Selective Ion Capturing via Carbon Nanotubes Charging

[Image: see text] We present a phenomenon consisting of the synergistic effects of a capacitive material, such as carbon nanotubes (CNTs), and an ion-selective, thin-layer membrane. CNTs can trigger a charge disbalance and propagate this effect into a thin-layer membrane domain under mildly polariza...

Descripción completa

Detalles Bibliográficos
Autores principales: Wiorek, Alexander, Cuartero, Maria, Crespo, Gaston A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American Chemical Society 2022
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9161223/
https://www.ncbi.nlm.nih.gov/pubmed/35579547
http://dx.doi.org/10.1021/acs.analchem.2c00797
Descripción
Sumario:[Image: see text] We present a phenomenon consisting of the synergistic effects of a capacitive material, such as carbon nanotubes (CNTs), and an ion-selective, thin-layer membrane. CNTs can trigger a charge disbalance and propagate this effect into a thin-layer membrane domain under mildly polarization conditions. With the exceptional selectivity and the fast establishment of new concentration profiles provided by the thin-layer membrane, a selective ion capture from the solution is expected, which is necessarily linked to the charge generation on the CNTs lattice. As a proof-of-concept, we investigated an arrangement based on a layer of CNTs modified with a nanometer-sized, potassium-selective membrane to conform an actuator that is in contact with a thin-layer aqueous solution (thickness of 50 μm). The potassium ion content was fixed in the solution (0.1–10 mM range), and the system was operated for 120 s at −400 mV (with respect to the open circuit potential). A 10-fold decrease from the initial potassium concentration in the thin-layer solution was detected through either a potentiometric potassium-selective sensor or an optode confronted to the actuator system. This work is significant, because it provides empirical evidence for interconnected charge transfer processes in CNT–membrane systems (actuators) that result in controlled ion uptake from the solution, which is monitored by a sensor. One potential application of this concept is the removal of ionic interferences in a sample by means of the actuator to enhance precision of analytical assessments of a charged or neutral target in the sample with the sensor.