Dentine sensitivity caused by illumination of intraoral scanner and light curing unit
BACKGROUND: Patients often compliant sensitivity to high-intensity light irradiated application during dental procedures. This study aims to investigate tooth sensitivity caused by high-intensity light irradiated from an intraoral scanner (IOS) and a light-curing unit (LCU). MATERIAL AND METHODS: Fo...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medicina Oral S.L.
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9094722/ https://www.ncbi.nlm.nih.gov/pubmed/35582349 http://dx.doi.org/10.4317/jced.59251 |
Sumario: | BACKGROUND: Patients often compliant sensitivity to high-intensity light irradiated application during dental procedures. This study aims to investigate tooth sensitivity caused by high-intensity light irradiated from an intraoral scanner (IOS) and a light-curing unit (LCU). MATERIAL AND METHODS: Forty-five teeth from 45 healthy volunteers were included. These were equally classified into three groups using the cold test (4 ± 1 °C) and NRS pain assessment; A=control, B=cavity without sensitivity, and C=cavity with sensitivity. Two thermocouple probes were attached to the cervical area of the experimental and control tooth with a composite resin. Tooth sensitivity response by participant grip force was monitored. The digital oscilloscope was used to record two surface temperatures and a pain response during an IOS or a LCU irradiation. The high-intensity light from a LCU and an IOS was randomly applied at 2 mm above the cervical area for 20s. The data were compared statistical with two-way repeated measures ANOVA and Pearson’s correlation. RESULTS: The illumination caused increasing surface temperatures of about 22.98 ± 3.20 °C for a LCU and 5.86 ± 1.46 °C for an IOS from a 29.5 °C baseline. As the light intensity from the LCU generated more heat, participants reported more tooth sensitivity with a shorter response time (2.10s to 18.70s). There was a positive correlation between surface temperature and pain response (R2 = 0.232; p<0.01). CONCLUSIONS: The heat from high-intensity light from a LCU and an IOS can cause tooth sensitivity in some individuals especially those who had a cervical cavity with dentine sensitivity. The higher light intensity would raise the surface temperature and cause a higher sensitivity response. Key words:Light curing unit, intraoral scanner, tooth temperature, dentine sensitivity. |
---|