Implementation and Validation of Anisotropic Analytical Algorithm in Eclipse Treatment Planning System for Indigenous Telecobalt Machine (Bhabhatron II)
BACKGROUND: The photon dose calculation model anisotropic analytical algorithm (AAA) available with eclipse integrated treatment planning system (TPS) (Varian) supports telecobalt dose calculation from Version 13.6 onward. Formerly, pencil beam convolution (PBC) was used for modeling telecobalt mach...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9084584/ https://www.ncbi.nlm.nih.gov/pubmed/35548036 http://dx.doi.org/10.4103/jmp.jmp_95_21 |
_version_ | 1784703640739512320 |
---|---|
author | Hajare, Raghavendra Ali, Md. Basit Sreelakshmi, K K Kumar, M Anil Kalita, Rituraj Mahantshetty, Umesh M |
author_facet | Hajare, Raghavendra Ali, Md. Basit Sreelakshmi, K K Kumar, M Anil Kalita, Rituraj Mahantshetty, Umesh M |
author_sort | Hajare, Raghavendra |
collection | PubMed |
description | BACKGROUND: The photon dose calculation model anisotropic analytical algorithm (AAA) available with eclipse integrated treatment planning system (TPS) (Varian) supports telecobalt dose calculation from Version 13.6 onward. Formerly, pencil beam convolution (PBC) was used for modeling telecobalt machines. Eclipse TPS no longer supports PBC dose calculation algorithm in v13.6 and above. The AAA dose calculation model is a three-dimensional PBC/superposition algorithm. Its configuration is based on Monte-Carlo-determined basic physical parameters that are tailored to measured clinical beam data. AIM: The study investigated the feasibility of clinical implementation of AAA in Eclipse TPS for Bhabhatron II. MATERIALS AND METHODS: The indigenous telecobalt machine, Bhabhatron II, was configured as a generic machine because an inbuilt machine model for the same was not available in Varian Eclipse TPS algorithm library. In such a scenario, it was necessary to evaluate and validate dosimetric parameters of the TPS because improper tailoring would cause errors in dose calculations. Beam data measurements of the machine were carried out which were used for configuration of the algorithm. RESULT: After successful configuration, a variety of plans created in TPS were executed on the machine and subsequently evaluated. CONCLUSION: From this study, we concluded that AAA-based dose calculation in TPS is very well suited for accurate dose calculations for telecobalt machine and can be implemented for clinical use. |
format | Online Article Text |
id | pubmed-9084584 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Wolters Kluwer - Medknow |
record_format | MEDLINE/PubMed |
spelling | pubmed-90845842022-05-10 Implementation and Validation of Anisotropic Analytical Algorithm in Eclipse Treatment Planning System for Indigenous Telecobalt Machine (Bhabhatron II) Hajare, Raghavendra Ali, Md. Basit Sreelakshmi, K K Kumar, M Anil Kalita, Rituraj Mahantshetty, Umesh M J Med Phys Original Article BACKGROUND: The photon dose calculation model anisotropic analytical algorithm (AAA) available with eclipse integrated treatment planning system (TPS) (Varian) supports telecobalt dose calculation from Version 13.6 onward. Formerly, pencil beam convolution (PBC) was used for modeling telecobalt machines. Eclipse TPS no longer supports PBC dose calculation algorithm in v13.6 and above. The AAA dose calculation model is a three-dimensional PBC/superposition algorithm. Its configuration is based on Monte-Carlo-determined basic physical parameters that are tailored to measured clinical beam data. AIM: The study investigated the feasibility of clinical implementation of AAA in Eclipse TPS for Bhabhatron II. MATERIALS AND METHODS: The indigenous telecobalt machine, Bhabhatron II, was configured as a generic machine because an inbuilt machine model for the same was not available in Varian Eclipse TPS algorithm library. In such a scenario, it was necessary to evaluate and validate dosimetric parameters of the TPS because improper tailoring would cause errors in dose calculations. Beam data measurements of the machine were carried out which were used for configuration of the algorithm. RESULT: After successful configuration, a variety of plans created in TPS were executed on the machine and subsequently evaluated. CONCLUSION: From this study, we concluded that AAA-based dose calculation in TPS is very well suited for accurate dose calculations for telecobalt machine and can be implemented for clinical use. Wolters Kluwer - Medknow 2022 2022-03-31 /pmc/articles/PMC9084584/ /pubmed/35548036 http://dx.doi.org/10.4103/jmp.jmp_95_21 Text en Copyright: © 2022 Journal of Medical Physics https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Original Article Hajare, Raghavendra Ali, Md. Basit Sreelakshmi, K K Kumar, M Anil Kalita, Rituraj Mahantshetty, Umesh M Implementation and Validation of Anisotropic Analytical Algorithm in Eclipse Treatment Planning System for Indigenous Telecobalt Machine (Bhabhatron II) |
title | Implementation and Validation of Anisotropic Analytical Algorithm in Eclipse Treatment Planning System for Indigenous Telecobalt Machine (Bhabhatron II) |
title_full | Implementation and Validation of Anisotropic Analytical Algorithm in Eclipse Treatment Planning System for Indigenous Telecobalt Machine (Bhabhatron II) |
title_fullStr | Implementation and Validation of Anisotropic Analytical Algorithm in Eclipse Treatment Planning System for Indigenous Telecobalt Machine (Bhabhatron II) |
title_full_unstemmed | Implementation and Validation of Anisotropic Analytical Algorithm in Eclipse Treatment Planning System for Indigenous Telecobalt Machine (Bhabhatron II) |
title_short | Implementation and Validation of Anisotropic Analytical Algorithm in Eclipse Treatment Planning System for Indigenous Telecobalt Machine (Bhabhatron II) |
title_sort | implementation and validation of anisotropic analytical algorithm in eclipse treatment planning system for indigenous telecobalt machine (bhabhatron ii) |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9084584/ https://www.ncbi.nlm.nih.gov/pubmed/35548036 http://dx.doi.org/10.4103/jmp.jmp_95_21 |
work_keys_str_mv | AT hajareraghavendra implementationandvalidationofanisotropicanalyticalalgorithmineclipsetreatmentplanningsystemforindigenoustelecobaltmachinebhabhatronii AT alimdbasit implementationandvalidationofanisotropicanalyticalalgorithmineclipsetreatmentplanningsystemforindigenoustelecobaltmachinebhabhatronii AT sreelakshmikk implementationandvalidationofanisotropicanalyticalalgorithmineclipsetreatmentplanningsystemforindigenoustelecobaltmachinebhabhatronii AT kumarmanil implementationandvalidationofanisotropicanalyticalalgorithmineclipsetreatmentplanningsystemforindigenoustelecobaltmachinebhabhatronii AT kalitarituraj implementationandvalidationofanisotropicanalyticalalgorithmineclipsetreatmentplanningsystemforindigenoustelecobaltmachinebhabhatronii AT mahantshettyumeshm implementationandvalidationofanisotropicanalyticalalgorithmineclipsetreatmentplanningsystemforindigenoustelecobaltmachinebhabhatronii |