Cargando…
Importance of Observing the Progress Curve During Enzyme Assay in an Automated Clinical Chemistry Analyzer: A Case Study
In the case presented here the reported result for total creatine phosphokinase led to the physician calling for report confirmation. The repeated test result was in keeping with the clinical picture and thus the previous erroneous result was amended. The incorrect result from auto analyzer was iden...
Autor principal: | Pant, Vivek |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Communications and Publications Division (CPD) of the IFCC
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9092719/ https://www.ncbi.nlm.nih.gov/pubmed/35645694 |
Ejemplares similares
-
Performance evaluation of enzyme immunoassay for voriconazole therapeutic drug monitoring with automated clinical chemistry analyzers
por: Jeon, Yongbum, et al.
Publicado: (2017) -
Validation of a novel and accurate ApoE4 assay for automated chemistry analyzers
por: Veiga, Sergio, et al.
Publicado: (2020) -
Automation in Clinical Chemistry
por: Polesky, Herbert F.
Publicado: (1964) -
Kinetic modelling: an integrated approach to analyze enzyme activity assays
por: Boeckx, Jelena, et al.
Publicado: (2017) -
Evaluation of a species-specific C-reactive protein assay for the dog on the ABX Pentra 400 clinical chemistry analyzer
por: Hindenberg, Sarah, et al.
Publicado: (2017)