A Magnetorheological Duckbill Valve Micropump for Drug Delivery Applications
In this study, we propose a duckbill valve microfluidic pump that relies on an electromagnetic actuation mechanism. An FEA/CFD-based approach was adopted for the design of the device due to the coupled electromagnetic–solid–fluid interactions in the device. The simulation methodology was confirmed w...