Differences and Limits of Agreement among Pentacam, Corvis-ST, and IOL-Master 700 Optical Biometric Devices regarding Central Corneal Thickness Measurements

PURPOSE: To investigate the differences and limits of agreement in measuring corneal thickness using Pentacam, Corvis, and intraocular lens (IOL)-Master 700 devices. METHODS: This study was conducted on 37 right eyes of 21 males and 16 females (n = 37) with a mean age of 52.11 ± 6.30 years. The cent...

Descripción completa

Detalles Bibliográficos
Autores principales: Rajabi, Sattar, Asharlous, Amir, Riazi, Abbas, Khabazkhoob, Mehdi, Moalej, Alireza
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2022
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9128434/
https://www.ncbi.nlm.nih.gov/pubmed/35620377
http://dx.doi.org/10.4103/joco.joco_96_21
_version_ 1784712560543531008
author Rajabi, Sattar
Asharlous, Amir
Riazi, Abbas
Khabazkhoob, Mehdi
Moalej, Alireza
author_facet Rajabi, Sattar
Asharlous, Amir
Riazi, Abbas
Khabazkhoob, Mehdi
Moalej, Alireza
author_sort Rajabi, Sattar
collection PubMed
description PURPOSE: To investigate the differences and limits of agreement in measuring corneal thickness using Pentacam, Corvis, and intraocular lens (IOL)-Master 700 devices. METHODS: This study was conducted on 37 right eyes of 21 males and 16 females (n = 37) with a mean age of 52.11 ± 6.30 years. The central corneal thickness was measured using three optical biometric devices, including Pentacam, Corvis, and IOL-Master 700. The inclusion criteria were normal eyes without any ophthalmological abnormalities, history of ocular pathology, or ocular surgery. The data obtained from these three devices were compared two by two. The correlation and agreement limits among them were analyzed using statistical techniques. RESULTS: The mean standard deviation differences between Pentacam and Corvis, Pentacam and IOL-Master 700, as well as Corvis and IOL-Master 700 regarding the corneal thickness measurement, were 22.13 ± 8.05, 7.91 ± 8.02, and 14.21 ± 9.85 μm, respectively, which were statistically significant (P < 0.0001). Based on the investigation of the limits of agreement according to the Bland Altman method, the corresponding values between Pentacam and Corvis, Pentacam and IOL-Master 700, and Corvis and IOL-Master 700 were -16.2 to +15.4, -15.8 to +16.3, and -20.1 to +20.0 μm, respectively. Furthermore, the correlation coefficients of the measurements obtained by Pentacam and Corvis, Pentacam and IOL-Master 700, as well as Corvis and IOL-Master 700 were determined 0.957, 0.964, and 0.948, respectively (P < 0.0001). CONCLUSION: The results from this study indicate that the interchangeable use of these three devices is not appropriate due to statistically significant differences and broad limits of agreement among the three devices, especially between Corvis and IOL-Master 700.
format Online
Article
Text
id pubmed-9128434
institution National Center for Biotechnology Information
language English
publishDate 2022
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-91284342022-05-25 Differences and Limits of Agreement among Pentacam, Corvis-ST, and IOL-Master 700 Optical Biometric Devices regarding Central Corneal Thickness Measurements Rajabi, Sattar Asharlous, Amir Riazi, Abbas Khabazkhoob, Mehdi Moalej, Alireza J Curr Ophthalmol Original Article PURPOSE: To investigate the differences and limits of agreement in measuring corneal thickness using Pentacam, Corvis, and intraocular lens (IOL)-Master 700 devices. METHODS: This study was conducted on 37 right eyes of 21 males and 16 females (n = 37) with a mean age of 52.11 ± 6.30 years. The central corneal thickness was measured using three optical biometric devices, including Pentacam, Corvis, and IOL-Master 700. The inclusion criteria were