scanned CGM (isCGM) or estimated average glucose (eAG) derived by conventional self-monitored blood glucose (SMBG) devices. METHODS: A cross-sectional study was conducted at the Diabetes Treatment Center, Prince Sultan Military Medical City, Saudi Arabia, between May and December 2020 with 81 patients with diabetes who used the isCGM system (n = 30) or conventional finger-pricking SMBG system (n = 51). At the same visit, venous and capillary blood samples were taken for routine HbA1c analysis by the standard laboratory and POCT methods, respectively. Also, for isCGM users, the GMI data for 28 days (GMI-28) and 90 days (GMI-90) were obtained, while for SMBG users, eAG data for 30 days (eAG-30) and 90 days (eAG-90) were calculated. The limits of agreement in different HbA1c measurements were evaluated using a Bland-Altman analysis. Pearson correlation and multivariate linear regression analyses were also performed. RESULTS: Based on the Bland-Altman analysis, HbA1c levels for 96.7% and 96.1% of the patients analyzed by the POCT and the standard laboratory methods were within the range of the 95% limit of agreement in both isCGM and conventional SMBG users, respectively. About 93.3% of the GMI measurements were within the 95% limit of agreement. Also, about 94.12% of the eAG-30 and 90.2% of the eAG-90 measurements were within the 95% limit of agreement. Moreover, the correlation analysis revealed a statistically significant positive correlation and linear regression among Lab-HbA1c, POCT-HbA1c, GMI, and eAG in both conventional SMBG and isCGM users (all p < 0.001). These positive results persisted significantly after adjusting for different factors (all p < 0.001). CONCLUSION: GMI derived by isCGM or eAG derived by conventional SMBG systems, as well as the POCT-HbA1c measurements, showed a high level of agreement; therefore, we recommend them as potential methods for diabetes monitoring, especially when a rapid result is needed or with patients with uncontrolled diabetes or on intensive insulin therapy.
format Online
Article
Text
id pubmed-9076797
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Springer Healthcare
record_format MEDLINE/PubMed
spelling pubmed-90767972022-05-08 Level of Agreement and Correlation Between the Estimated Hemoglobin A1c Results Derived by Continuous or Conventional Glucose Monitoring Systems Compared with the Point-of-Care or Laboratory-Based Measurements: An Observational Study Al Hayek, Ayman A. Sobki, Samia H. Al-Saeed, Abdulghani H. Alzahrani, Wael M. Al Dawish, Mohamed A. Diabetes Ther Original Research INTRODUCTION: Hemoglobin A1C (HbA1c) is an important marker for diabetes care management. With the increasing use of new technologies such as continuous glucose monitoring (CGM) and point-of-care testing (POCT), patients and their physicians have been able to monitor and continuously check their blood glucose levels in an efficient and timely manner. This study aimed to investigate the level of agreement between the standard laboratory test for HbA1c (Lab-HbA1c) with point-of-care testing (POCT-HbA1c) and glucose monitoring index (GMI) derived by intermittently scanned CGM (isCGM) or estimated average glucose (eAG) derived by conventional self-monitored blood glucose (SMBG) devices. METHODS: A cross-sectional study was conducted at the Diabetes Treatment Center, Prince Sultan Military Medical City, Saudi Arabia, between May and December 2020 with 81 patients with diabetes who used the isCGM system (n = 30) or conventional finger-pricking SMBG system (n = 51). At the same visit, venous and capillary blood samples were taken for routine HbA1c analysis by the standard laboratory and POCT methods, respectively. Also, for isCGM users, the GMI data for 28 days (GMI-28) and 90 days (GMI-90) were obtained, while for SMBG users, eAG data for 30 days (eAG-30) and 90 days (eAG-90) were calculated. The limits of agreement in different HbA1c measurements were evaluated using a Bland-Altman analysis. Pearson correlation and multivariate linear regression analyses were also performed. RESULTS: Based on the Bland-Altman analysis, HbA1c levels for 96.7% and 96.1% of the patients analyzed by the POCT and the standard laboratory methods were within the range of the 95% limit of agreement in both isCGM and conventional SMBG users, respectively. About 93.3% of the GMI measurements were within the 95% limit of agreement. Also, about 94.12% of the eAG-30 and 90.2% of the eAG-90 measurements were within the 95% limit of agreement. Moreover, the correlation analysis revealed a statistically significant positive correlation and linear regression among Lab-HbA1c, POCT-HbA1c, GMI, and eAG in both conventional SMBG and isCGM users (all p < 0.001). These positive results persisted significantly after adjusting for different factors (all p < 0.001). CONCLUSION: GMI derived by isCGM or eAG derived by conventional SMBG systems, as well as the POCT-HbA1c measurements, showed a high level of agreement; therefore, we recommend them as potential methods for diabetes monitoring, especially when a rapid result is needed or with patients with uncontrolled diabetes or on intensive insulin therapy. Springer Healthcare 2022-03-20 2022-05 /pmc/articles/PMC9076797/ /pubmed/35306640 http://dx.doi.org/10.1007/s13300-022-01240-0 Text en © The Author(s) 2022 https://creativecommons.org/licenses/by-nc/4.0/Open AccessThis article is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License, which permits any non-commercial use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) .
