Development of a low-cost device for testing glove and condom leakage
BACKGROUND: Currently, hospital-acquired infections in healthcare workers and patients are a major concern. On the other hand, sexually transmitted infections and diseases, unwanted pregnancies, and unsafe abortions continue to be a public health concern, particularly in developing countries. Gloves...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2022
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9127158/ https://www.ncbi.nlm.nih.gov/pubmed/35620039 http://dx.doi.org/10.1016/j.amsu.2022.103791 |
_version_ | 1784712287649529856 |
---|---|
author | Seyoum, Mekides Abera Belachew, Menen Teshome Dessale, Meron Berihun Adugna, Meti Wakjira Keto, Tizita Yohannes Wassie, Yezbalem Getnet Dobamo, Habtamu Abafoge Getaneh, Melese Uma Nemomssa, Hundessa Daba |
author_facet | Seyoum, Mekides Abera Belachew, Menen Teshome Dessale, Meron Berihun Adugna, Meti Wakjira Keto, Tizita Yohannes Wassie, Yezbalem Getnet Dobamo, Habtamu Abafoge Getaneh, Melese Uma Nemomssa, Hundessa Daba |
author_sort | Seyoum, Mekides Abera |
collection | PubMed |
description | BACKGROUND: Currently, hospital-acquired infections in healthcare workers and patients are a major concern. On the other hand, sexually transmitted infections and diseases, unwanted pregnancies, and unsafe abortions continue to be a public health concern, particularly in developing countries. Gloves are among the most commonly used personal protective equipment to safeguard healthcare workers' hands from contagious infections, and using a condom is strongly advised for people who have sexual relations with more than one partner. However, the quality of gloves and condoms in developing countries is a subject of concern. The usage of quality test instruments such as glove leakage test apparatus (GLTA), leakage testers by water level, the Ammonia leak testing method (ALTM), conductivity-based leakage testers, and water hang testers in developing countries is limited owing to cost, accessibility, and safety. The main purpose of this study was to develop and test a low-cost integrated device to test glove and condom leakage that is safe and easily accessible in resource poor settings. METHOD: In this study, an integrated glove and condom leakage testing device for detecting pin holes and leakages is proposed. The device automatically fills a randomly selected condom and glove with a predetermined volume of water based on International Organization for Standardization (ISO) criteria. RESULTS: The prototype of the proposed device was successfully developed and tested. The accuracy of 98.66% for filling condom samples with 300 ml of water and 99.29% for filling glove samples with 1000 ml of water was achieved. CONCLUSION: The implementation of the developed prototype in resource poor settings to test gloves and condom leakage has the potential to improve the safety of healthcare workers, patients, and the general public. |
format | Online Article Text |
id | pubmed-9127158 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2022 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-91271582022-05-25 Development of a low-cost device for testing glove and condom leakage Seyoum, Mekides Abera Belachew, Menen Teshome Dessale, Meron Berihun Adugna, Meti Wakjira Keto, Tizita Yohannes Wassie, Yezbalem Getnet Dobamo, Habtamu Abafoge Getaneh, Melese Uma Nemomssa, Hundessa Daba Ann Med Surg (Lond) Experimental Research BACKGROUND: Currently, hospital-acquired infections in healthcare workers and patients are a major concern. On the other hand, sexually transmitted infections and diseases, unwanted pregnancies, and unsafe abortions continue to be a public health concern, particularly in developing countries. Gloves are among the most commonly used personal protective equipment to safeguard healthcare workers' hands from contagious infections, and using a condom is strongly advised for people who have sexual relations with more than one partner. However, the quality of gloves and condoms in developing countries is a subject of concern. The usage of quality test instruments such as glove leakage test apparatus (GLTA), leakage testers by water level, the Ammonia leak testing method (ALTM), conductivity-based leakage testers, and water hang testers in developing countries is limited owing to cost, accessibility, and safety. The main purpose of this study was to develop and test a low-cost integrated device to test glove and condom leakage that is safe and easily accessible in resource poor settings. METHOD: In this study, an integrated glove and condom leakage testing device for detecting pin holes and leakages is proposed. The device automatically fills a randomly selected condom and glove with a predetermined volume of water based on International Organization for Standardization (ISO) criteria. RESULTS: The prototype of the proposed device was successfully developed and tested. The accuracy of 98.66% for filling condom samples with 300 ml of water and 99.29% for filling glove samples with 1000 ml of water was achieved. CONCLUSION: The implementation of the developed prototype in resource poor settings to test gloves and condom leakage has the potential to improve the safety of healthcare workers, patients, and the general public. Elsevier 2022-05-17 /pmc/articles/PMC9127158/ /pubmed/35620039 http://dx.doi.org/10.1016/j.amsu.2022.103791 Text en © 2022 Published by Elsevier Ltd on behalf of IJS Publishing Group Ltd. https://creativecommons.org/licenses/by/4.0/This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Experimental Research Seyoum, Mekides Abera Belachew, Menen Teshome Dessale, Meron Berihun Adugna, Meti Wakjira Keto, Tizita Yohannes Wassie, Yezbalem Getnet Dobamo, Habtamu Abafoge Getaneh, Melese Uma Nemomssa, Hundessa Daba Development of a low-cost device for testing glove and condom leakage |
title | Development of a low-cost device for testing glove and condom leakage |
title_full | Development of a low-cost device for testing glove and condom leakage |
title_fullStr | Development of a low-cost device for testing glove and condom leakage |
title_full_unstemmed | Development of a low-cost device for testing glove and condom leakage |
title_short | Development of a low-cost device for testing glove and condom leakage |
title_sort | development of a low-cost device for testing glove and condom leakage |
topic | Experimental Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9127158/ https://www.ncbi.nlm.nih.gov/pubmed/35620039 http://dx.doi.org/10.1016/j.amsu.2022.103791 |
work_keys_str_mv | AT seyoummekidesabera developmentofalowcostdevicefortestinggloveandcondomleakage AT belachewmenenteshome developmentofalowcostdevicefortestinggloveandcondomleakage AT dessalemeronberihun developmentofalowcostdevicefortestinggloveandcondomleakage AT adugnametiwakjira developmentofalowcostdevicefortestinggloveandcondomleakage AT ketotizitayohannes developmentofalowcostdevicefortestinggloveandcondomleakage AT wassieyezbalemgetnet developmentofalowcostdevicefortestinggloveandcondomleakage AT dobamohabtamuabafoge developmentofalowcostdevicefortestinggloveandcondomleakage AT getanehmeleseuma developmentofalowcostdevicefortestinggloveandcondomleakage AT nemomssahundessadaba developmentofalowcostdevicefortestinggloveandcondomleakage |