Accuracy of atrial fibrillation detection by an insertable cardiac monitor in patients undergoing catheter ablation: Results of the BioVAD study

BACKGROUND: Insertable cardiac monitors (ICMs) are increasingly used to evaluate the atrial fibrillation (AF) burden after catheter ablation of AF. BioMonitor III (BM3) is an ICM with a long sensing vector, which enhances sensing capabilities. The AF detection algorithm of the BM3 is based on R–R in...

Descripción completa

Detalles Bibliográficos
Autores principales: Assaf, Amira, Theuns, Dominic A. M. J., Sakhi, Rafi, Bhagwandien, Rohit E., Szili‐Torok, Tamas, Yap, Sing‐Chien
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9107077/
https://www.ncbi.nlm.nih.gov/pubmed/35481956
http://dx.doi.org/10.1111/anec.12960
Descripción
Sumario:BACKGROUND: Insertable cardiac monitors (ICMs) are increasingly used to evaluate the atrial fibrillation (AF) burden after catheter ablation of AF. BioMonitor III (BM3) is an ICM with a long sensing vector, which enhances sensing capabilities. The AF detection algorithm of the BM3 is based on R–R interval variability. OBJECTIVE: To evaluate the performance of the AF detection algorithm of BM3 in patients before and after catheter ablation of AF using simultaneous Holter recordings. METHODS: In this prospective study, we enrolled patients scheduled for catheter ablation of paroxysmal or persistent AF. After BM3 implantation, patients had a 4 days Holter registration before and 3 months after ablation. All true AF episodes ≥2 min on the Holter were annotated and matched with BM3 detected AF detections. RESULTS: Thirty‐one patients were enrolled (mean age 60 ± 8, 74% male, 68% paroxysmal AF). Fifty‐six Holter registrations were performed in 30 patients. Twelve patients demonstrated at least one true AF episode with a total AF duration of 570 h. The AF burden accuracy of BM3 before catheter ablation was 99.6%, with a duration sensitivity of 98.6% and a duration specificity of 99.9%. The AF burden accuracy of BM3 after catheter ablation was 99.8%, with a duration sensitivity of 90.2% and a duration specificity of 99.9%. Overall, the AF burden detected on the Holter and BM3 demonstrated a high Pearson correlation coefficient of 0.996. CONCLUSION: BM3 accurately detects AF burden in patients before and after catheter ablation of AF.