normal eyes without any ophthalmological abnormalities, history of ocular pathology, or ocular surgery. The data obtained from these three devices were compared two by two. The correlation and agreement limits among them were analyzed using statistical techniques. RESULTS: The mean standard deviation differences between Pentacam and Corvis, Pentacam and IOL-Master 700, as well as Corvis and IOL-Master 700 regarding the corneal thickness measurement, were 22.13 ± 8.05, 7.91 ± 8.02, and 14.21 ± 9.85 μm, respectively, which were statistically significant (P < 0.0001). Based on the investigation of the limits of agreement according to the Bland Altman method, the corresponding values between Pentacam and Corvis, Pentacam and IOL-Master 700, and Corvis and IOL-Master 700 were -16.2 to +15.4, -15.8 to +16.3, and -20.1 to +20.0 μm, respectively. Furthermore, the correlation coefficients of the measurements obtained by Pentacam and Corvis, Pentacam and IOL-Master 700, as well as Corvis and IOL-Master 700 were determined 0.957, 0.964, and 0.948, respectively (P < 0.0001). CONCLUSION: The results from this study indicate that the interchangeable use of these three devices is not appropriate due to statistically significant differences and broad limits of agreement among the three devices, especially between Corvis and IOL-Master 700. Wolters Kluwer - Medknow 2022-04-16 /pmc/articles/PMC9128434/ /pubmed/35620377 http://dx.doi.org/10.4103/joco.joco_96_21 Text en Copyright: © 2022 Journal of Current Ophthalmology https://creativecommons.org/licenses/by-nc-sa/4.0/This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Rajabi, Sattar
Asharlous, Amir
Riazi, Abbas
Khabazkhoob, Mehdi
Moalej, Alireza
Differences and Limits of Agreement among Pentacam, Corvis-ST, and IOL-Master 700 Optical Biometric Devices regarding Central Corneal Thickness Measurements
title Differences and Limits of Agreement among Pentacam, Corvis-ST, and IOL-Master 700 Optical Biometric Devices regarding Central Corneal Thickness Measurements
title_full Differences and Limits of Agreement among Pentacam, Corvis-ST, and IOL-Master 700 Optical Biometric Devices regarding Central Corneal Thickness Measurements
title_fullStr Differences and Limits of Agreement among Pentacam, Corvis-ST, and IOL-Master 700 Optical Biometric Devices regarding Central Corneal Thickness Measurements
title_full_unstemmed Differences and Limits of Agreement among Pentacam, Corvis-ST, and IOL-Master 700 Optical Biometric Devices regarding Central Corneal Thickness Measurements
title_short Differences and Limits of Agreement among Pentacam, Corvis-ST, and IOL-Master 700 Optical Biometric Devices regarding Central Corneal Thickness Measurements
title_sort differences and limits of agreement among pentacam, corvis-st, and iol-master 700 optical biometric devices regarding central corneal thickness measurements
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9128434/
https://www.ncbi.nlm.nih.gov/pubmed/35620377
http://dx.doi.org/10.4103/joco.joco_96_21
work_keys_str_mv AT rajabisattar differencesandlimitsofagreementamongpentacamcorvisstandiolmaster700opticalbiometricdevicesregardingcentralcornealthicknessmeasurements
AT asharlousamir differencesandlimitsofagreementamongpentacamcorvisstandiolmaster700opticalbiometricdevicesregardingcentralcornealthicknessmeasurements
AT riaziabbas differencesandlimitsofagreementamongpentacamcorvisstandiolmaster700opticalbiometricdevicesregardingcentralcornealthicknessmeasurements
AT khabazkhoobmehdi differencesandlimitsofagreementamongpentacamcorvisstandiolmaster700opticalbiometricdevicesregardingcentralcornealthicknessmeasurements
AT moalejalireza differencesandlimitsofagreementamongpentacamcorvisstandiolmaster700opticalbiometricdevicesregardingcentralcornealthicknessmeasurements