Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers

This study evaluated and compared the overall equipment effectiveness (OEE), sensitivity, specificity, and efficiency of the high-end hematology analyzers, Yumizen H2500, DxH 800, DxH 900 and XN-9000 (XN-10). A total of 400 anonymized left over’s K2 EDTA whole blood samples were analyzed for complet...

Descripción completa

Detalles Bibliográficos
Autores principales: Rastogi, Shubham, Chandran Nair, Sukesh, Murugan, Pandiyan, Sukumar, Asady Sukanya, Mammen, Joy J., Mullai, Saravanan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9117814/
https://www.ncbi.nlm.nih.gov/pubmed/35601458
http://dx.doi.org/10.1016/j.plabm.2022.e00275
_version_ 1784710391267328000
author Rastogi, Shubham
Chandran Nair, Sukesh
Murugan, Pandiyan
Sukumar, Asady Sukanya
Mammen, Joy J.
Mullai, Saravanan
author_facet Rastogi, Shubham
Chandran Nair, Sukesh
Murugan, Pandiyan
Sukumar, Asady Sukanya
Mammen, Joy J.
Mullai, Saravanan
author_sort Rastogi, Shubham
collection PubMed
description This study evaluated and compared the overall equipment effectiveness (OEE), sensitivity, specificity, and efficiency of the high-end hematology analyzers, Yumizen H2500, DxH 800, DxH 900 and XN-9000 (XN-10). A total of 400 anonymized left over’s K2 EDTA whole blood samples were analyzed for complete blood count. Of 400 samples, 200 were tested on Yumizen H2500; DxH 800 & DxH 900 while the other 200 were tested on Yumizen H2500 & XN-9000 (XN-10), respectively. The OEE was good and comparable for all the hematology analyzers except DxH 800 showing an average status. The sensitivity (%), specificity (%) and turnaround time (in minutes) for Yumizen H2500, DxH 800, DxH 900 and XN-9000 (XN-10) were 91.67, 61.11 & 103; 66.67, 54.55, & 149; 83.33, 27.27 & 136; 83.33, 28.57 & 122, respectively. Confusion matrix highlights the difficulty for DxH 800 and DxH 900 to discriminate left shift or blasts with large hyper-segmented neutrophils. The flags triggered by Yumizen H2500 were markedly changed to large hyper-segmented neutrophils. Lymphoblast caused more confusion for XN-9000 (XN-10), as it came out to be atypical lymphocytes, or hypersegmented neutrophils. Although comparable in OEE index to other analyzers, the Yumizen H2500 seems to be more reliable in detecting the abnormal cells as it has high sensitivity, specificity and less turnaround time. Thus, analysis adding specificity, sensitivity, and efficiency parameters to the OEE index provides more reliable information of the analyzers.
format Online
Article
Text
id pubmed-9117814
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-91178142022-05-20 Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers Rastogi, Shubham Chandran Nair, Sukesh Murugan, Pandiyan Sukumar, Asady Sukanya Mammen, Joy J. Mullai, Saravanan Pract Lab Med Original Research Article This study evaluated and compared the overall equipment effectiveness (OEE), sensitivity, specificity, and efficiency of the high-end hematology analyzers, Yumizen H2500, DxH 800, DxH 900 and XN-9000 (XN-10). A total of 400 anonymized left over’s K2 EDTA whole blood samples were analyzed for complete blood count. Of 400 samples, 200 were tested on Yumizen H2500; DxH 800 & DxH 900 while the other 200 were tested on Yumizen H2500 & XN-9000 (XN-10), respectively. The OEE was good and comparable for all the hematology analyzers except DxH 800 showing an average status. The sensitivity (%), specificity (%) and turnaround time (in minutes) for Yumizen H2500, DxH 800, DxH 900 and XN-9000 (XN-10) were 91.67, 61.11 & 103; 66.67, 54.55, & 149; 83.33, 27.27 & 136; 83.33, 28.57 & 122, respectively. Confusion matrix highlights the difficulty for DxH 800 and DxH 900 to discriminate left shift or blasts with large hyper-segmented neutrophils. The flags triggered by Yumizen H2500 were markedly changed to large hyper-segmented neutrophils. Lymphoblast caused more confusion for XN-9000 (XN-10), as it came out to be atypical lymphocytes, or hypersegmented neutrophils. Although comparable in OEE index to other analyzers, the Yumizen H2500 seems to be more reliable in detecting the abnormal cells as it has high sensitivity, specificity and less turnaround time. Thus, analysis adding specificity, sensitivity, and efficiency parameters to the OEE index provides more reliable information of the analyzers. Elsevier 2022-04-18 /pmc/articles/PMC9117814/ /pubmed/35601458 http://dx.doi.org/10.1016/j.plabm.2022.e00275 Text en © 2022 The Authors https://creativecommons.org/licenses/by/4.0/This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Original Research Article
Rastogi, Shubham
Chandran Nair, Sukesh
Murugan, Pandiyan
Sukumar, Asady Sukanya
Mammen, Joy J.
Mullai, Saravanan
Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers
title Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers
title_full Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers
title_fullStr Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers
title_full_unstemmed Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers
title_short Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers
title_sort overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers
topic Original Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9117814/
https://www.ncbi.nlm.nih.gov/pubmed/35601458
http://dx.doi.org/10.1016/j.plabm.2022.e00275
work_keys_str_mv AT rastogishubham overallequipmenteffectivenessefficiencyandslidereviewanalysisofhighendhematologyanalyzers
AT chandrannairsukesh overallequipmenteffectivenessefficiencyandslidereviewanalysisofhighendhematologyanalyzers
AT muruganpandiyan overallequipmenteffectivenessefficiencyandslidereviewanalysisofhighendhematologyanalyzers
AT sukumarasadysukanya overallequipmenteffectivenessefficiencyandslidereviewanalysisofhighendhematologyanalyzers
AT mammenjoyj overallequipmenteffectivenessefficiencyandslidereviewanalysisofhighendhematologyanalyzers
AT mullaisaravanan overallequipmenteffectivenessefficiencyandslidereviewanalysisofhighendhematologyanalyzers