Cargando…

Conebeam CT‐guided 3D MLC‐based spatially fractionated radiation therapy for bulky masses

For fast, safe, and effective management of large and bulky (≥8 cm) non‐resectable tumors, we have developed a conebeam CT‐guided three‐dimensional (3D)‐conformal MLC‐based spatially fractionated radiation therapy (SFRT) treatment. Using an in‐house MLC‐fitting algorithm, Millennium 120 leaves were...

Descripción completa

Detalles Bibliográficos
Autores principales: Pokhrel, Damodar, Bernard, Mark E, Mallory, Richard, St Clair, William, Kudrimoti, Mahesh
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9121033/
https://www.ncbi.nlm.nih.gov/pubmed/35446479
http://dx.doi.org/10.1002/acm2.13608
_version_ 1784711067952218112
author Pokhrel, Damodar
Bernard, Mark E
Mallory, Richard
St Clair, William
Kudrimoti, Mahesh
author_facet Pokhrel, Damodar
Bernard, Mark E
Mallory, Richard
St Clair, William
Kudrimoti, Mahesh
author_sort Pokhrel, Damodar
collection PubMed
description For fast, safe, and effective management of large and bulky (≥8 cm) non‐resectable tumors, we have developed a conebeam CT‐guided three‐dimensional (3D)‐conformal MLC‐based spatially fractionated radiation therapy (SFRT) treatment. Using an in‐house MLC‐fitting algorithm, Millennium 120 leaves were fitted to the gross tumor volume (GTV) generating 1‐cm diameter holes at 2‐cm center‐to‐center distance at isocenter. SFRT plans of 15 Gy were generated using four to six coplanar crossfire gantry angles 60° apart with a 90° collimator, differentially weighted with 6‐ or 10‐MV beams. A dose was calculated using AcurosXB algorithm, generating sieve‐like dose channels without post‐processing the physician‐drawn GTV contour within an hour of CT simulation allowing for the same day treatment. In total, 50 extracranial patients have been planned and treated using this method, comprising multiple treatment sites. This novel MLC‐fitting algorithm provided excellent dose parameters with mean GTV (V7.5 Gy) and mean GTV doses of 53.2% and 7.9 Gy, respectively, for 15 Gy plans. Average peak‐to‐valley dose ratio was 3.2. Mean beam‐on time was 3.32 min, and treatment time, including patient setup and CBCT to beam‐off, was within 15 min. Average 3D couch correction from original skin‐markers was <1.0 cm. 3D MLC‐based SFRT plans enhanced target dose for bulky masses, including deep‐seated large tumors while protecting skin and adjacent critical organs. Additionally, it provides the same day, safe, effective, and convenient treatment by eliminating the risk to therapists and patients from heavy gantry‐mounted physical GRID‐block—we recommend other centers to use this simple and clinically useful method. This rapid SFRT planning technique is easily adoptable in any radiation oncology clinic by eliminating the need for plan optimization and patient‐specific quality assurance times while providing dosimetry information in the treatment planning system. This potentially allows for dose‐escalation to deep‐seated masses to debulk unresectable large tumors providing an option for neoadjuvant treatment. An outcome study of clinical trial is underway.
format Online
Article
Text
id pubmed-9121033
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-91210332022-05-21 Conebeam CT‐guided 3D MLC‐based spatially fractionated radiation therapy for bulky masses Pokhrel, Damodar Bernard, Mark E Mallory, Richard St Clair, William Kudrimoti, Mahesh J Appl Clin Med Phys Technical Notes For fast, safe, and effective management of large and bulky (≥8 cm) non‐resectable tumors, we have developed a conebeam CT‐guided three‐dimensional (3D)‐conformal MLC‐based spatially fractionated radiation therapy (SFRT) treatment. Using an in‐house MLC‐fitting algorithm, Millennium 120 leaves were fitted to the gross tumor volume (GTV) generating 1‐cm diameter holes at 2‐cm center‐to‐center distance at isocenter. SFRT plans of 15 Gy were generated using four to six coplanar crossfire gantry angles 60° apart with a 90° collimator, differentially weighted with 6‐ or 10‐MV beams. A dose was calculated using AcurosXB algorithm, generating sieve‐like dose channels without post‐processing the physician‐drawn GTV contour within an hour of CT simulation allowing for the same day treatment. In total, 50 extracranial patients have been planned and treated using this method, comprising multiple treatment sites. This novel MLC‐fitting algorithm provided excellent dose parameters with mean GTV (V7.5 Gy) and mean GTV doses of 53.2% and 7.9 Gy, respectively, for 15 Gy plans. Average peak‐to‐valley dose ratio was 3.2. Mean beam‐on time was 3.32 min, and treatment time, including patient setup and CBCT to beam‐off, was within 15 min. Average 3D couch correction from original skin‐markers was <1.0 cm. 3D MLC‐based SFRT plans enhanced target dose for bulky masses, including deep‐seated large tumors while protecting skin and adjacent critical organs. Additionally, it provides the same day, safe, effective, and convenient treatment by eliminating the risk to therapists and patients from heavy gantry‐mounted physical GRID‐block—we recommend other centers to use this simple and clinically useful method. This rapid SFRT planning technique is easily adoptable in any radiation oncology clinic by eliminating the need for plan optimization and patient‐specific quality assurance times while providing dosimetry information in the treatment planning system. This potentially allows for dose‐escalation to deep‐seated masses to debulk unresectable large tumors providing an option for neoadjuvant treatment. An outcome study of clinical trial is underway. John Wiley and Sons Inc. 2022-04-21 /pmc/articles/PMC9121033/ /pubmed/35446479 http://dx.doi.org/10.1002/acm2.13608 Text en © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Notes
Pokhrel, Damodar
Bernard, Mark E
Mallory, Richard
St Clair, William
Kudrimoti, Mahesh
Conebeam CT‐guided 3D MLC‐based spatially fractionated radiation therapy for bulky masses
title Conebeam CT‐guided 3D MLC‐based spatially fractionated radiation therapy for bulky masses
title_full Conebeam CT‐guided 3D MLC‐based spatially fractionated radiation therapy for bulky masses
title_fullStr Conebeam CT‐guided 3D MLC‐based spatially fractionated radiation therapy for bulky masses
title_full_unstemmed Conebeam CT‐guided 3D MLC‐based spatially fractionated radiation therapy for bulky masses
title_short Conebeam CT‐guided 3D MLC‐based spatially fractionated radiation therapy for bulky masses
title_sort conebeam ct‐guided 3d mlc‐based spatially fractionated radiation therapy for bulky masses
topic Technical Notes
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9121033/
https://www.ncbi.nlm.nih.gov/pubmed/35446479
http://dx.doi.org/10.1002/acm2.13608
work_keys_str_mv AT pokhreldamodar conebeamctguided3dmlcbasedspatiallyfractionatedradiationtherapyforbulkymasses
AT bernardmarke conebeamctguided3dmlcbasedspatiallyfractionatedradiationtherapyforbulkymasses
AT malloryrichard conebeamctguided3dmlcbasedspatiallyfractionatedradiationtherapyforbulkymasses
AT stclairwilliam conebeamctguided3dmlcbasedspatiallyfractionatedradiationtherapyforbulkymasses
AT kudrimotimahesh conebeamctguided3dmlcbasedspatiallyfractionatedradiationtherapyforbulkymasses