spellingShingle Original Research
Al Hayek, Ayman A.
Sobki, Samia H.
Al-Saeed, Abdulghani H.
Alzahrani, Wael M.
Al Dawish, Mohamed A.
Level of Agreement and Correlation Between the Estimated Hemoglobin A1c Results Derived by Continuous or Conventional Glucose Monitoring Systems Compared with the Point-of-Care or Laboratory-Based Measurements: An Observational Study
title Level of Agreement and Correlation Between the Estimated Hemoglobin A1c Results Derived by Continuous or Conventional Glucose Monitoring Systems Compared with the Point-of-Care or Laboratory-Based Measurements: An Observational Study
title_full Level of Agreement and Correlation Between the Estimated Hemoglobin A1c Results Derived by Continuous or Conventional Glucose Monitoring Systems Compared with the Point-of-Care or Laboratory-Based Measurements: An Observational Study
title_fullStr Level of Agreement and Correlation Between the Estimated Hemoglobin A1c Results Derived by Continuous or Conventional Glucose Monitoring Systems Compared with the Point-of-Care or Laboratory-Based Measurements: An Observational Study
title_full_unstemmed Level of Agreement and Correlation Between the Estimated Hemoglobin A1c Results Derived by Continuous or Conventional Glucose Monitoring Systems Compared with the Point-of-Care or Laboratory-Based Measurements: An Observational Study
title_short Level of Agreement and Correlation Between the Estimated Hemoglobin A1c Results Derived by Continuous or Conventional Glucose Monitoring Systems Compared with the Point-of-Care or Laboratory-Based Measurements: An Observational Study
title_sort level of agreement and correlation between the estimated hemoglobin a1c results derived by continuous or conventional glucose monitoring systems compared with the point-of-care or laboratory-based measurements: an observational study
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9076797/
https://www.ncbi.nlm.nih.gov/pubmed/35306640
http://dx.doi.org/10.1007/s13300-022-01240-0
work_keys_str_mv AT alhayekaymana levelofagreementandcorrelationbetweentheestimatedhemoglobina1cresultsderivedbycontinuousorconventionalglucosemonitoringsystemscomparedwiththepointofcareorlaboratorybasedmeasurementsanobservationalstudy
AT sobkisamiah levelofagreementandcorrelationbetweentheestimatedhemoglobina1cresultsderivedbycontinuousorconventionalglucosemonitoringsystemscomparedwiththepointofcareorlaboratorybasedmeasurementsanobservationalstudy
AT alsaeedabdulghanih levelofagreementandcorrelationbetweentheestimatedhemoglobina1cresultsderivedbycontinuousorconventionalglucosemonitoringsystemscomparedwiththepointofcareorlaboratorybasedmeasurementsanobservationalstudy
AT alzahraniwaelm levelofagreementandcorrelationbetweentheestimatedhemoglobina1cresultsderivedbycontinuousorconventionalglucosemonitoringsystemscomparedwiththepointofcareorlaboratorybasedmeasurementsanobservationalstudy
AT aldawishmohameda levelofagreementandcorrelationbetweentheestimatedhemoglobina1cresultsderivedbycontinuousorconventionalglucosemonitoringsystemscomparedwiththepointofcareorlaboratorybasedmeasurementsanobservationalstudy