Evaluation of the detection accuracy of set-up for various treatment sites using surface-guided radiotherapy system, VOXELAN: a phantom study

The purpose of this study is to evaluate the detection accuracy of a 3-dimensional (3D) body scanner, VOXELAN, in surface-guided radiotherapy (SGRT) of each part of the human body using a whole-body human phantom. We used A Resusci Anne was used as the whole-body phantom. The detection accuracy of V...

Descripción completa

Detalles Bibliográficos
Autores principales: Saito, Masahide, Ueda, Koji, Suzuki, Hidekazu, Komiyama, Takafumi, Marino, Kan, Aoki, Shinichi, Sano, Naoki, Onishi, Hiroshi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9124621/
https://www.ncbi.nlm.nih.gov/pubmed/35467750
http://dx.doi.org/10.1093/jrr/rrac015
Descripción
Sumario:The purpose of this study is to evaluate the detection accuracy of a 3-dimensional (3D) body scanner, VOXELAN, in surface-guided radiotherapy (SGRT) of each part of the human body using a whole-body human phantom. We used A Resusci Anne was used as the whole-body phantom. The detection accuracy of VOXELAN in a radiotherapy treatment room with a linear accelerator (LINAC) was evaluated for two reference images: reconstruction of the planning computed tomography (CT) image (CT reference) and scanning by VOXELAN before the treatment (scan reference). The accuracy of the translational and rotational directions was verified for four treatment sites (open face shell, breast, abdomen, and arm), using the magnitude of the 6D robotic couch movement as the true value. Our results showed that the detection accuracy improved as the displacement from the reference position decreased for all the sites. Using the scan reference, the average accuracy of the translational and rotational axes was within 1.44 mm and 0.41°, respectively, for all sites except the arms. Similarly, using the CT reference, the average accuracy was within 2.45 mm and 1.35°, respectively. Additionally, it was difficult for both reference images to recognize misalignment of the arms. In conclusion we discovered that VOXELAN achieved a high detection accuracy for the head with an open face shell, chest, and abdomen, indicating that the system is useful in a clinical setting. However, it is necessary to pay attention to location matching for areas with few features, such as surface irregularities and potential errors, when the reference image is created from CT.