Evaluation of the detection accuracy of set-up for various treatment sites using surface-guided radiotherapy system, VOXELAN: a phantom study

The purpose of this study is to evaluate the detection accuracy of a 3-dimensional (3D) body scanner, VOXELAN, in surface-guided radiotherapy (SGRT) of each part of the human body using a whole-body human phantom. We used A Resusci Anne was used as the whole-body phantom. The detection accuracy of V...

Descripción completa

Detalles Bibliográficos
Autores principales: Saito, Masahide, Ueda, Koji, Suzuki, Hidekazu, Komiyama, Takafumi, Marino, Kan, Aoki, Shinichi, Sano, Naoki, Onishi, Hiroshi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9124621/
https://www.ncbi.nlm.nih.gov/pubmed/35467750
http://dx.doi.org/10.1093/jrr/rrac015
_version_ 1784711775764086784
author Saito, Masahide
Ueda, Koji
Suzuki, Hidekazu
Komiyama, Takafumi
Marino, Kan
Aoki, Shinichi
Sano, Naoki
Onishi, Hiroshi
author_facet Saito, Masahide
Ueda, Koji
Suzuki, Hidekazu
Komiyama, Takafumi
Marino, Kan
Aoki, Shinichi
Sano, Naoki
Onishi, Hiroshi
author_sort Saito, Masahide
collection PubMed
description The purpose of this study is to evaluate the detection accuracy of a 3-dimensional (3D) body scanner, VOXELAN, in surface-guided radiotherapy (SGRT) of each part of the human body using a whole-body human phantom. We used A Resusci Anne was used as the whole-body phantom. The detection accuracy of VOXELAN in a radiotherapy treatment room with a linear accelerator (LINAC) was evaluated for two reference images: reconstruction of the planning computed tomography (CT) image (CT reference) and scanning by VOXELAN before the treatment (scan reference). The accuracy of the translational and rotational directions was verified for four treatment sites (open face shell, breast, abdomen, and arm), using the magnitude of the 6D robotic couch movement as the true value. Our results showed that the detection accuracy improved as the displacement from the reference position decreased for all the sites. Using the scan reference, the average accuracy of the translational and rotational axes was within 1.44 mm and 0.41°, respectively, for all sites except the arms. Similarly, using the CT reference, the average accuracy was within 2.45 mm and 1.35°, respectively. Additionally, it was difficult for both reference images to recognize misalignment of the arms. In conclusion we discovered that VOXELAN achieved a high detection accuracy for the head with an open face shell, chest, and abdomen, indicating that the system is useful in a clinical setting. However, it is necessary to pay attention to location matching for areas with few features, such as surface irregularities and potential errors, when the reference image is created from CT.
format Online
Article
Text
id pubmed-9124621
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Oxford University Press
record_format MEDLINE/PubMed
spelling pubmed-91246212022-05-23 Evaluation of the detection accuracy of set-up for various treatment sites using surface-guided radiotherapy system, VOXELAN: a phantom study Saito, Masahide Ueda, Koji Suzuki, Hidekazu Komiyama, Takafumi Marino, Kan Aoki, Shinichi Sano, Naoki Onishi, Hiroshi J Radiat Res Oncology / Medicine The purpose of this study is to evaluate the detection accuracy of a 3-dimensional (3D) body scanner, VOXELAN, in surface-guided radiotherapy (SGRT) of each part of the human body using a whole-body human phantom. We used A Resusci Anne was used as the whole-body phantom. The detection accuracy of VOXELAN in a radiotherapy treatment room with a linear accelerator (LINAC) was evaluated for two reference images: reconstruction of the planning computed tomography (CT) image (CT reference) and scanning by VOXELAN before the treatment (scan reference). The accuracy of the translational and rotational directions was verified for four treatment sites (open face shell, breast, abdomen, and arm), using the magnitude of the 6D robotic couch movement as the true value. Our results showed that the detection accuracy improved as the displacement from the reference position decreased for all the sites. Using the scan reference, the average accuracy of the translational and rotational axes was within 1.44 mm and 0.41°, respectively, for all sites except the arms. Similarly, using the CT reference, the average accuracy was within 2.45 mm and 1.35°, respectively. Additionally, it was difficult for both reference images to recognize misalignment of the arms. In conclusion we discovered that VOXELAN achieved a high detection accuracy for the head with an open face shell, chest, and abdomen, indicating that the system is useful in a clinical setting. However, it is necessary to pay attention to location matching for areas with few features, such as surface irregularities and potential errors, when the reference image is created from CT. Oxford University Press 2022-04-25 /pmc/articles/PMC9124621/ /pubmed/35467750 http://dx.doi.org/10.1093/jrr/rrac015 Text en © The Author(s) 2022. Published by Oxford University Press on behalf of The Japanese Radiation Research Society and Japanese Society for Radiation Oncology. https://creativecommons.org/licenses/by/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Oncology / Medicine
Saito, Masahide
Ueda, Koji
Suzuki, Hidekazu
Komiyama, Takafumi
Marino, Kan
Aoki, Shinichi
Sano, Naoki
Onishi, Hiroshi
Evaluation of the detection accuracy of set-up for various treatment sites using surface-guided radiotherapy system, VOXELAN: a phantom study
title Evaluation of the detection accuracy of set-up for various treatment sites using surface-guided radiotherapy system, VOXELAN: a phantom study
title_full Evaluation of the detection accuracy of set-up for various treatment sites using surface-guided radiotherapy system, VOXELAN: a phantom study
title_fullStr Evaluation of the detection accuracy of set-up for various treatment sites using surface-guided radiotherapy system, VOXELAN: a phantom study
title_full_unstemmed Evaluation of the detection accuracy of set-up for various treatment sites using surface-guided radiotherapy system, VOXELAN: a phantom study
title_short Evaluation of the detection accuracy of set-up for various treatment sites using surface-guided radiotherapy system, VOXELAN: a phantom study
title_sort evaluation of the detection accuracy of set-up for various treatment sites using surface-guided radiotherapy system, voxelan: a phantom study
topic Oncology / Medicine
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9124621/
https://www.ncbi.nlm.nih.gov/pubmed/35467750
http://dx.doi.org/10.1093/jrr/rrac015
work_keys_str_mv AT saitomasahide evaluationofthedetectionaccuracyofsetupforvarioustreatmentsitesusingsurfaceguidedradiotherapysystemvoxelanaphantomstudy
AT uedakoji evaluationofthedetectionaccuracyofsetupforvarioustreatmentsitesusingsurfaceguidedradiotherapysystemvoxelanaphantomstudy
AT suzukihidekazu evaluationofthedetectionaccuracyofsetupforvarioustreatmentsitesusingsurfaceguidedradiotherapysystemvoxelanaphantomstudy
AT komiyamatakafumi evaluationofthedetectionaccuracyofsetupforvarioustreatmentsitesusingsurfaceguidedradiotherapysystemvoxelanaphantomstudy
AT marinokan evaluationofthedetectionaccuracyofsetupforvarioustreatmentsitesusingsurfaceguidedradiotherapysystemvoxelanaphantomstudy
AT aokishinichi evaluationofthedetectionaccuracyofsetupforvarioustreatmentsitesusingsurfaceguidedradiotherapysystemvoxelanaphantomstudy
AT sanonaoki evaluationofthedetectionaccuracyofsetupforvarioustreatmentsitesusingsurfaceguidedradiotherapysystemvoxelanaphantomstudy
AT onishihiroshi evaluationofthedetectionaccuracyofsetupforvarioustreatmentsitesusingsurfaceguidedradiotherapysystemvoxelanaphantomstudy