Determination of tobramycin in eye drops with an open-source hardware ion mobility spectrometer
The analysis of tobramycin was demonstrated successfully as an example for electrospray ionization on an open-source hardware ion mobility spectrometer. This instrument was assembled inexpensively in-house, and required only very few purpose-made components. The quantitative determination of tobramy...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9124657/ https://www.ncbi.nlm.nih.gov/pubmed/35381854 http://dx.doi.org/10.1007/s00216-022-04050-2 |
Sumario: | The analysis of tobramycin was demonstrated successfully as an example for electrospray ionization on an open-source hardware ion mobility spectrometer. This instrument was assembled inexpensively in-house, and required only very few purpose-made components. The quantitative determination of tobramycin required 20 s for a reading. The calibration curve for the range from 50 to 200 μM was found to be linear with a correlation coefficient of r = 0.9994. A good reproducibility was obtained (3% relative standard deviation) and the limit of detection was determined as 8 μM. As the concentration of the active ingredient in the eye drops (ophthalmic solutions) is too high for the sensitivity of the instrument, the samples had to be diluted appropriately. GRAPHICAL ABSTRACT: [Image: see text] |
---|