Biomechanical effect of anatomical tibial component design on load distribution of medial proximal tibial bone in total knee arthroplasty: finite element analysis indicating anatomical design prevents stress-shielding

AIMS: This study aimed to identify the effect of anatomical tibial component (ATC) design on load distribution in the periprosthetic tibial bone of Koreans using finite element analysis (FEA). METHODS: 3D finite element models of 30 tibiae in Korean women were created. A symmetric tibial component (...

Descripción completa

Detalles Bibliográficos
Autores principales: Cho, Byung W., Kang, Kyoung-Tak, Kwon, Hyuck M., Lee, Woo-Suk, Yang, Ick H., Nam, Ji H., Koh, Yong-Gon, Park, Kwan K.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The British Editorial Society of Bone & Joint Surgery 2022