Optimizing Calibration for a Capacitance-Based Void Fraction Sensor with Asymmetric Electrodes under Horizontal Flow in a Smoothed Circular Macro-Tube

In this study, a technique that uses a capacitance sensor with an asymmetric electrode to measure the void fraction of a refrigerant was developed. It is known that the void fraction and flow pattern affect the measured capacitance. Therefore, the relationship between the void fraction and capacitan...

Descripción completa

Detalles Bibliográficos
Autores principales: Kim, Moojoong, Komeda, Kanta, Jeong, Jongsoo, Oinuma, Mizuki, Sato, Tetsuya, Saito, Kiyoshi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9103446/
https://www.ncbi.nlm.nih.gov/pubmed/35591200
http://dx.doi.org/10.3390/s22093511
_version_ 1784707559111786496
author Kim, Moojoong
Komeda, Kanta
Jeong, Jongsoo
Oinuma, Mizuki
Sato, Tetsuya
Saito, Kiyoshi
author_facet Kim, Moojoong
Komeda, Kanta
Jeong, Jongsoo
Oinuma, Mizuki
Sato, Tetsuya
Saito, Kiyoshi
author_sort Kim, Moojoong
collection PubMed
description In this study, a technique that uses a capacitance sensor with an asymmetric electrode to measure the void fraction of a refrigerant was developed. It is known that the void fraction and flow pattern affect the measured capacitance. Therefore, the relationship between the void fraction and capacitance is not linear; hence, a calibration method for obtaining accurate measurements is necessary. A calibration method was designed in this study based on repeated capacitance measurements and the bimodal temporal distribution to calibrate the atypical and repetitive flow patterns of slug flow and its transition to the intermittent flow regime. The calibration method also considers the weighted-average relation for the gradual transition of the intermittent to annular flow pattern according to the change from low to high quality. The proposed method was experimentally analyzed under the conditions of R32 refrigerant, a tube inner diameter of 7.1 mm, saturation temperature of 25 °C, mass flux of 100–400 kg m(−2) s(−1), and vapor quality of 0.025–0.900, and it was validated using a quick-closing valve (QCV) system under identical conditions. A relative error of 2.99% was obtained for the entire system, indicating good agreement between the proposed and QCV-based methods.
format Online
Article
Text
id pubmed-9103446
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-91034462022-05-14 Optimizing Calibration for a Capacitance-Based Void Fraction Sensor with Asymmetric Electrodes under Horizontal Flow in a Smoothed Circular Macro-Tube Kim, Moojoong Komeda, Kanta Jeong, Jongsoo Oinuma, Mizuki Sato, Tetsuya Saito, Kiyoshi Sensors (Basel) Article In this study, a technique that uses a capacitance sensor with an asymmetric electrode to measure the void fraction of a refrigerant was developed. It is known that the void fraction and flow pattern affect the measured capacitance. Therefore, the relationship between the void fraction and capacitance is not linear; hence, a calibration method for obtaining accurate measurements is necessary. A calibration method was designed in this study based on repeated capacitance measurements and the bimodal temporal distribution to calibrate the atypical and repetitive flow patterns of slug flow and its transition to the intermittent flow regime. The calibration method also considers the weighted-average relation for the gradual transition of the intermittent to annular flow pattern according to the change from low to high quality. The proposed method was experimentally analyzed under the conditions of R32 refrigerant, a tube inner diameter of 7.1 mm, saturation temperature of 25 °C, mass flux of 100–400 kg m(−2) s(−1), and vapor quality of 0.025–0.900, and it was validated using a quick-closing valve (QCV) system under identical conditions. A relative error of 2.99% was obtained for the entire system, indicating good agreement between the proposed and QCV-based methods. MDPI 2022-05-05 /pmc/articles/PMC9103446/ /pubmed/35591200 http://dx.doi.org/10.3390/s22093511 Text en © 2022 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Kim, Moojoong
Komeda, Kanta
Jeong, Jongsoo
Oinuma, Mizuki
Sato, Tetsuya
Saito, Kiyoshi
Optimizing Calibration for a Capacitance-Based Void Fraction Sensor with Asymmetric Electrodes under Horizontal Flow in a Smoothed Circular Macro-Tube
title Optimizing Calibration for a Capacitance-Based Void Fraction Sensor with Asymmetric Electrodes under Horizontal Flow in a Smoothed Circular Macro-Tube
title_full Optimizing Calibration for a Capacitance-Based Void Fraction Sensor with Asymmetric Electrodes under Horizontal Flow in a Smoothed Circular Macro-Tube
title_fullStr Optimizing Calibration for a Capacitance-Based Void Fraction Sensor with Asymmetric Electrodes under Horizontal Flow in a Smoothed Circular Macro-Tube
title_full_unstemmed Optimizing Calibration for a Capacitance-Based Void Fraction Sensor with Asymmetric Electrodes under Horizontal Flow in a Smoothed Circular Macro-Tube
title_short Optimizing Calibration for a Capacitance-Based Void Fraction Sensor with Asymmetric Electrodes under Horizontal Flow in a Smoothed Circular Macro-Tube
title_sort optimizing calibration for a capacitance-based void fraction sensor with asymmetric electrodes under horizontal flow in a smoothed circular macro-tube
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9103446/
https://www.ncbi.nlm.nih.gov/pubmed/35591200
http://dx.doi.org/10.3390/s22093511
work_keys_str_mv AT kimmoojoong optimizingcalibrationforacapacitancebasedvoidfractionsensorwithasymmetricelectrodesunderhorizontalflowinasmoothedcircularmacrotube
AT komedakanta optimizingcalibrationforacapacitancebasedvoidfractionsensorwithasymmetricelectrodesunderhorizontalflowinasmoothedcircularmacrotube
AT jeongjongsoo optimizingcalibrationforacapacitancebasedvoidfractionsensorwithasymmetricelectrodesunderhorizontalflowinasmoothedcircularmacrotube
AT oinumamizuki optimizingcalibrationforacapacitancebasedvoidfractionsensorwithasymmetricelectrodesunderhorizontalflowinasmoothedcircularmacrotube
AT satotetsuya optimizingcalibrationforacapacitancebasedvoidfractionsensorwithasymmetricelectrodesunderhorizontalflowinasmoothedcircularmacrotube
AT saitokiyoshi optimizingcalibrationforacapacitancebasedvoidfractionsensorwithasymmetricelectrodesunderhorizontalflowinasmoothedcircularmacrotube