A novel methodology for the optimization of transmission and dosimetric leaf gap parameters

PURPOSE: Optimization of dosimetric leaf gap (DLG) and transmission is commonly performed through a manual trial and error process, which can lead to sub‐optimal values. The purpose of this work is to create an alternative automated optimization process that provides the optimal DLG and transmission...

Descripción completa

Detalles Bibliográficos
Autores principales: DiCostanzo, Dominic J., Ayan, Ahmet S.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9121040/
https://www.ncbi.nlm.nih.gov/pubmed/35226392
http://dx.doi.org/10.1002/acm2.13565
_version_ 1784711069472653312
author DiCostanzo, Dominic J.
Ayan, Ahmet S.
author_facet DiCostanzo, Dominic J.
Ayan, Ahmet S.
author_sort DiCostanzo, Dominic J.
collection PubMed
description PURPOSE: Optimization of dosimetric leaf gap (DLG) and transmission is commonly performed through a manual trial and error process, which can lead to sub‐optimal values. The purpose of this work is to create an alternative automated optimization process that provides the optimal DLG and transmission pair for use in a clinical setting. METHODS: Utilizing the treatment planning system application programming interface, a phase space of clinically viable DLG and transmission pairs was generated. The phase space contained 51,051 dose planes for DLGs between 0.0  and 2.5 mm and transmission values between 0.01% and 2.5%. Thirteen plans were measured for multiple multileaf collimator types and nominal beam energies. The optimization minimized the mean γ‐index and maximized the γ‐index pass rate. The optimized values were validated using five plans excluded from the optimization. RESULTS: Of the nominal beam energies and multileaf collimator system (MLC)‐type combinations tested, 6/7 showed an increase in γ‐index pass rate and a decrease in mean γ‐index signifying better agreement between measurement and calculation. When comparing the optimized DLG and transmission values to the clinically implemented values identified via an iterative method, 5/7 energy, and MLC type combinations showed no statistically significant changes. In addition, the optimized values were benchmarked against three Task Group 119 plans with published γ‐index pass rates, which had been held out of the optimization. For those plans, the optimized DLG and transmission values provided the same or better γ‐index pass rates. CONCLUSION: We presented a novel and viable automated alternative to current approaches of selecting the DLG and transmission parameters. This method will reduce the time required to determine the clinically acceptable DLG and transmission parameters and ensure optimality for the plans included in the optimization.
format Online
Article
Text
id pubmed-9121040
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-91210402022-05-21 A novel methodology for the optimization of transmission and dosimetric leaf gap parameters DiCostanzo, Dominic J. Ayan, Ahmet S. J Appl Clin Med Phys Radiation Oncology Physics PURPOSE: Optimization of dosimetric leaf gap (DLG) and transmission is commonly performed through a manual trial and error process, which can lead to sub‐optimal values. The purpose of this work is to create an alternative automated optimization process that provides the optimal DLG and transmission pair for use in a clinical setting. METHODS: Utilizing the treatment planning system application programming interface, a phase space of clinically viable DLG and transmission pairs was generated. The phase space contained 51,051 dose planes for DLGs between 0.0  and 2.5 mm and transmission values between 0.01% and 2.5%. Thirteen plans were measured for multiple multileaf collimator types and nominal beam energies. The optimization minimized the mean γ‐index and maximized the γ‐index pass rate. The optimized values were validated using five plans excluded from the optimization. RESULTS: Of the nominal beam energies and multileaf collimator system (MLC)‐type combinations tested, 6/7 showed an increase in γ‐index pass rate and a decrease in mean γ‐index signifying better agreement between measurement and calculation. When comparing the optimized DLG and transmission values to the clinically implemented values identified via an iterative method, 5/7 energy, and MLC type combinations showed no statistically significant changes. In addition, the optimized values were benchmarked against three Task Group 119 plans with published γ‐index pass rates, which had been held out of the optimization. For those plans, the optimized DLG and transmission values provided the same or better γ‐index pass rates. CONCLUSION: We presented a novel and viable automated alternative to current approaches of selecting the DLG and transmission parameters. This method will reduce the time required to determine the clinically acceptable DLG and transmission parameters and ensure optimality for the plans included in the optimization. John Wiley and Sons Inc. 2022-02-28 /pmc/articles/PMC9121040/ /pubmed/35226392 http://dx.doi.org/10.1002/acm2.13565 Text en © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
DiCostanzo, Dominic J.
Ayan, Ahmet S.
A novel methodology for the optimization of transmission and dosimetric leaf gap parameters
title A novel methodology for the optimization of transmission and dosimetric leaf gap parameters
title_full A novel methodology for the optimization of transmission and dosimetric leaf gap parameters
title_fullStr A novel methodology for the optimization of transmission and dosimetric leaf gap parameters
title_full_unstemmed A novel methodology for the optimization of transmission and dosimetric leaf gap parameters
title_short A novel methodology for the optimization of transmission and dosimetric leaf gap parameters
title_sort novel methodology for the optimization of transmission and dosimetric leaf gap parameters
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9121040/
https://www.ncbi.nlm.nih.gov/pubmed/35226392
http://dx.doi.org/10.1002/acm2.13565
work_keys_str_mv AT dicostanzodominicj anovelmethodologyfortheoptimizationoftransmissionanddosimetricleafgapparameters
AT ayanahmets anovelmethodologyfortheoptimizationoftransmissionanddosimetricleafgapparameters
AT dicostanzodominicj novelmethodologyfortheoptimizationoftransmissionanddosimetricleafgapparameters
AT ayanahmets novelmethodologyfortheoptimizationoftransmissionanddosimetricleafgapparameters